Precautions
WARNING
Electrical shock hazard. All repairs should be done by service
personnel authorized by GE. Do not open any covers or replace
parts unless specifically stated in the user documentation.
WARNING
Use only approved parts. Only spare parts and accessories that
are approved or supplied by GE may be used for maintaining or
servicing the product.
WARNING
For continued protection against injury risks due to fluid jets, burst
pipes or potentially explosive atmosphere, the user must test the
piping system for leakage at maximum operating pressure.
•
Always perform a leakage test after assembly or maintenance.
•
Always perform a leakage test before operation or CIP.
WARNING
Hazardous chemicals during run. When using hazardous chemi-
cals, flush the entire system tubing with distilled water, before ser-
vice and maintenance.
WARNING
Decontaminate before maintenance. To avoid personnel being
exposed to potentially hazardous substances, make sure that the
column is properly decontaminated and sanitized before mainte-
nance or service.