background image

CO

2

Technology

Supports Novametrix CapnoStat (mainstream) and CapnoFlex LF (low-flow sidestream) CO

2

 technologies

Principle of operation 

 

Non-dispersive infrared (NDIR) single beam optics,

 

 

 

dual wavelength and no moving parts

Warm-up time 

 

2 minutes warm-up time to meet accuracy specifications; waveform immediate

 

 

 

upon power up, calculated end tidal after two breaths

 

Cable length (mainstream)   

8 foot (2.4 m)

Sample line length (low-flow sidestream) 7 foot (2.1 m)

Information displayed

Inspired and expired CO

2

 concentrations in %, mmHg or kPa; respiratory rate, continuous CO

2

 waveform

Measurement range (at 760 mmHg at an ambient temperature of 25°C)

0-100 mmHg, 0-13%, 0-12.5 kPa
PiCO

2

 /FiCO

2

  0-50 mmHg, 0-6.5%, 0-6.25 kPa

Respiration rate range 

 

Low-flow SS  0-150 breaths/min

 

 

 

Mainstream  0-120 breaths/min

Accuracy (at 760 mmHg at an ambient temperature of 25°C)

MS 

 

 

±2 mmHg or 5%, whichever is greater

SS 

 

 

0-40 mmHg ± 2 mmHg; 41-70 mmHg ± 5% of reading;

 

 

 

71-100 mmHg ± 8% of reading; all specifications ± 12%

 

 

 

of actual from 80-150 BrPM

Display resolution 

 

1 mmHg

Rise time 

 

 

Less than 200 ms (low-flow sidestream); less than 60 ms

 

 

 

(mainstream adult reusable or SPU); less than 50 ms

 

 

 

(mainstream infant reusable or SPU)

Respiration rate accuracy   

± 1 breath/min

Compensations

Automatic barometric pressure 

± 25 mmHg from 530-785 mmHg

Operator-selectable O2/N2O compensation

Calibration

Mainstream   

 

No routine user calibration required. 15 second airway adapter zero performed

 

 

 

when changing  to a different style of airway adapter.

 

Low-flow sidestream 

 

No routine user calibration required

Airway adapters and sample lines – mainstream (airway adapters)

Types 

 

 

Adult reusable (standard), adult disposable, infant

Deadspace   

 

Adult reusable/disposable < 5 cc

 

 

 

Infant disposable < 1 cc

 

 

 

Taper meets ISO 5356-1

Low-flow sidestream airway adapters

Types 

 

 

Adult reusable (standard), adult disposable, infant

Deadspace   

 

Adult reusable/disposable < 7.3 cc infant disposable < 1 cc

Adult, pediatric and infant   

Nasal CO2 and nasal CO2/O2

Adult and pediatric 

 

Nasal/oral CO2 and nasal/oral CO2/O2

Alarms

CO

2

 

 

 

High inspired CO

2

; high/low expired CO

2

Respiratory rate 

 

Adjustable high and low

Содержание Dash 3000

Страница 1: ... revolutionize patient care and assessment by combining the most complete selection of gold standard patient monitoring parameters with leading edge cardiac technology Enterprise networking Hard wired and wireless network connectivity including access to CIS CVIS PACS RIS HIS and more than 350 beds without central station support contributes to the Dash monitors unprecedented ability to adapt to c...

Страница 2: ...Default and individual Silencing 1 minute current alarm only Pause 5 minutes in Adult ICU mode 3 minutes in Neonatal ICU mode and 5 minute 15 minute or permanent pause in OR mode Volume Default 70 dB measured at 1 meter Invasive blood pressure Number of channels 1 to 4 optional Transducer sites Arterial femoral artery pulmonary arterial central venous right atrial left atrial intracranial and spec...

Страница 3: ... is limited by resolution to 40 Hz 3 dB 25 mm s Specified upper frequency limits may vary by 2 Hz Diagnostic mode 0 67 Hz 0 4 dB to 100 Hz 3 dB For compliance with China 1 0 Hz 0 4dB to 75 Hz 3 dB National Standard Monitoring mode 0 67 0 4 dB to 40 Hz 3 dB Moderate mode 0 67 0 4 dB to 25 Hz 3 dB Maximum mode 5 0 Hz 0 3 dB to 25 Hz 3 dB Noise 30 µV referred to input Pacemaker detection rejection In...

Страница 4: ...jectate temperature and trial number Range Cardiac output 0 2 15 liters per minute Blood temperature 30 42 C Injectate temperature 0 30 C Accuracy Cardiac output 5 Blood temperature 0 2 C Injectate temperature 0 3 C Frequency response dc to 15 Hz 2 Hz Pulse oximetry Parameters monitored Arterial oxygen saturation SpO2 and peripheral pulse rate PPR SpO2 range Nellcor 1 100 Masimo 30 100 GE Ohmeda 3...

Страница 5: ...te 10 110 mmHg Pulse rate as displayed in tabular trends or 3 wave form display Adult 30 200 bpm Pediatric 30 200 bpm Neonate 30 200 bpm Other specifications Overall system accuracy Meets or exceeds SP 10 1992 AAMI standards Automatic cycle times 0 4 hours Tubing length 12 feet adult 8 feet neonatal Automatic cuff deflation Cycle time exceeding 3 minutes 90 seconds neonatal French mode Cycle time ...

Страница 6: ...all specifications 12 of actual from 80 150 BrPM Display resolution 1 mmHg Rise time Less than 200 ms low flow sidestream less than 60 ms mainstream adult reusable or SPU less than 50 ms mainstream infant reusable or SPU Respiration rate accuracy 1 breath min Compensations Automatic barometric pressure 25 mmHg from 530 785 mmHg Operator selectable O2 N2O compensation Calibration Mainstream No rout...

Страница 7: ...ency 2 4 to 2 5 GHz Transmit power 100 mW Data rate 802 11 1Mbps and 2Mbps per channel 802 11b 1 2 5 5 11 Mbps Radio technology Frequency hopping spread spectrum Communication protocol IEEE 802 11 or IEEEE 802 11b UL 1950 Listed ITE 9B97 CE Mark RF Standard US CAN FCC Part 15 Class B RSS 210 Europe ETSI EN 300 328 Japan RCD STD 33R Battery Battery type Exchangeable Lithium Ion Maximum number of ba...

Страница 8: ...dical Device Directive C22 2 No 601 1 2028034 002 2006 03 V2 0 Storage conditions Do not exceed Maximum 70 C 158 F at 95 relative humidity Minimum 40 C 40 F CO2 sensor 30 to 65 C 22 to 149 F Batteries 20 to 60 C 4 to 140 F Physical specifications Height Depth Width Weight Dash 5000 28 7 cm 11 3 in 23 9 cm 9 4 in 30 7 cm 12 2 in 6 4 kg 14 lbs Dash 4000 27 4 cm 10 8 in 24 3 cm 9 6 in 29 3 cm 11 5 in...

Отзывы: