Issue 1
4-6 Pressure indicator operation (MC 620)
K0449 - [EN] English
2.
Set the applicable software options; see
Section 4.4.1
(Procedure overview)
. This example shows two pressure
functions:
•
Pressure functions P1 and P2 are set up.
3.
To attach the external equipment, see
Section 4.3.1
.
4.5 Error indications
If the display shows <<<< (under range) or >>>> (over range):
•
Make sure that the range is correct.
•
Make sure that all the related equipment and connections
are serviceable.
<<<<<
Under range: The display shows this symbol for this condition:
Reading < Negative FS - (10% of negative FS)
>>>>>
Over range: The display shows this symbol for this condition:
Reading > Positive FS + (10% of positive FS)