Description
Part
Outlets for fractionation
5 to 9
Function
The outlet valve module contains a number of membrane valves to control the flow path
during fractionation. For more information about fractionation, see
. The first outlet port denoted 1-Waste is used as waste outlet in all
pre-defined methods.
Up to five outlet valves can be installed in the system, giving up to 15 outlets.
If additional outlet valve modules are installed, the outlets that lead to the next
module must be connected to the inlet of next module inlet. Keep the tubing
between the outlet valve modules as short as possible. For more information,
see separate installation instructions for the outlet valve module.
Note:
Outlet tubing
The outlets are connected with SNAP connectors to outlet tubing with an inner diameter
of 3.2 mm.
When running at low flow rates or collecting small fractions, it is recommended to use
an adapter to be able to use outlet tubing with smaller diameter. If even smaller diameter
is required, use suitable nipple and ferrule to further reduce the diameter.
Reduced tubing diameter increases the system pressure.
Note:
58
ÄKTA pilot 600 User Manual 29274559 AA
3 Description of modules
3.12 Outlet valves