D
IRECTION
5750007-1EN, R
EV
. 1
LOGIQ E10 B
ASIC
S
ERVICE
M
ANUAL
Chapter 4 Functional Checks
4 - 27
ECG Checks
Introduction
The ECG capability on this LOGIQ E10 is intended as a trigger for measurements, but can also be
viewed on the screen.
Parts needed
-
ECG Harness, for P/Ns, see: Section, Options.
-
ECG Pads, (3 pc)
or
-
ECG simulator
Preparations
None
ECG Checks
3D/4D and Volume Navigation Functional Checks
For complete information, refer to the appropriate version of the LOGIQ E10 Basic User Manual,
Chapters 5 and 6, or the appropriate LOGIQ E10 Release Notes. See:
Section 1-5 "Label locations" on
Shear Wave Elastography (Shear Elasto) Functional Checks
If the Capacitor Pack is not present or the Power Supply to Capacitor Pack Cable is not connected, or
connected properly, the option to select the “Shear Elasto” soft button after pressing the Elasto button,
the Touch Panel key “Shear Elasto/strain Elasto” button WILL NOT be present.
•
Select a probe - L2-9-D, C1-6-D or ML4-16-D.
•
Select ELASTO button.
•
In the Touch Panel, select Shear Wave button if not already selected.
•
Verify that the Shear Wave button is available.
Table 4-10 ECG Checks
Step
Task
Expected Result(s)
1.
Select
Preset
, Select
Cardiac
, Select
Scan
.
2.
If you do not see the ECG line, select the
ECG
tab, you should see selection on touch
screen for DISPLAY ECG. Select and press.
You should then see the line.
3.
Plug in the ECG cables with nothing
attached so you pick up noise.
The line should display noise.
4.
Connect the ECG harness to the connector
on the front of the Ultrasound system.
The LOGIQ E10 displays a straight curve along
the bottom edge of the image sector on the
screen.
5.
Connect the three leads to a ECG simulator,
or fasten the three ECG Pads to your body
and connect the three leads to respective
ECG Pad.
When connecting, the signal on the screen will be
noisy. When the connection is completed, a
typical clean ECG signal is displayed.
Содержание 5935000E10
Страница 1: ...LOGIQ E10 Basic Service Manual 5750007 1EN Rev 1 Version R1 2018 GENERAL ELECTRIC COMPANY ...
Страница 2: ......
Страница 3: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL i Important Precautions TRANSLATION POLICY ...
Страница 4: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL ii ...
Страница 5: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL iii ...
Страница 6: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL iv ...
Страница 7: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL v ...
Страница 8: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL vi ...
Страница 9: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL vii ...
Страница 10: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL viii ...
Страница 11: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL ix ...
Страница 12: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL x ...
Страница 13: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL xi ...
Страница 14: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL xii ...
Страница 18: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL xvi This page was intentionally left blank ...
Страница 34: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL xxxii Hardware Kit 9 49 ...
Страница 776: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL Index 4 ...
Страница 777: ......
Страница 778: ... 2018 General Electric Company GE Healthcare 9900 Innovation Drive RP 2156 Wauwatosa WI 53226 USA www gehealthcare com ...