5 Setups, alignments and adjustments
32551-IMG rev 6
GE Healthcare
49
INSTRUMENTARIUM Imaging, MGF 101/110 information, SW v.7.x1a, 5-Feb-2001
========================================================================
DEVICE#: 05000 ALPHA MODE: RT(5)
COUNTRY CODE: 011 ANODE BRAKING IS OFF
CARM BOARD: MCU
RADIATION EFFECIENCY 10.0(Mo), 10.0(Rh) mR/mAs at 25 kV
AUTOREL TIME(s): 006 SW gain for AEC-board: 1.00
LABEL TYPE: 000 I/O DEVICE: D16+PC
LABEL TEXT IS:
IQC multiplier: 1.00 IQC MODE: EXPOSURE WITH 027 kV
WANTED DENSITY: 1.30
DEFAULT FILTER: Molybdenium
EXPOSURE COUNTER: 0000000
ERROR CNT: E01 E02 E03 E04 E07 E08 E09 E10
00000 00000 00000 00000 00000 00000 00012 00000
E11 E12 SLO ER1 ER2 ER3 ER4 ER5 EAF SYS
00011 00011 00000 00000 00001 00000 00000 00000 00001 00000
PROGRAMMED VALUES: Full
BUCKY(Mo) MAGN(Mo) CONTACT(Mo)
kV den kV den kV den
A= 00004 00100 00004 00100 00004 00100
B= 00004 00100 00004 00100 00004 00100
C= 00004 00100 00004 00100 00004 00100
BUCKY(Rh) MAGN(Rh) CONTACT(Rh)
kV den kV den kV den
A= 00004 00100 00004 00100 00004 00100
B= 00004 00100 00004 00100 00004 00100
C= 00004 00100 00004 00100 00004 00100
PROGRAMMED VALUES: Semi
BUCKY(Mo) MAGN(Mo) CONTACT(Mo)
kV den kV den kV den
A= 00027 00100 00027 00100 00027 00100
B= 00027 00100 00027 00100 00027 00100
C= 00027 00100 00027 00100 00027 00100
BUCKY(Rh) MAGN(Rh) CONTACT(Rh)
kV den kV den kV den
A= 00027 00100 00027 00100 00027 00100
B= 00027 00100 00027 00100 00027 00100
C= 00027 00100 00027 00100 00027 00100
PROGRAMMED VALUES: Man
BUCKY(Mo) MAGN(Mo) CONTACT(Mo)
kV mAs kV mAs kV mAs
00027 00012 00027 00012 00027 00012
BUCKY(Rh) MAGN(Rh) CONTACT(Rh)
kV mAs kV mAs kV mAs
00027 00012 00027 00012 00027 00012
USER DENSITY STEP: 0.10
FULL-AEC TARGET TIME var limits:
BUCKY: Mo: 0.25 +0.10 0.50 3.50
BUCKY: Rh: 0.25 +0.10 0.50 3.50
MAGN: Mo: 0.25 +0.30 0.50 5.00
MAGN: Rh: 0.25 +0.30 0.50 5.00
CONTACT:Mo: 0.25 +0.10 0.50 3.50
CONTACT:Rh: 0.25 +0.10 0.50 3.50
THICKNESS COMPENSATION variables:
BUCKY: Mo: 1.00 1.00 1.00 250 250
BUCKY: Rh: 1.00 1.00 1.00 250 250
MAGN: Mo: 1.00 1.00 1.00 250 250
MAGN: Rh: 1.00 1.00 1.00 250 250
CONTACT:Mo: 1.00 1.00 1.00 250 250
CONTACT:Rh: 1.00 1.00 1.00 250 250
KV COMPENSATION:
kV: 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35
BUCKY:
Mo: 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100
Rh: 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100
MAGN:
Mo: 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100
Rh: 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100
CONTACT:
Mo: 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100
Rh: 100 100 100 100 100 100 098 100 100 100 100 100 100 100 100 100
FULL-AEC kV-TABLES:
mm: 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95
BUCKY:
Mo: 23 23 23 23 24 24 24 25 25 25 26 26 26 27 27 27 28 28 29 30
Rh: 23 23 23 23 24 24 24 25 25 25 26 26 26 27 27 27 28 28 29 30
MAGN:
Mo: 25 25 25 25 25 25 25 25 26 26 26 27 27 28 28 28 29 29 30 30
Rh: 25 25 25 25 25 25 25 25 26 26 26 27 27 28 28 28 29 29 30 30
CONTACT:
Mo: 23 23 23 23 23 23 24 24 24 25 25 25 26 26 26 27 27 27 28 28
Rh: 23 23 23 23 23 23 24 24 24 25 25 25 26 26 26 27 27 27 28 28
EPROM CHECKSUM: Calc = 05925 Set = 05925
Содержание Alpha RT 32551
Страница 1: ...GE Healthcare Alpha RT Installation Instructions 32551 IMG rev 6 0459 ...
Страница 2: ......
Страница 4: ......
Страница 42: ...3 Installation 36 GE Healthcare 32551 IMG rev 6 ...
Страница 51: ...4 Alpha RT softstart procedure 32551 IMG rev 6 GE Healthcare 45 Fig 4 10 Inverter board ...
Страница 52: ...4 Alpha RT softstart procedure 46 GE Healthcare 32551 IMG rev 6 ...
Страница 71: ...5 Setups alignments and adjustments 32551 IMG rev 6 GE Healthcare 65 Fig 5 2 Magnetic brake ...
Страница 94: ...5 Setups alignments and adjustments 88 GE Healthcare 32551 IMG rev 6 ...
Страница 96: ...6 Alpha RT installation step by step 90 GE Healthcare 32551 IMG rev 6 ...
Страница 97: ......
Страница 98: ......