Voluson™ P8 / Voluson™ P6 

Basic Service Manual

• Voluson™ P8  systems

• Voluson™ P6 systems

Document Number: 5723243

Revision: 6

0459

Содержание 5669999

Страница 1: ...Voluson P8 Voluson P6 Basic Service Manual Voluson P8 systems Voluson P6 systems Document Number 5723243 Revision 6 0459 ...

Страница 2: ......

Страница 3: ...AU CLIENT QU IL INCOMBE DE LE FAIRE TRADUIRE NE PAS TENTER D INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE MANUEL SERVICE N A PAS ÉTÉ CONSULTÉ ET COMPRIS LE NON RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE TECHNICIEN L OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES DANGERS ÉLECTRIQUES MÉCANIQUES OU AUTRES DIESES KUNDENDIENST HANDBUCH EXISTIERT NUR IN ENGLISCHER SPRACHE FALLS EIN FREMDER KU...

Страница 4: ...EGURANÇA DO TÉCNICO OPERADOR OU PACIENTE DEVIDO A CHOQUES ELÉTRICOS MECÂNICOS OU OUTROS IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN INGLESE SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEHC RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA IL CLIENTE È TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE SI PROCEDA ALLA MANUTENZIONE DELL APPARECCHIATURA SOLO DOPO AVER CONSULTATO IL PRESENTE ...

Страница 5: ... ANNARS TUNGUMÁLS EN ENSKU ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG SKILIÐ ÞESSA ÞJÓNUSTUHANDBÓK EF EKKI ER FARIÐ AÐ ÞESSARI VIÐVÖRUN GETUR ÞAÐ VALDIÐ MEIÐSLUM ÞJÓNUSTUVEITANDA STJÓRNANDA EÐA SJÚKLINGS VEGNA RAFLOSTS VÉLRÆNNAR EÐA ANNARRAR HÆTTU VAROITUS FI ΠΡΟΕΙ ΟΠΟΙΗΣΗ EL FIGYELMEZTETÉS HU VIÐVÖRUN IS ...

Страница 6: ...KTRISK MEKANISK ELLER ANDEN FARE FOR TEKNIKEREN OPERATØREN ELLER PATIENTEN DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS VERKRIJGBAAR ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST DAN IS DE KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE ONDERHOUDSHANDLEIDING WERD GERAADPLEEGD EN BEGREPEN IS INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOL...

Страница 7: ...ER ET ANNET SPRÅK ER DET KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT SERVICELEVERANDØREN OPERATØREN ELLER PASIENTEN SKADES PÅ GRUNN AV ELEKTRISK STØT MEKANISKE ELLER ANDRE FARER ĮSPĖJIMAS LT ADVARSEL NO OSTRZEŻENIE PL ATENŢIE RO ...

Страница 8: ...GELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA ÖVERSÄTTNINGSTJÄNSTER FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNINGEN OM DU INTE HAR LÄST OCH FÖRSTÅR DEN HÄR SERVICEHANDBOKEN OM DU INTE TAR HÄNSYN TILL DEN HÄR VARNINGEN KAN DET RESULTERA I SKADOR PÅ SERVICETEKNIKERN OPERATÖREN ELLER PATIENTEN TILL FÖLJD AV ELEKTRISKA STÖTAR MEKANISKA FAROR ELLER ANDRA FAROR ОСТОРОЖНО RU UPOZORNENIE SK VARNING SV DİKKAT T...

Страница 9: ...GE VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 BASIC SERVICE MANUAL vii JA ZH CN KO ...

Страница 10: ...rical work on these products will comply with the requirements of the applicable electrical codes The purchaser of GE equipment shall only utilize qualified personnel i e GE s field engineers personnel of third party service companies with equivalent training or licensed electricians to perform electrical servicing on the equipment OMISSIONS ERRORS If there are any omissions errors or suggestions ...

