Chapter 2. Installation
10 | ADTS 405 R/F Mk2 Instruction Manual-English
© 2017 General Electric Company – All Rights Reserved.
3.
The power supply connector on the rear panel of the unit will not be accessible when
fitted in the equipment rack.
2.4 Positioning of the ADTS
2.4.1 ADTS 405R Mk2 Rack-mounted Version
It is important that the position of the ADTS 405R Mk2 in relation to the UUT is known. An
altitude correction must be made to allow for the difference in height between the reference
level, indicated on the mimic panel, and the UUT. See Section 7.3.11, “SETUP, ALT,” on page 87
for details of altitude correction.
Figure 2-2: ADTS 405R Mk2 Altitude Reference
The corrected altitude output equals the true altitude output increased by height difference.
See equation below:
Note:
The Δ
h
altitude correction value should be negative when the ADTS is positioned
higher than the unit under test.
WARNING
Observe the appropriate safety instructions and procedures
detailed in the aircraft maintenance manual and component maintenance
manuals.
1
Rack-mount ADTS 405R Mk2
2
UUT
h
ADTS
ADTS panel marked reference level
h
UUT
UUT reference level
∆
h
Altitude correction height difference
1
2
h
h
ADTS
h
UUT
h
h
ADTS
h
UUT
–
=
Содержание Druck ADTS 405 F Mk2
Страница 2: ......
Страница 16: ...xiv ADTS 405 R F Mk2 Instruction Manual English 2017 General Electric Company All Rights Reserved ...
Страница 18: ...xvi ADTS 405 R F Mk2 Instruction Manual English 2017 General Electric Company All Rights Reserved ...
Страница 118: ...Chapter 7 Reference 100 ADTS 405 R F Mk2 Instruction Manual English 2017 General Electric Company All Rights Reserved ...
Страница 134: ...Appendix A Menus 116 ADTS 405 R F Mk2 Instruction Manual English 2017 General Electric Company All Rights Reserved ...
Страница 139: ......