GE
DRAFT
V
OLUSON
™ P8 / V
OLUSON
™ P6
DIRECTION 5723243, R
EVISION
6
DRAFT (M
AY
23, 2018)
B
ASIC
S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-135
a.) Enter the "Serial key" (alphanumeric letters based on system serial number and DRFM
Hardware ID) in the field "Serial Number" (e.g., AEE1193428AA6078123456SU7). Then press
ENTER. it takes about several seconds to get response.
b.) Enter the "Permanent Key" in the field Permanent key, click the
SUBMIT
button and afterwards
SAVE & EXIT
.
Confirm the Message box "BT-key state has been changed." with YES. The system reboots.
4.) Check System Serial number, System Type, Hardware ID and Software options
a.) Press the
F2
key on upper OPIO, select
SYSTEM SETUP
, select
ADMINISTRATION
and then
select the
OPTION
tab.
System Type must be correct depending on Voluson™ P8 / Voluson™ P6.
Software options and serial number must be correct.
5.) Perform:
FRU8-29 Replacement of the DRFM - Functional Tests.
NOTICE
!! NOTICE:
If the previous mentioned steps was done correctly, the system will show the correct SNO and the prior
installed options are enabled in the
OPTIONS
page.
NOTICE
!! NOTICE:
DO NOT
reboot the system before you have changed the serial number successfully!