Страница 11: ... in any form in whole or in part without prior written permission of Voluson P8 Voluson P6 Voluson P8 Voluson P6 may revise this publication from time to time without written notice TRADEMARKS All products and their name brands are trademarks of their respective holders COPYRIGHTS All Material Copyright 2016 18 by General Electric Inc All Rights Reserved ...

Страница 12: ...y 2018 Update contents Pages Revision Pages Revision Pages Revision Title 6 Chapter 3 Setup Instructions pages 1 72 6 Chapter 8 Replacement Procedures pages 1 214 6 Important Precautions pages i x 6 Chapter 4 Functional Checks pages 1 68 6 Chapter 9 Renewal Parts pages 1 26 6 Table of Contents pages i xxvi 6 Chapter 5 Components and Functions Theory pages 1 78 6 Chapter 10 Care Maintenance pages 1...

Страница 13: ...Icons 1 5 Safety Considerations 1 7 Introduction 1 7 Human Safety 1 7 Mechanical Safety 1 7 Electrical Safety 1 8 Safe Practices 1 8 Probes 1 8 Auxiliary Devices Safety 1 8 Labels Locations 1 10 Main Label 1 11 TUV mark 1 12 Multi Caution Label 1 13 Dangerous Procedure Warnings 1 14 Lockout Tagout Requirements For USA Only 1 14 Returning Shipping System Probes and Repair Parts 1 14 Battery Safety ...

Страница 14: ...Power Plug 2 3 EMI Limitations 2 4 Probe Environmental Requirements 2 5 Time and Manpower Requirements 2 5 System Specifications 2 5 Physical Dimensions of Voluson P8 Voluson P6 2 5 Acoustic Noise Output 2 5 Electrical Specifications 2 5 Facility Needs 2 6 Purchaser Responsibilities 2 6 Mandatory Site Requirements 2 7 Site Recommendations 2 7 Recommended Ultrasound Room Layout 2 8 Networking Setup...

Страница 15: ...Printer 3 14 Connection Scheme B W Printer 3 15 Connecting the Color Printer 3 16 Connection Scheme Color Printer 3 17 Connecting the DeskJet Color Printer 3 18 Connection via Bluetooth Adapter 3 18 Connecting the Wireless Network Adapter 3 20 Connecting the Footswitch 3 21 Connecting the USB Flash Memory Stick 3 22 Connecting the external USB Hard disk Handydrive 3 22 General Remarks and Hints wh...

Страница 16: ...r Settings 3 48 UP D897 UP D898MD Printer Settings 3 49 UP D25MD Printer Settings 3 50 Remote Control Selection 3 52 Report Printer Selection 3 53 Cellular Modem Installation 3 54 Overview 3 54 Tools Required 3 54 Parts Required 3 54 Removal procedure 3 55 Removal procedure for Cellular Modem installation 3 55 Installation procedure 3 56 System Configuration 3 62 System Setup 3 62 To invoke the Se...

Страница 17: ...ing to the WLAN 3 76 Disconnecting from the WLAN 3 77 Adding a WLAN Profile 3 77 Refreshing a WLAN Network 3 78 Setting a WLAN Network as Non Preferable 3 78 Removing a WLAN Profile 3 79 Customizing an existing WLAN Profile 3 79 Available WLAN Channels 3 80 Monitoring the WLAN 3 80 WLAN Diagnostic 3 81 Repairing the WLAN 3 82 How to Setup the Cellular Modem 3 83 How to Setup E mail 3 85 How to Set...

Страница 18: ...s and Images 4 11 Save Small Backup Scan Settings 4 12 Load Small Backup Scan Settings 4 13 Preparations 4 13 Load Complete Backup 4 14 Load only parts of the Complete Backup 4 15 Save Full System Configuration Full Backup 4 16 Load Full System Configuration Full Backup 4 18 Delete Full System Configuration Full Backup 4 20 Archiving Images 4 21 Save Image Archive 4 21 Load Image Archive 4 23 Soft...

