[EN] English - K0449
Electrical and IDOS operations 3-5
Issue 1
3.2.6 Example procedure: Source DC voltage (CH1)
These examples (A and B) show Channel 1 (CH1) set up to
source a DC voltage.
1.
Set the applicable software options; see
Section 3.2.1
(Procedure overview)
.
2.
Complete the electrical connections.
3.
To continue, set the applicable output value; see
Section 2.11.2
.
3.2.7 Example procedure: Measure or source current with internal 24V loop power
These examples (A and B) show Channel 2 (CH2) set up to
measure or source a current with internal loop power (24 V).
2
1
3
Source DC volts on
channel 1 (CH1)
Range: 0 to 12 V (DC)
Automation
:
Nudge
(
Section 2.11.2
)
A
Source DC millivolts on
channel 1 (CH1)
Range: 0 to 2000 mV (DC)
Automation
:
Nudge
(
Section 2.11.2
)
B
1
3
2
2
1
Measure current on
channel 2 (CH2) + internal
loop power (24 V)
Range: ±55 mA
A
Maximum: 24 mA
Source current on
channel 2 (CH2) + internal
loop power (24 V)
Range: 0 to 24 mA
Automation
:
Nudge
(
Section 2.11.2
)
2
B
3
1
Содержание Druck DPI 620
Страница 1: ...GE Sensing Inspection Technologies Druck DPI 620 advanced modular calibrator user manual K0449 ...
Страница 8: ...Issue 1 viii About this manual K0449 EN English ...
Страница 18: ...Issue 1 1 6 Instrument parts accessories and options K0449 EN English ...
Страница 38: ...Issue 1 2 20 Prepare the instrument K0449 EN English ...
Страница 52: ...Issue 1 3 14 Electrical and IDOS operations K0449 EN English ...
Страница 60: ...Issue 1 5 2 Instrument communications K0449 EN English ...
Страница 64: ...Issue 1 6 4 Datalog operation K0449 EN English ...
Страница 68: ...Issue 1 7 4 Documenting functions K0449 EN English ...
Страница 80: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Страница 82: ...Issue 1 10 2 General specification K0449 EN English ...
Страница 83: ...Customer service Visit our web site www gesensinginspection com ...
Страница 84: ......