AMX−4+ FUNCTIONAL CHECK
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173224−100
GE HEALTHCARE
REV 8
17
SECTION 2
VISUAL INSPECTION
Requirements are pass or fail. Service is required if any Requirement is failed. After each
Requirement is verified, place a check in its box. Record notes and comments next to
the Requirement.
Inspect
Requirement
Inspector’s Notes
2−1
Operator’s
Console
□
Switch and display markings clearly legible.
□
Covering over kVp Switch, mAs Switch,
and Displays free from cuts or punctures
□
Verify good condition of ON and OFF
symbols on console label.
□
Verify
good condition of X−ray
emission
symbol on console label.
2−2
Collimator
□
Field Light and SID markings clearly legible.
□
Field Light switch and Field Size knobs not
loose or damaged.
□
Guard Arms perpendicular to collimator
bottom.
□
SID indicating tape attached to left side.
□
Rating Plate installed on right side of
collimator.
□
Rating Plate legible.
□
Rotational Detent knob on right side of
Collimator.
□
Must be no movement between collimator
and tube unit.
□
Verify good condition of Transverse (Lateral)
AND Longitudinal Collimator Blade symbols
on collimator labels.
□
Verify good condition of Field Light symbol
on collimator label.
□
Verify good condition of Positioning Control
Lock symbol on labels, one on EACH
collimator handle.