Table 4: Indicator Status  

Status 

Warmer 
Covers 

Disposable 
Cartridge 

Power 
LED 
Indicator 

Temperature 
LED 
Indicator 

Audible 
Indicator 

Description 

Action Required 

Ready 

Open or 
Closed 

None 

Flashing 
Green 
every 3 
seconds 

Unlit 

None 

Warmer unit has 
power, but is not 
heating 

None 

Power up 
of device 

Closed 

In place 

Red ½ 
second 

Red ½ 
second 

 ½ second 
beep 

Successful device 
power up and over 
temperature circuit 
test  

Observe. If the 
LED does not flash 
red, replace the 
Warmer; and 
contact Technical 
Support. 

In 
Operation 

Closed 

In place 

Solid 
Green 

Flashing Blue 

None 

Infusate 
temperature is  
< 33 °C. 

None 

In 
Operation 

Closed 

In place 

Solid 
Green 

Solid Blue 

None 

Infusate 
temperature is  

 33 °C and < 35 °C. 

None 

In 
Operation 

Closed 

In place 

Solid 
Green 

Solid or 
Flashing Blue 
>30 seconds 

None 

Warmer is unable 
to heat the infusate 
within operational 
range. Infusate 
temperature is  
< 20 °C and/or the 
flow rate is > 200 
mL/min. 

Reduce the flow 
rate if possible. If 
there is no 
change in 
operational 
temperature, 
consider replacing 
the Warmer and 
contact Technical 
Support. 

In 
Operation 

Closed 

In place 

Solid 
Green 

Solid Green 

None 

Infusate 
temperature is  
≥ 35 °C and ≤ 42 °C. 

None 

In 
Operation 

Closed 

In place 

Solid 
Green 

Solid Yellow 

None 

Infusate (and/or 
ambient 
temperature) is > 
42 °C but less than 
an “Over Temp” 
condition.  

Observe. This 
state whereby the 
infusate is > 42 ºC 
should only be 
entered 
periodically during 
changes in flow 
rate or infusate 
temperature. 

In 
Operation 

Closed 

In place 

Flashing 
Red 
 
 

 

Continuous 
audible 
burst  

Internal failure in 
the Warmer 

Replace the 
Warmer if this 
occurs, and 
contact Technical 
Support. 

Continuous 
Operation 

Closed 

In place 

Solid 
Green 

Flashing Red  
 
High Priority 
Alarm 

Continuous 
audible 
burst  

Infusate (and/or 
ambient 
temperature) is > 
45 °C signifying an 
“Over Temp” 
condition. 

Stop the fluid flow, 
and slide the 
Warmer covers 
open to stop 
warming. Replace 
the Warmer if this 
occurs, and 
contact Technical 
Support. 

Содержание enFlow IV

Страница 1: ...enFlow IV Fluid Blood Warming System Operator s Manual Vital Signs a GE Healthcare Company...

Страница 2: ...ements If you have questions or concerns regarding this manual or product please contact one of the following for assistance Customer Service E mail vitalsignscustomerservice ge com Phone 1 800 932 07...

Страница 3: ...ds a reference point Close Di 2 ethylhexyl phthalate Free In transport applications it is advised to cushion and insulate the Warmer from the patient s skin and apply the Warmer as loosely as acceptab...

Страница 4: ...eclaring compliance with all applicable directives Safety EMC Machinery Medical and others Interek is accredited by OSHA as a NRTL as well as by the Standards Council of Canada This mark indicates tha...

Страница 5: ...STEM 9 ENFLOW CONTROLLER MODEL 120 SERIES INDICATORS AND OPERATION 11 ENFLOW WARMER MODEL 100 SERIES INDICATORS AND OPERATION 13 CLEANING THE ENFLOW IV FLUID BLOOD WARMING SYSTEM COMPONENTS 15 STORING...

Страница 6: ...ood The Controller serves as a power supply for the Warmer unit The Controller is designed to mount on an IV pole or sit on a table top The front panel includes a Controller reading in degrees Celsius...

