AMX−4+ FUNCTIONAL CHECK
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173224−100
GE HEALTHCARE
REV 8
31
Appendix 2: Test Equipment
This section contains the illustrations of the meters used in the IEC 62353 tests.
Fluke ESA612 or equivalent (For more details refer the product manuals and video
explanation at
http://www.flukebiomedical.com/
Fluke ESA612
Appendix 3: AC Power plug troubleshooting
Tools and Test Equipment
•
Standard tool kit
Required Conditions
•
Perform LOTO on the system. Leave the AC Plug (E1) exposed.
Procedure
•
Remove screws on the top of plug marked
•
Check the cable core if any open strands or broken wires, connect properly into plug
•
Ensure all ground connections are properly tightened.
•
Inspect cord reel box and termination of ground cables.
Note:
If system using molded power cord type then check ground connection, Cord reel box and termination
of ground cables.