Section 10-7 - Electrical Safety Tests
shows a typical format for recording the patient lead to ground leakage current.
Measurements should be recorded from each lead and lead combination under each set of test
conditions specified in:
•
•
, as applicable.
Record all data on the Inspection Certificate.
Table 10-17 Typical Data Format for Recording Patient Lead to Ground Leakage
System Under Test:
________________________
Date of Test:
_____________________
NOTE:
Values in italics font are given as examples only.
ON
NORM
OPEN
ON
REVERSE
CLOSED
ON
REVERSE
OPEN
OFF
NORM
CLOSED
OFF
NORM
OPEN
OFF
REVERSE
CLOSED
OFF
REVERSE
OPEN
Test Conditions
Patient Lead or Combination Measured
System Power
Grounding/PE
Right Arm
(RA)
Left Arm
(LA)
Left Leg
(LL)
RA+LA+LL
Mains on
open
RF on
open
Table 10-16 Testing Power Conditions
ECG Power
Meter’s Polarity Switch
Meter’s Neutral Switch