Communications Panel Options
© 2017 General Electric Company – All Rights Reserved.
ADTS 405 R/F Mk2 Instruction Manual-English | 41
accessible within the communications adapter IP port at this time, however the functionality
of the port may be assessed and tested if required.
The system is supplied with factory default IP addresses as follows:
The ports are most simply tested using a network ‘ping’ to these addresses.
Connect the ADTS to a computer equipped with a network interface, using a UTP network
cable. The network interface should be configured with an IP address on the same subnet.
Example:
If the ADTS has the default IP addresses then the host network address should be configured
as 192.168.2.1/255.255.255. The method for doing this may vary but the following example
screens show the method under Windows 7. It is suggested to take careful note of the
original computer settings so they may be restored after the check.
Figure 4-7: Configuring Windows For ADTS Subnet IP Address
Table 4-1: Default IP Address Configuration
IP Address
Function
192.168.2.2
Supervisor
192.168.2.3
Communications Adapter
Содержание Druck ADTS 405 F Mk2
Страница 2: ......
Страница 16: ...xiv ADTS 405 R F Mk2 Instruction Manual English 2017 General Electric Company All Rights Reserved ...
Страница 18: ...xvi ADTS 405 R F Mk2 Instruction Manual English 2017 General Electric Company All Rights Reserved ...
Страница 118: ...Chapter 7 Reference 100 ADTS 405 R F Mk2 Instruction Manual English 2017 General Electric Company All Rights Reserved ...
Страница 134: ...Appendix A Menus 116 ADTS 405 R F Mk2 Instruction Manual English 2017 General Electric Company All Rights Reserved ...
Страница 139: ......