Optimize pressure control
settings
To optimize pressure control parameters, change the settings manually and test the
new settings with a manual run. Follow the steps below:
Action
Step
Open the Advanced:Pressure control parameters section of Manual instruc-
tions in System Control.
1
Adjust the parameters as required.
2
Activate pressure control and select the appropriate Pressure control pa-
rameter in the Pumps:System Flow section of Manual instructions.
3
Run Pumps:System flow at a suitable flow rate to test the effect of your
adjustments.
4
When you are satisfied, update the pressure control parameters in System
Settings so that the optimized settings will apply automatically in the future.
5
Set pressure control parameters
Follow the steps below to change the pressure control parameters in System:Settings.
Settings changed in this way will apply as default for both manual and method runs.
Action
Step
Open System:Settings in the UNICORN System Control module.
1
ÄKTA pilot 600 User Manual 29274559 AA
85
4 Operation
4.5 Pressure control