Страница 19: ...N P8 VOLUSO P6 DIRECTION 5723243 REVISION 6 BASIC SERVICE MANUAL Table of Contents vii Mode Transition 4 42 B Mode 4 42 M Mode 4 42 B CF Mode 4 42 PW Mode 4 42 CW Optional 4 42 3D 4D Optional 4 42 Site Log 4 43 ...

Страница 20: ...on 5 8 3D Rendering 5 9 Block diagram Voluson P8 Voluson P6 5 10 Front End Block diagram 5 11 Back End Block diagram 5 11 OPIO Block diagram 5 11 Data Flow Control Description 5 12 B Mode 5 12 Special B Mode Techniques 5 13 M Mode 5 13 D Mode Pulsed Wave and Continuous Wave Doppler 5 13 D Mode Autotrace draws PC calculated envelope to D Spectrum 5 14 CFM Mode Color Flow Mode 5 15 3D Mode Freezes a...

Страница 21: ...ions 5 20 CW Continuous Wave Doppler 5 20 ECG Module 5 20 Wireless Network Adapter WLAN Wireless Local Area Network 5 20 DPS2 Battery Assy 5 20 Data Location 5 21 FrontEnd Processor 5 22 T DPI Probe Interface board 5 23 DBM64 Beam former board 5 24 DRFM Radio Frequency interface and System controller 5 25 DRFM FPGA 5 26 BackEnd Processor 5 27 Hard Disk Drive 5 28 SOM System On Module 5 28 Congatec...

Страница 22: ...r Up Sequence Description 5 43 Power Down Sequence Description 5 44 Mechanical Concept and Overview 5 44 Functions of ACFE 5 44 ACDC primary DC Power Supply 5 45 Function of ACDC 5 45 DCDC main DC Power Supply 5 45 4D Module 5 45 Service Platform InSite Exc 5 46 Introduction 5 46 Access Security 5 46 Local Access 5 47 Remote Access 5 48 Common Service Desktop CSD 5 49 Internationalization 5 49 Ser...

Страница 23: ...u 5 59 LINUX Voluson Maintenance System 5 61 Overview 5 61 General Operation and Selection in LINUX and BIOS menu 5 61 Access Security 5 61 Before Recovering or Initializing the Hard disk 5 62 Voluson Maintenance System Main Menu 5 64 Media 5 64 Advanced 5 64 About 5 64 Exit Reboot 5 64 Halt 5 64 Media Menu 5 65 Exit 5 65 Restore 5 65 NewDisk 5 69 BIOS Boot Priority Order Setting 5 70 New Hard Dis...

Страница 24: ...GE VOLUSON P8 VOLUSO P6 DIRECTION 5723243 REVISION 6 BASIC SERVICE MANUAL xii Table of Contents CollectLogs 5 79 CheckLnkSpeed 5 79 After Recovering or Initializing the Hard disk 5 80 ...

Страница 25: ...6 BASIC SERVICE MANUAL Table of Contents xiii CHAPTER 6 Service Adjustments Overview 6 1 Purpose of Chapter 6 6 1 Regulatory 6 1 LCD Monitor Adjustment 6 2 Modification of Keyboard Layout 6 3 Setup the Voluson P8 Voluson P6 Keyboard Language Layout 6 3 ...

Страница 26: ...ort Log s and System Data 7 14 Export System Data by pressing the ALT D key 7 14 Export Log s and System Data via Service Page 7 16 Dump file 7 16 Remote Access to Service Platform 7 17 General 7 17 How the Customer enables disables Disruptive Mode and VCO 7 17 How the FE remotely enables disables Disruptive Mode and VCO 7 18 How the FE remotely Connect to Device Total Access Session via RSvP 7 22...

Страница 27: ... 39 ADC Analog Digital Converter 7 39 SignalGen 7 39 TxPulser 7 39 CW Continues Wave 7 40 DPI 7 40 Voltage DPI 7 40 DPS 7 40 Voltage DPS 7 40 ADM Interface 7 40 HV Linearity 7 40 HV Stop 7 40 DC4D Interface 7 41 15W protection 7 41 OPIO Overall Test 7 41 VPD OPIO Test 7 41 Battery 7 41 Battery Interface 7 41 Battery Status 7 41 Battery Discharging to Shipment Level 7 41 HW Test Dialog 7 42 Apertur...

