5 Setups, alignments and adjustments
32551-IMG rev 6
GE Healthcare
77
1
Check that the fast turn-around times are equal at both ends. If
adjustment is necessary, move the slotted metal sheet travelling
through the optocoupler.
2
The grid travelling time is adjusted from
R5
located on the bucky-
control board. In the CC position travel time of the grid should be
1.6
+/- 0.1 sec.
The fast travel time should be 0.17 +/- 0.1 sec.
3
Make sure that the grid moves well in lateral positions (C-arm in +/-
90°angle) as well.
Fig 5.12.
Bucky control board
5.2.14
AEC adjustment
Test equipment needed: Sensitometer, Densitometer, Acrylic blocks, Film
& Cassette.
1
Attach Bucky device.
2
Press
Den+
&
Den-
for 3 sec. to enter program (Pr) mode.
CAUTION!
Store all default values in the tables on the next pages (molybdenum filter)
and (rhodium filter).
3
Press
Autoexp
to
AA
mode and channel a. Select
Molybdenum
filter.
4
Select max. kV change range of initial kV (factory default “4 kV) with
kV-
and
kV+
keys. Value can be selected between 0...9 kV. 0 kV
means that initial kV does not change at all during the exposure.
Value 9 means that kV can decrease or increase 9 kV from initial kV
(which was selected according the breast thickness) during
exposure.
5
Select the internal density correction factor within a range of 50% to
200% (0.50 on screen means 50% and 2.00 on mAs-screen means
200% change to exposure time). The default value is 1.00.
Содержание Alpha RT 32551
Страница 1: ...GE Healthcare Alpha RT Installation Instructions 32551 IMG rev 6 0459 ...
Страница 2: ......
Страница 4: ......
Страница 42: ...3 Installation 36 GE Healthcare 32551 IMG rev 6 ...
Страница 51: ...4 Alpha RT softstart procedure 32551 IMG rev 6 GE Healthcare 45 Fig 4 10 Inverter board ...
Страница 52: ...4 Alpha RT softstart procedure 46 GE Healthcare 32551 IMG rev 6 ...
Страница 71: ...5 Setups alignments and adjustments 32551 IMG rev 6 GE Healthcare 65 Fig 5 2 Magnetic brake ...
Страница 94: ...5 Setups alignments and adjustments 88 GE Healthcare 32551 IMG rev 6 ...
Страница 96: ...6 Alpha RT installation step by step 90 GE Healthcare 32551 IMG rev 6 ...
Страница 97: ......
Страница 98: ......