Column valve options
See the list in the image above.
The Waste 2 option is used, for example, during pump wash and air trap drain.
The Bypass 1 and Bypass 2 options represent different flow paths in the valve. When
priming and cleaning the system, make sure that Bypass 1 is selected first, followed by
Bypass 2.
The Bypass both option is used during system wash.
The first four column run options make it possible to run a column separately with up
or down flow.
The Column 1 to column 2 downflow and Column 2 to column 1 downflow options
make it possible to run two columns in series, for example, to maximize column capacity.
The two options can also be used to connect a Superloop, see
.
The two packing options are used for Intelligent packing, see
ÄKTA pilot 600 User Manual 29274559 AA
49
3 Description of modules
3.8 Column valve