5 Setups, alignments and adjustments
32551-IMG rev 6
GE Healthcare
83
4
Make exposures (do not develop the film) and adjust the trimmer
R17
(CCW to increase frequency) on the AEC board so that the AEC
frequency will be appr.
32kHz
(that is the second value in the mAs
window reading about
320
).
Fig 5.14.
Location of AEC adjustment potentiometer
Also the exposure time is shown in the timer display (for example 315 =
315 ms, 1.54 = 1.54 sec.
5.2.16
Adjusting the optical density (=O.D)
1
Reload the cassette with a new film and make a new exposure with
the
clinical kV
(for example 26kV). Develop the film and mark the
optical density of the film (=O.D) to the graph next page.
2
Alter the AEC Gain parameter ‘g in the service program
#64
to 1.10,
reload the cassette with a new film and make a new exposure.
Develop the film and mark the optical density of the film (=O.D) to the
graph next page.
Содержание Alpha RT 32551
Страница 1: ...GE Healthcare Alpha RT Installation Instructions 32551 IMG rev 6 0459 ...
Страница 2: ......
Страница 4: ......
Страница 42: ...3 Installation 36 GE Healthcare 32551 IMG rev 6 ...
Страница 51: ...4 Alpha RT softstart procedure 32551 IMG rev 6 GE Healthcare 45 Fig 4 10 Inverter board ...
Страница 52: ...4 Alpha RT softstart procedure 46 GE Healthcare 32551 IMG rev 6 ...
Страница 71: ...5 Setups alignments and adjustments 32551 IMG rev 6 GE Healthcare 65 Fig 5 2 Magnetic brake ...
Страница 94: ...5 Setups alignments and adjustments 88 GE Healthcare 32551 IMG rev 6 ...
Страница 96: ...6 Alpha RT installation step by step 90 GE Healthcare 32551 IMG rev 6 ...
Страница 97: ......
Страница 98: ......