Maintenance 5
Disassembly and assembly 5.4
5.4 Disassembly and assembly
Disassembly and assembly of the FineLINE 35
oligo
column is very straightforward and
therefore only basic instructions are necessary.
WARNING
Hazardous substances
. When using hazardous chemical and
biological agents, take all suitable protective measures, such as
wearing protective glasses and gloves resistant to the substances used.
Follow local and/or national regulations for safe operation and
maintenance of the system.
CAUTION
After assembly, ensure that the column is tight and not leaking.
NOTICE
When working with sealing, the nets and surfaces in contact with these
components:
- Never use sharp tools.
- Do not scratch any such components or surfaces.
NOTICE
If the support rods between the two end-pieces have been removed,
they must be re-tightened with a torque key to 1.45 to 1.55 Nm for safe
operation.
NOTICE
Before use, it is recommended to clean the column with 20% ethanol.
NOTICE
The inlet and outlet tubing are connected with 5/16” fittings. To avoid
bending the outlet tube, it should be tucked into one of the slots on the
bottom end-piece.
Содержание FineLINE 35 oligo
Страница 1: ...GE Healthcare Operating Instructions Original instructions FineLINE 35 oligo column ...
Страница 2: ......
Страница 4: ...Table of Contents 4 FineLINE 35 oligo column Operating Instructions 28 9649 57 AA ...
Страница 18: ...3 Installation 18 FineLINE 35 oligo column Operating Instructions 28 9649 57 AA ...
Страница 34: ...6 Troubleshooting 34 FineLINE 35 oligo column Operating Instructions 28 9649 57 AA ...
Страница 43: ......