D
IRECTION
5750007-1EN, R
EV
. 1
LOGIQ E10 B
ASIC
S
ERVICE
M
ANUAL
Chapter 7 Diagnostics/Troubleshooting
7 - 29
Network Capture
Network Capture
displays network traffic between the LOGIQ E10 and configured devices. A network
capture outputs two log files: one for main logging with no protected information and another including
protected information. These log files are useful when debugging connectivity issues. Because these
log files can be large, they are only kept for one week.
The information on
Network Capture
is available to all service class licenses.
To access
Network Capture
, select
Utility > Service > Utilities > Network Capture
.
Figure 7-17 Network Capture
This table shows all the elements available on
Network Capture
with descriptions.
Table 7-9
Network Capture
Element
DESCRIPTION
Network Capture Parameters
Maximum Size
Allowed size of the generated log file. Valid value are:
• 256MB
• 512MB
• 1024MB
Devices
DICOM-configured devices for which you want to capture information. If no
additional devices are configured, only All will be available.
Start Network Capture
Select to start the process. This causes the network capture to start, enables the
Stop
button, and updates the
Network Capture Status
pane and changes the
Status to
Running
.
Stop Network Capture
Select to stop the process.
Network Capture Status
Displays information about the status of the network capture. The language setting
for this information is set in Windows and not through the Service desktop or LOGIQ
E10 application software.
Displays the current status of the network capture. Valid values are:
• Not Running
• Running
Содержание 5935000E10
Страница 1: ...LOGIQ E10 Basic Service Manual 5750007 1EN Rev 1 Version R1 2018 GENERAL ELECTRIC COMPANY ...
Страница 2: ......
Страница 3: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL i Important Precautions TRANSLATION POLICY ...
Страница 4: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL ii ...
Страница 5: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL iii ...
Страница 6: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL iv ...
Страница 7: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL v ...
Страница 8: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL vi ...
Страница 9: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL vii ...
Страница 10: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL viii ...
Страница 11: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL ix ...
Страница 12: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL x ...
Страница 13: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL xi ...
Страница 14: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL xii ...
Страница 18: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL xvi This page was intentionally left blank ...
Страница 34: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL xxxii Hardware Kit 9 49 ...
Страница 776: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL Index 4 ...
Страница 777: ......
Страница 778: ... 2018 General Electric Company GE Healthcare 9900 Innovation Drive RP 2156 Wauwatosa WI 53226 USA www gehealthcare com ...