Testing instructions for KISS™ and KISS™ Multilead
Testing Electrical Function (KISS™ Multilead)
Resistance between the leadwire connector and the KISS™ electrode connector. The
measured resistance must be 10 kohm ±1kohm.
CAUTION:
It is prudent to make this measurement without electrodes as the silver-silver
chloride electrodes must not come into contact with other metals.
Leadwire connector (KISS™ Multilead)
KISS™ - Disassembly, Assembly
Disassembling and Assembling the Controller and the End
Piece on the Distributor
Remove all electrode leads on the distributor. There are 4 spring catches to be opened
for each lead.
First of all lift up 2 spring catches on one side with a pair of pointed tweezers (Caution:
Do not overstretch!) and lift component slightly, then lift up the next 2 spring catches
on the other side and remove component (controller or end piece).
22745010H
KISS™ / KISS™ Multilead
25
Содержание KISS
Страница 14: ...Functional Description 14 KISS KISS Multilead 22745010H ...
Страница 15: ...2 Operating Controls and System Components KISS 22745010H KISS KISS Multilead 15 ...
Страница 18: ...Operating Controls and System Components 18 KISS KISS Multilead 22745010H ...
Страница 36: ...Testing instructions for KISS and KISS Multilead 36 KISS KISS Multilead 22745010H ...
Страница 40: ...Maintenance Cleaning and Troubleshooting KISS KISS Multilead 40 KISS KISS Multilead 22745010H ...
Страница 44: ...Spare Parts List KISS KISS Multilead 44 KISS KISS Multilead 22745010H ...
Страница 52: ...Schematics 52 KISS KISS Multilead 22745010H ...
Страница 53: ......