Section 8-14 - Cables replacement
Main Cable (Harness) installation (cont’d)
Table 8-393 Main Cable (Harness) Routing
Steps
Corresponding Graphic
1.
Lay out the cables as shown. DO NOT
cross any cables or allow any excess length
in the Lower OP Frame, except the XYZ
USB Cable (shown below). Make sure the
cables are not crossed coming into the
Lower OP Frame
.
Route the USB Cable for the XYZ Switch as
shown. Use the markings on the cable to
position in the tab.
2.
There should only be enough cable for the
cable ground shielding to be under the
Grounding Strap Clamp when it is secured.
Also, the USB cables have markings and
should be in the same area.
Notice that the cables are not crossed or
snug when routed into the Lower OP
Frame.
NOTE: Apply Clamp loosely, then connect
cables. Adjust lengths as necessary and
tighten the Clamp.
Содержание 5935000E10
Страница 1: ...LOGIQ E10 Basic Service Manual 5750007 1EN Rev 1 Version R1 2018 GENERAL ELECTRIC COMPANY ...
Страница 2: ......
Страница 3: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL i Important Precautions TRANSLATION POLICY ...
Страница 4: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL ii ...
Страница 5: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL iii ...
Страница 6: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL iv ...
Страница 7: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL v ...
Страница 8: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL vi ...
Страница 9: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL vii ...
Страница 10: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL viii ...
Страница 11: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL ix ...
Страница 12: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL x ...
Страница 13: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL xi ...
Страница 14: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL xii ...
Страница 18: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL xvi This page was intentionally left blank ...
Страница 34: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL xxxii Hardware Kit 9 49 ...
Страница 776: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL Index 4 ...
Страница 777: ......
Страница 778: ... 2018 General Electric Company GE Healthcare 9900 Innovation Drive RP 2156 Wauwatosa WI 53226 USA www gehealthcare com ...