GE H
EALTHCARE
RAFT
V
OLUSON
E8 / V
OLUSON
E6
D
IRECTION
KTD102576, R
EVISION
7
DRAFT (A
UGUST
23, 2012)
S
ERVICE
M
ANUAL
7-18
Section 7-6 - How to use the Auto Tester program
Section 7-6 How to use the Auto Tester program
The following screen will appear.
10.)Select the “Disc-Save” icon. The file will be saved in D:\Export\AutotesterScripts\*.*.
11.)Press
ALT
+
D
to export system data and the needed logs.
Operation see:
Section 7-4-2-1 "Export System Data (by pressing the ALT + D key)" on page 7-10
12.)If desired, save the script file also on DVD/CD. Therefore:
-
Insert an empty DVD/CD+R/RW in the Drive and select the “Burn on CD” icon.
-
Enter a Filename.
-
After the DVD/CD write is finished click the OK button.
13.)Close the “Auto Tester” program.
7-6-1
Limitation of the Auto Tester
The following information will not be recorded. Depending on the moment the Auto Tester is activated,
this must be provided by the customer or the field engineer.
•
Which probe is in use and which probes are connected?
•
Which Mode is activated?
•
Which peripherals are connected (Dicom, Printer, etc..)?
Collected information from all steps above + exported system data and needed logs (
ALT
+
D
) can be
sent to the Make Center. With this information the Make Center can see how the customer is using the
system and reproduce potential failures.
NOTE:
Providing all information at once will help the Make Center find the root cause and speeds up finding a
solution for the customer.
Figure 7-21 Auto Tester Finished
NOTICE
The standalone recorded "Auto Tester" file makes only sense with the
ALT
+
D
log!
To analyze the workflow the exact date and time of occurrence needs to be documented!
“DVD/CD” (Burn)
icon
“DISC” (Save)
icon