![GE Druck ADTS 405 Скачать руководство пользователя страница 45](http://html.mh-extra.com/html/ge/druck-adts-405/druck-adts-405_service-manual_128378045.webp)
Druck ADTS 405 Service Manual
K171 Issue No. 1
Page 4-5
Key-pad Check
13.
This verifies the hand terminal key-pad operation.
(1)
Press the MACH key and check the display changes to show Altitude and Mach
measured values.
(2)
Press each key in turn in the order shown below and check that the hand terminal
beeps on each key press.
Note:
The display shows a warning message when the EPR key is pressed.
This message should be ignored. Do not press the QUIT key to clear the
warning message.
F1, F2, F3, F4, ALT, SPEED, MACH, EPR, ROC, RATE TIME, HOLD, RATE,
LEAK MEASURE, GROUND, PORT, LOCAL, PRINT, EXEC. TEST PROGRAM, HELP,
SETUP, UP ARROW, 7, 8, 9, DOWN ARROW, 4, 5, 6, 000, 1, 2, 3, CLEAR, ., 0, ENTER
14.
Change the units to mbar as follows:
(1)
Press SETUP then select [UNITS]. Repeatedly select [NEXT] until the display
shows “mbar”.
(2)
Select [SAVE] and then press QUIT twice and the display now shows Ps and Pt in
units of mbar.
Controller stability
15.
This section verifies the control stability.
(1)
Press CONTROL to turn the pressure controllers ON.
“AIM” replaces “LEAK MEASURE” on the display.
(2)
Enter a Ps aim of 510 mbar with a rate of change of 204 mbar/min. by pressing the
following: Ps, 5, 1, 0, ENTER, Ps RATE, 2, 0, 4, ENTER.
(3)
Enter a Qc aim of 0 mbar with a rate of change of 204 mbar/min. by pressing the
following: Qc, 0, ENTER, RATE, 2, 0, 4, ENTER.
(4)
Press Ps and wait for the aim values to be achieved.
(5)
If auto zero is enabled, wait for an auto zero to take place. During auto zero, the
valve state changes on the mimic panel and the display shows “ZERO” in place of
the airspeed aim value. The time to auto zero should be less than 1 minute from
achieving the aim value.
(6)
Wait for a minimum of 20 seconds after auto zero before proceeding.
(7)
Observe the measured values of Ps and Qc on the display for 1 min.
(8)
Check that the displayed value of Ps remains within ±0.068 mbar.
(9)
Check that the displayed value of Qc remains within ±0.068 mbar.
Note:
If a further auto zero takes place, allow the display to stabilise following the
auto zero before proceeding.
Содержание Druck ADTS 405
Страница 2: ... 2008 General Electric Company All rights reserved ...
Страница 8: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page vi ...
Страница 11: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page 1 3 Figure 1 2 General view Figure 1 1 General view ...
Страница 13: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page 1 5 Figure 1 3 ADTS 405 L N ...
Страница 24: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page 1 16 ...
Страница 48: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page 4 8 ...
Страница 54: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page 5 6 ...
Страница 58: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page 6 4 Figure 6 2 ADTS 405F Pump Rack 4 2 3 1 ...
Страница 61: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page 6 7 Figure 6 3 ADTS 405 6 4 10 5 11 9 3 2 3 3 ...