Chapter 4 - Functional Checks
4-25
4-6-2
Measure Setup
Press the
F2
key on the control panel and then select MEASURE SETUP.
The Measure Setup desktop offers three different pages to check.
NOTICE
!! NOTICE:
Parameters and possible adjustments depend on the selected Application.
To view, add, delete, reorder, edit or when creating a new parameter (in the
MEASURE & CALC
page), it
is very important that all items are chosen correctly and that the relevant item is highlighted.
For further information refer to Chapter 18 in the Basic User Manual of Voluson™ P8 / Voluson™ P6.
Table 4-6
Measurement Setup Checks
Step
Task
Expected Result(s)
1
Measure & Calc:
Check all settings for all applications
setting assigned as desired by the customer
2
Application Parameters:
Check all settings for all applications
setting assigned as desired by the customer
3
Global Parameters: Check all settings
setting assigned as desired by the customer