Operating Instructions
28-9649-57
AA
31
Maintenance 5
Component maintenance 5.5
Bottom filter unit replacement
Adapter filter unit replacement
Step
Action
1
Use the wrench supplied to remove
the old bottom filter unit.
2
Wet the new bottom filter unit with ethanol and insert into the column tube.
3
Tighten the bottom filter.
Step
Action
1
Unscrew the two screws that hold the
adapter filter unit and remove it from
the adapter rod.
2
Gently pull out the filter unit.