3300/16 Dual Vibration Monitor Operation Manual
38
NOTE:
The gap zero position voltage must be within the limits shown in the table below
for the applicable transducer and gap full scale range or your gap full scale range may
go outside the transducer OK limits.
GAP
FULL-SCALE
RANGES
3300 OR 7200 5 & 8mm OR
3300XL 8mm TRANSDUCER
w/external w/o
external
barriers
barriers
7200 11MM
TRANSDUCER
7200 14MM OR
3300 16mm HTPS
OR 3300 XL 11mm
TRANSDUCER
3000
TRANSDUCER
15-0-15 mils
-6.18 to-12.92 V
-6.3 to -12.8 V
NA
NA
-5.1 TO -7.5 V
25-0-25 mils
-8.10 to -11.00 V
-8.3 to -10.8V
-6.18 to -16.1 V
-5.8 to -13.3 V
NA
50-0-50 mils
NA
NA
-8.68 to -13.6 V
-8.3 to -10.8 V
NA
300-0-300µm
-5.57 to -13.53 V
-5.66 to -13.44 V
NA
NA
-4.46 TO -8.14 V
600-0-600µm
-7.84 to -11.26V
-8.02 to -11.08 V
-6.04 to -16.24 V
-5.66 to -13.44 V
NA
1000-0-1000µm NA
NA
-7.62 to -14.66 V
-7.24 to -11.86
NA
GAP
FULL-SCALE
RANGES
3300 RAM
TRANSDUCER
w/ external
w/o external
barriers barriers
15-0-15 mils
-4.98 to -7.62 V
-5.1 to -7.5 V
25-0-25 mils
NA
NA
50-0-50 mils
NA
NA
300-0-300µm
-4.37 to -8.23 V
-4.46 to -8.14 V
600-0-600µm NA
NA
1000-0-1000µm NA
NA