D
IRECTION
5750007-1EN, R
EV
. 1
LOGIQ E10 B
ASIC
S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8 - 189
8.
Turn the tension adjustment screw (
8
), so that the
flat side faces the brake shoe (8a), to further release
the lock (brake) motor assembly.
NOTE: When replacing the brake on the right rear
side (
A
) and the brake on the left front side (
B
), the
tension adjustment screw (
C
) turns counter-
clockwise.
When replacing the brake on the right front side (
A
)
and the brake on the left rear side (
B
), the tension
adjustment screw (
D
) turns clockwise.
After the motor assembly is released, loosen the
retaining screw (
8b
) a little, then gently pry up on the
XY Hinge (
8c
) with a flat blade screwdriver. Alternate
both steps until the brake can be rocked out
completely.
The brake assembly needs to be gently pried out on
both ends, a little at a time, to rock it out of the arm
cavity.
(Alignment for removal {
E
}).
9.
Before installing the new lock motor and after the old
motor assembly is removed, place a piece of sticky
tape (circled) on the tension adjustment screw on the
replacement motor. The tape is for holding the screw
in place and keep it from falling, while inserting the
motor into the frogleg arm.
NOTE: Remove the tape after the replacement brake
is secured into the frogleg.
10.
Install a new lock motor assembly by aligning the XY
Hinge (
10
) into the slot in the brake shoe (
10a
).
When the lock motor assembly is properly aligned
and inserted, push the assembly in by the XY Hinge
as the retaining screw (
10b
) is tightened.
NOTE: The lock assembly may have to be rocked in
from side-to-side and end-to-end for proper
insertion.
11.
Reconnect the wire connectors from the lock motor assembly to the frogleg wiring (refer to Step 7 for
reference).
12.
Plug the power cord back into the wall outlet and power up the system.
Steps
Corresponding Graphic
Содержание 5935000E10
Страница 1: ...LOGIQ E10 Basic Service Manual 5750007 1EN Rev 1 Version R1 2018 GENERAL ELECTRIC COMPANY ...
Страница 2: ......
Страница 3: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL i Important Precautions TRANSLATION POLICY ...
Страница 4: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL ii ...
Страница 5: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL iii ...
Страница 6: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL iv ...
Страница 7: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL v ...
Страница 8: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL vi ...
Страница 9: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL vii ...
Страница 10: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL viii ...
Страница 11: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL ix ...
Страница 12: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL x ...
Страница 13: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL xi ...
Страница 14: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL xii ...
Страница 18: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL xvi This page was intentionally left blank ...
Страница 34: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL xxxii Hardware Kit 9 49 ...
Страница 776: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL Index 4 ...
Страница 777: ......
Страница 778: ... 2018 General Electric Company GE Healthcare 9900 Innovation Drive RP 2156 Wauwatosa WI 53226 USA www gehealthcare com ...