Druck ADTS 405 Service Manual
K171 Issue No. 1
Page 5-5
Completion
18.
When the calibration check is complete, repeatedly use the QUIT key to return to the first
calibration screen.
i.
If necessary, carry out a rate calibration or press QUIT can be pressed again to exit
calibration mode.
ii.
Select go to ground or stay at the current pressure.
Point
Calibration Pressure
inHg
mbar
1
1
34
2
5.1
173
3
10.2
345
4
20.4
691
5
30.5
1033
6*
38.7
1311
7«
50.9
1724
8«
63.1
2137
9«
77.4
2621
10«
89.6
3034
11«
103.4
3502
Table 5-1
Table 5-2
Calibration Check Points
Ps Calibration Check Points
(both channels)
Point
Calibration Pressure
inHg
mbar
1
1
34
2
5.1
173
3
10.2
345
4
15.3
518
5
20.4
691
6
25.4
860
7
30.5
1033
8*
38.7
1311
9«
50.9
1724
10«
63.1
2137
11«
80
2709
Point
Calibration Pressure
inHg
mbar
1
1
34
2
3.1
105
3
5.1
173
4
7.1
240
5
10.2
345
6
15.3
518
7
20.4
691
8
25.4
860
9
30.5
1033
10
34.6
1172
11
40
1355
Table 5-3
Calibration Check Points
Special version
(both channels)
* Ps END
« Pt value only
Содержание Druck ADTS 405
Страница 2: ... 2008 General Electric Company All rights reserved ...
Страница 8: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page vi ...
Страница 11: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page 1 3 Figure 1 2 General view Figure 1 1 General view ...
Страница 13: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page 1 5 Figure 1 3 ADTS 405 L N ...
Страница 24: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page 1 16 ...
Страница 48: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page 4 8 ...
Страница 54: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page 5 6 ...
Страница 58: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page 6 4 Figure 6 2 ADTS 405F Pump Rack 4 2 3 1 ...
Страница 61: ...Druck ADTS 405 Service Manual K171 Issue No 1 Page 6 7 Figure 6 3 ADTS 405 6 4 10 5 11 9 3 2 3 3 ...