Page
18
Chapter 2 — Operation
If you wish to perform dew point calibrations in the test chamber,
the display may be reprogrammed to read dew point.
B. CONTROL STATUS Indicator.
The green control indicator advises
you that the control loop that maintains the dew layer on the
chilled mirror is operating correctly, and that there is a proper layer.
When lit, you may consider it to be an indication that the chilled
mirror humidity measurement is valid. Conversely, if the green
indicator is not lit, you should wait until it lights before performing
calibrations.
C. STATUS Window (humidity).
A green letter “
S
” (for Service) in this
window advises the operator that the sensor mirror has become
contaminated, and requires cleaning.
If you are readjusting the sensor optical balance, a light bar appears
in this window. See
details.
This window will show various indications while reprogramming
the display units or analog outputs. See Appendix B.
D. DISPLAY Switch.
The
DISPLAY
Switch is normally used in the
center
OPER
ate position.
This rocker switch is used to reprogram measurement units and to
check the scale factor of analog outputs.
In the standard configuration, the analog output scale factors can
only be changed using the RS-232 port, to ensure that they are not
changed accidentally. If you wish to reconfigure the system so that
these parameters
can
be changed via the front panel, see Appendix
B.
E. BALANCE Switch.
This switch is normally operated in the lower
AUTO
matic position.
In the
AUTO
position, the instrument will go through a normal
Pacer® cycle every 12 hours, rebalancing the dew point sensor
bridge and determining whether it requires cleaning.
In the
MAN
ual center position, there will be no automatic Pacer
cycles, only those initiated manually (if desired).
Содержание HumiLab
Страница 43: ...Chapter 5 Maintenance and Troubleshooting Page 39 Figure 13 Mirror Check and Mirror Sensor Balancing Procedure ...
Страница 75: ...Appendix D Humidity Equations and Conversion Charts Page 71 Figure 16 Graphical Conversion Chart ...
Страница 76: ...Page 72 Appendix D Humidity Equations and Conversion Figure 17 Psychrometric Chart ...
Страница 80: ...Page 76 Appendix E Warranty information and return ...