AMX−4+ FUNCTIONAL CHECK
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173224−100
GE HEALTHCARE
REV 8
19
SECTION 3
FUNCTIONAL CHECK
Requirements are pass or fail. Service is required if any Requirement is failed. After each
Requirement is verified, place a check in its box. Record notes and comments next to
the Requirement.
Inspect
Requirement
Inspector’s Notes
3−1
Power on
ON
OFF
□
Changing Key Switch position from off to on
makes the Operators Display flash, then display
TESTING COMPLETE.
Recharge if display reads
RECHARGE
RECOMMENDED
or
RECHARGE IMMEDIATELY.
□
Pressing
kVp
UP increases kVp display.
□
Pressing
kVp
DOWN decreases kVp display.
□
Pressing
mAs
UP increases mAs display.
□
Pressing
mAs
DOWN decreases mAs display.
3−2
Drive
□
Drive Enable bar raises and lowers easily.
□
Drives forward, reverse, left, and right.
□
Drive speed is easily controlled.
□
Drives only in reverse when bumper is pushed in.
□
Drive speed reduced when tube arm is
removed from latch.
□
Drive brakes release when brake release
switch is depressed.
3−3
Tube Column and
Arm
□
Horizontal Tube Arm Latch operates smoothly
and holds arm securely.
□
Vertical Tube Column rotates 270 degrees from
latch position to latch position.
□
Tube Arm moves smoothly from bottom to top
of tube column.
□
Tube Arm extends and retracts smoothly.
□
Tube Unit rotates 90 degrees to horizontal in
both directions.
□
Tube Unit rotates back 10 and forward 110
degrees.
□
Latch holds column in drive position.
□
Lock prevents vertical travel.
□
Lock prevents tube extension.
□
Lock prevents column rotation.