Страница 7: ...the Disposable Cartridge from the Warmer Re prime the entire IV system using aseptic techniques Ensure all the air is removed from both the line and the Disposable Cartridge Replace the Disposable Ca...

Страница 8: ...hould have a base diameter of no less than 61 cm 24 in A Controller mounted too high on the IV pole may cause instability IV pole accessories or the attachment of fluid bags may also cause instability...

Страница 9: ...nFlow IV Fluid Blood Warming System a Place the Controller on a firm level surface or on an IV pole with an outside dimension of no greater than 3 0 cm 1 25 in Two Controllers may also be mounted next...

Страница 10: ...ion set to the patient and place the Disposable Cartridge into the Warmer h Completely close the covers on the Warmer by sliding them inward toward each other until the covers meet Upon closing the co...

Страница 11: ...s blankets or drapes l Opening the Warmer covers immediately stops the heating but not the flow m To turn off the device use the switch located at the rear of the controller enFlow Controller Model 12...

Страница 12: ...to Warmer Temp LED Fault detected System Fault XX If a system fault message is on the Controller display contact Technical Support Red High Priority Alarm Table 3 Controller Display Setup Mode Activi...

Страница 13: ...n the top of the Warmer there are two status indicator lights multicolored LEDs which reflect the following Power indicates the power and operational status of the Warmer Temperature indicates that th...

Страница 14: ...ational range Infusate temperature is 20 C and or the flow rate is 200 mL min Reduce the flow rate if possible If there is no change in operational temperature consider replacing the Warmer and contac...

Страница 15: ...taraldehyde based cleaners Chlorhexidine Cleaning the Warmer Wipe down and or wash 1 After each use clean the Warmer only as required In many instances it may only need to be wiped clean 2 If the warm...

Страница 16: ...nly approved cleaning solutions 2 Moisten a clean cloth with the cleaning solution do not spray or pour cleaning solutions directly on to the Controller 3 Wipe the surface of the Controller taking car...

Страница 17: ...ow of 100 20 mL min Wait for the temperature probes to stabilize Record the input fluid temperature Input fluid temperature 20 C 2 C Record the output fluid temperature Output fluid temperature 40 C 2...

Страница 18: ...Service Representative will discuss repair replacement options You will be issued a Return Goods Authorization RGA number You will be instructed to return the Product in packaging sufficient to preven...

Страница 19: ...a maximum of 300 Watts Controller 110 120 or 220 240 VAC Temperature Set Point 40 C Over Temperature Set Point ASTM F 2172 02 Alarms IEC60601 1 8 2006 Input Current 5 A Input Frequency Range Warmer D...

Страница 20: ...limits are designed to provide reasonable protection against electromagnetic interference EMI in a typical medical installation The enFlow System generates uses and can radiate radio frequency energy...

Страница 21: ...kV I Os 2kV Mains 1kV I Os Mains power quality should be that of a typical commercial or hospital environment Surge EN IEC 61000 4 5 1kV Differential 2kV Common 1kV Differential 2kV Common Mains power...

Страница 22: ...t and the enFlow The enFlow 100 with enFlow 120 is intended for use in the electromagnetic environment in which radiated disturbances are controlled The customer or user of the enFlow 100 with enFlow...

Страница 23: ...e rain chamber 9 Unplug the power supply and wipe excess water away from the outside 10 Perform validation test to confirm proper operation Record serial numbers of test items TPTR_083 11 Using a scre...

Страница 24: ...in degrees C as well as containing a keypad which controls the clock and the mute feature Additionally it converts AC line power to 28 Volts DC and is used as a power source for the Warmer Disposable...

Страница 25: ...reen 42 C Green No 43 C Off Yellow 43 C Yellow No 44 C Off Yellow 44 C Yellow No 45 C Off Red Flashing High Priority Alarm after 20 seconds 45 C after 20 seconds Red Flashing Over Temp message After 2...

Страница 26: ...00EU Disposable cartridge 980202EU Disposable cartridge with IV extension set 980305VS Warmer Holder 980309VS Warmer Cord Clip 980307 enCheck Alarm testing Tool 44000024 Operator s manual USA 44000060...

Отзывы: