Action
Step
In the Calibration dialog, choose the pressure sensor in the Monitor to
calibrate list.
3
If the Current value is outside the range ± 0.02 MPa (0.2 bar, 2.9 psi), click
Reset pressure.
4
Result: The pressure reading is reset to zero.
Reconnect the tubing to the pressure sensor.
5
ÄKTA pilot 600 User Manual 29274559 AA
103
5 Maintenance
5.4 Calibration procedures
5.4.1 Reset the pressure sensors