AMX−4+ FUNCTIONAL CHECK
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173224−100
GE HEALTHCARE
REV 8
25
6.
Select IEC62353 standard is displayed, press
F1
button, on the blue screen press
Standard
to confirm the standard.
7.
Press
F4
button, on the blue screen press
Done
to complete the standard setup.
8.
Push
to reveal the resistance function menu.
9.
Connect one end of a test lead to the V/ Ω /A jack.
10.
If you use an accessories probe, connect it to the other end of the test lead and put the probe tip
into the
∅
/Null jack. If you use an alligator clip accessory, connect it to the other end of the test
lead, put the null post adapter in the
∅
/Null jack, and clamp the alligator clip to the null post
adapter.
Note:
The
∅
/Null jack does not accept the test leads supplied with the Product.