Electrical and IDOS operations 3-9
Issue 1
3.
If necessary change the
Settings
and continue with the
measure operation.
Values to set
:
TC Type (Set the applicable TC)
CJ compensation (Mode: Automatic/Manual).
Automatic
uses the internal cold junction. Use
Manual
mode when you
want to use an external cold junction.
CJ Value. For
Manual
mode, set an applicable value. The
value is not used in
Automatic
mode.
Example B
1.
Set the applicable software options; see
Section 3.2.1
(Procedure overview)
.
2.
Complete the electrical connections.
3.
If necessary change the
Source Settings
and continue with
source operation.
Values to set
:
TC Type (Set the applicable TC)
CJ compensation (Mode: Automatic/Manual).
Automatic
uses the internal cold junction. Use
Manual
mode when you
want to use an external cold junction.
CJ Value. For
Manual
mode, set an applicable value. The
value is not used in
Automatic
mode.
4.
Automation
: Set the applicable output value; see
Section 2.11.2
.
3.2.11 Example procedure: Switch test
CH1, P1, P2 and IDOS functions use the CH2 switch
connections; CH2 functions use the CH1 switch connections.
Switch operation
If you set the
Switch Test
utility on one channel, the software
automatically sets up the other channel for the switch
connections.
Note:
If there is a measure or source function on the switch
connection channel, it is automatically disabled; you get a
screen message “... Function Disabled”.
If you try to set a measure or source function on the switch
connection channel, you get a message “... Function not set”.
Example
This example shows a thermocouple switch:
Содержание Druck DPI 620
Страница 1: ...GE Sensing Inspection Technologies Druck DPI 620 advanced modular calibrator user manual K0449 ...
Страница 8: ...Issue 1 viii About this manual K0449 EN English ...
Страница 18: ...Issue 1 1 6 Instrument parts accessories and options K0449 EN English ...
Страница 38: ...Issue 1 2 20 Prepare the instrument K0449 EN English ...
Страница 52: ...Issue 1 3 14 Electrical and IDOS operations K0449 EN English ...
Страница 60: ...Issue 1 5 2 Instrument communications K0449 EN English ...
Страница 64: ...Issue 1 6 4 Datalog operation K0449 EN English ...
Страница 68: ...Issue 1 7 4 Documenting functions K0449 EN English ...
Страница 80: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Страница 82: ...Issue 1 10 2 General specification K0449 EN English ...
Страница 83: ...Customer service Visit our web site www gesensinginspection com ...
Страница 84: ......