[EN] English - K0449
HART
®
device operations 8-5
Issue 1
2.
Set the
HART Resistor
in the
Advanced
menu:
3.
Complete the electrical connections and continue with the
HART menu operation; see
Section 8.5
.
8.4.2 External loop power
In this example, there is an external power supply.
1.
Set the functions for channel 2 and HART:
e
2
Set to “On”
Section 2.9
a
b
c
d
Home display
TAP
Measure current on
channel 2 (CH2)
Range: ±55 mA
HART Function is set
250
:
resistor is set
Maximum: 30V
e
1
a
b
d
c
Set HART
Section 2.10.4
Set Current; Measure
Section 2.10.4
Home display
TAP