Product specifications
Display

Size 

 

 

Dash 3000 – 8.4 in., Dash 4000 – 10.4 in., Dash 5000 – 12.1 in.

Type 

 

 

Active-matrix color LCD

Resolution 

 

 

Dash 3000 and 4000: 640 by 480 dpi, Dash 5000: 800 by 600 dpi

Number of traces 

 

7 (maximum)

Number of seconds/trace 

 

4.9 at 25 mm/sec

Sweep speed   

 

6.25, 12.5, 25 mm/sec (with erase bar)

Controls

TrimKnob

®

 control

Five hard keys 

 

Standard Silence Alarm, Print, NBP Go/Stop, Zero All and Power On/Off. Dash 5000 adds 

 

 

 

Trends, NBP Auto, Admit/Discharge, Standby and Main View

Remote control option 

 

Available

Alarms

Categories 

 

 

Patient status and system status

Priority 

 

 

4 levels – Crisis, Warning, Advisory and Message

Notification   

 

Audible and visual

Setting 

 

 

Default and individual

Silencing 

 

 

1 minute, current alarm only

Pause 

 

 

5 minutes in Adult ICU mode, 3 minutes in Neonatal ICU mode and 5 minute,

  

 

 

15 minute, or permanent pause in OR mode

Volume 

 

 

Default 70 dB measured at 1 meter

Invasive blood pressure

Number of channels 

 

1 to 4 (optional)

Transducer sites 

 

Arterial, femoral artery, pulmonary arterial, central venous, right atrial, left atrial,

 

 

 

intracranial and special

Transducer requirements 

 

Excitation voltage: 5 V dc ± 0.1%

Transducer output 

 

5 µV/V/mmHg

Input specifications

Range 

 

 

-25 mmHg to 300 mmHg

Offset 

 

 

± 150 mmHg

Output specifications

Frequency response 

 

dc to 50 Hz (-3 dB)

Zero balance range 

 

± 150 mmHg

Zero balance accuracy 

 

± 1 mmHg

Zero balance drift 

 

± 1 mmHg over 24 hours

Accuracy 

 

 

± 2% or ± 1 mmHg, whichever is greater

 

 

 

(exclusive of transducer)

Alarms 

 

 

User-selectable upper and lower limits for systolic, diastolic and mean pressures

Содержание Dash 3000

Страница 1: ... revolutionize patient care and assessment by combining the most complete selection of gold standard patient monitoring parameters with leading edge cardiac technology Enterprise networking Hard wired and wireless network connectivity including access to CIS CVIS PACS RIS HIS and more than 350 beds without central station support contributes to the Dash monitors unprecedented ability to adapt to c...

Страница 2: ...Default and individual Silencing 1 minute current alarm only Pause 5 minutes in Adult ICU mode 3 minutes in Neonatal ICU mode and 5 minute 15 minute or permanent pause in OR mode Volume Default 70 dB measured at 1 meter Invasive blood pressure Number of channels 1 to 4 optional Transducer sites Arterial femoral artery pulmonary arterial central venous right atrial left atrial intracranial and spec...

Страница 3: ... is limited by resolution to 40 Hz 3 dB 25 mm s Specified upper frequency limits may vary by 2 Hz Diagnostic mode 0 67 Hz 0 4 dB to 100 Hz 3 dB For compliance with China 1 0 Hz 0 4dB to 75 Hz 3 dB National Standard Monitoring mode 0 67 0 4 dB to 40 Hz 3 dB Moderate mode 0 67 0 4 dB to 25 Hz 3 dB Maximum mode 5 0 Hz 0 3 dB to 25 Hz 3 dB Noise 30 µV referred to input Pacemaker detection rejection In...

Страница 4: ...jectate temperature and trial number Range Cardiac output 0 2 15 liters per minute Blood temperature 30 42 C Injectate temperature 0 30 C Accuracy Cardiac output 5 Blood temperature 0 2 C Injectate temperature 0 3 C Frequency response dc to 15 Hz 2 Hz Pulse oximetry Parameters monitored Arterial oxygen saturation SpO2 and peripheral pulse rate PPR SpO2 range Nellcor 1 100 Masimo 30 100 GE Ohmeda 3...

Страница 5: ...te 10 110 mmHg Pulse rate as displayed in tabular trends or 3 wave form display Adult 30 200 bpm Pediatric 30 200 bpm Neonate 30 200 bpm Other specifications Overall system accuracy Meets or exceeds SP 10 1992 AAMI standards Automatic cycle times 0 4 hours Tubing length 12 feet adult 8 feet neonatal Automatic cuff deflation Cycle time exceeding 3 minutes 90 seconds neonatal French mode Cycle time ...

Страница 6: ...all specifications 12 of actual from 80 150 BrPM Display resolution 1 mmHg Rise time Less than 200 ms low flow sidestream less than 60 ms mainstream adult reusable or SPU less than 50 ms mainstream infant reusable or SPU Respiration rate accuracy 1 breath min Compensations Automatic barometric pressure 25 mmHg from 530 785 mmHg Operator selectable O2 N2O compensation Calibration Mainstream No rout...

Страница 7: ...ency 2 4 to 2 5 GHz Transmit power 100 mW Data rate 802 11 1Mbps and 2Mbps per channel 802 11b 1 2 5 5 11 Mbps Radio technology Frequency hopping spread spectrum Communication protocol IEEE 802 11 or IEEEE 802 11b UL 1950 Listed ITE 9B97 CE Mark RF Standard US CAN FCC Part 15 Class B RSS 210 Europe ETSI EN 300 328 Japan RCD STD 33R Battery Battery type Exchangeable Lithium Ion Maximum number of ba...

Страница 8: ...dical Device Directive C22 2 No 601 1 2028034 002 2006 03 V2 0 Storage conditions Do not exceed Maximum 70 C 158 F at 95 relative humidity Minimum 40 C 40 F CO2 sensor 30 to 65 C 22 to 149 F Batteries 20 to 60 C 4 to 140 F Physical specifications Height Depth Width Weight Dash 5000 28 7 cm 11 3 in 23 9 cm 9 4 in 30 7 cm 12 2 in 6 4 kg 14 lbs Dash 4000 27 4 cm 10 8 in 24 3 cm 9 6 in 29 3 cm 11 5 in...