Product specifications
Display
Size
Dash 3000 – 8.4 in., Dash 4000 – 10.4 in., Dash 5000 – 12.1 in.
Type
Active-matrix color LCD
Resolution
Dash 3000 and 4000: 640 by 480 dpi, Dash 5000: 800 by 600 dpi
Number of traces
7 (maximum)
Number of seconds/trace
4.9 at 25 mm/sec
Sweep speed
6.25, 12.5, 25 mm/sec (with erase bar)
Controls
TrimKnob
®
control
Five hard keys
Standard Silence Alarm, Print, NBP Go/Stop, Zero All and Power On/Off. Dash 5000 adds
Trends, NBP Auto, Admit/Discharge, Standby and Main View
Remote control option
Available
Alarms
Categories
Patient status and system status
Priority
4 levels – Crisis, Warning, Advisory and Message
Notification
Audible and visual
Setting
Default and individual
Silencing
1 minute, current alarm only
Pause
5 minutes in Adult ICU mode, 3 minutes in Neonatal ICU mode and 5 minute,
15 minute, or permanent pause in OR mode
Volume
Default 70 dB measured at 1 meter
Invasive blood pressure
Number of channels
1 to 4 (optional)
Transducer sites
Arterial, femoral artery, pulmonary arterial, central venous, right atrial, left atrial,
intracranial and special
Transducer requirements
Excitation voltage: 5 V dc ± 0.1%
Transducer output
5 µV/V/mmHg
Input specifications
Range
-25 mmHg to 300 mmHg
Offset
± 150 mmHg
Output specifications
Frequency response
dc to 50 Hz (-3 dB)
Zero balance range
± 150 mmHg
Zero balance accuracy
± 1 mmHg
Zero balance drift
± 1 mmHg over 24 hours
Accuracy
± 2% or ± 1 mmHg, whichever is greater
(exclusive of transducer)
Alarms
User-selectable upper and lower limits for systolic, diastolic and mean pressures