5.5.2
Replace the Mixer module
Maintenance interval
Replace the Mixer module when the existing module malfunctions.
Required material
The following is required:
•
Mixer module (available as a spare part from GE)
•
7 mm socket tool
Procedure
Follow the steps below to replace the Mixer module.
Action
Step
Disconnect the tubing from the mixer chamber.
1
Pull the mixer chamber away from the module front.
2
Remove the retaining screw that holds the module in the instrument chassis
(see the Mixer Kit Installation Instruction supplied with the replacement
mixer).
3
Install the new module as described in the instructions provided with the
Mixer Kit.
4