AMX−4+ FUNCTIONAL CHECK
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173224−100
GE HEALTHCARE
REV 8
15
1-4
Inspection
To assure continued performance of this
x−ray
equipment, a periodic maintenance
program must be established in accordance with
2173227−100
AMX−4+
Periodic
Maintenance
. It is the
owner’s
responsibility to supply or arrange for this service.
Functional Check verifies that the AMX−4+ is functioning
correctly, that wear, or
damage has not been sustained during use, that diagnostic
self−test
can be passed, and
x−rays
can be made. It does not indicate that the unit is performing to specifications.
Refer to Direction 2173227−100,
AMX−4+
Periodic Maintenance,
for tests that verify
specified performance.
These checks are divided into three parts: Visual Inspection, Functional Check, and
Power Cord Check. Visual Inspection confirms legibility of operator indicators, and
general safety. Functional Check confirms equipment operation and function. Power
Cord Check verifies the integrity of the power cord for safe operation.
At the start of each shift, a Visual Inspection and Functional Check should be performed
to verify operation. Functional Check should also be used to verify operation after the
AMX−4+ is
serviced.