Страница 28: ...System does not boot up 7 50 Noise disturbs the Image 7 51 Trackball Impaired Sensitivity 7 52 Printer Malfunction 7 53 Monitor Troubleshooting 7 54 DVD CD R RW Drive Test 7 55 Network Troubleshooting 7 57 No Connection to the Network at All 7 57 GE remote service connection 7 57 Tech Tips 7 58 Storing SonoView images to Voluson P8 Voluson P6 7 58 Error Messages 7 59 ...

Страница 29: ...13 Overview 8 13 Update Procedure 8 13 Software and Functional Checks after Installation Upgrade Procedure 8 20 Image Settings Only Loading Procedure 8 22 Introduction 8 22 Loading Procedure 8 22 Full Backup Full System Configuration Loading Procedure 8 22 Introduction 8 22 Loading Procedure 8 22 Image Archive Loading Procedure 8 22 Introduction 8 22 Loading Procedure 8 22 Replacement or Activatio...

Страница 30: ... 8 29 Replacement of Front PEDESTAL COVER BT16 8 30 Manpower 8 30 Tools 8 30 Disassembly Procedure 8 30 Installation Procedure 8 31 Replacement of TRAY PEDESTAL CABINET PLASTIC BT16 8 32 Manpower 8 32 Tools 8 32 Disassembly Procedure 8 32 Replacement of REAR CABINET BEZEL ASSY BT16 8 34 Manpower 8 34 Tools 8 34 Disassembly Procedure 8 34 Installation Procedure 8 34 Replacement of ENCODER set 8 35 ...

Страница 31: ...ic lower cover assy 8 56 Manpower 8 56 Tools 8 56 Disassembly procedure 8 56 Installation procedure 8 61 Replacement of OPIO Plastic upper cover assy 8 62 Manpower 8 62 Tools 8 62 Disassembly procedure 8 62 Installation procedure 8 66 Replacement of Monitor 8 67 Manpower 8 67 Tools 8 67 Removal Procedure 8 67 Installation Procedure 8 68 Replacement of the MONITOR ARM 8 69 Manpower 8 69 Tools 8 69 ...

Страница 32: ...ent of the SPEAKER 8 91 Manpower 8 91 Tools 8 91 Preparations 8 91 Removal Procedure 8 91 Installation Procedure 8 91 Remove the PEDESTAL BOTTOM COVER BT16 8 92 Manpower 8 92 Tools 8 92 Preparations 8 92 Removal Procedure 8 92 Installation Procedure 8 92 Replacement of the NEST 8 93 Manpower 8 93 Tools 8 93 Preparations 8 93 Removal Procedure 8 93 Installation Procedure 8 97 Replacement of the AIR...

Страница 33: ...er 8 103 Tools 8 103 Preparations 8 103 Removal Procedure 8 103 Installation Procedure 8 105 Replacement of the Trackball 8 106 Manpower 8 106 Tools 8 106 Removal procedure 8 106 Installation Procedure 8 109 Replacement of the TGC BOARD 8 111 Manpower 8 111 Tools 8 111 Preparations 8 111 Removal Procedure 8 111 Installation Procedure 8 112 Replacement of AN KEYBOARD ASSY 8 113 Manpower 8 113 Tools...

Страница 34: ...oval Procedure 8 123 Installation Procedure 8 123 Replacement of the REAR HANDLE 8 124 Manpower 8 124 Tools 8 124 Preparations 8 124 Removal Procedure 8 124 Installation Procedure 8 126 Replacement of the SPEAKER COVER LEFT RIGHT 8 127 Manpower 8 127 Tools 8 127 Preparations 8 127 Removal Procedure 8 127 Installation Procedure 8 127 Replacement of the DPS DPS with Battery 8 128 Manpower 8 128 Tool...

Страница 35: ...8 139 Installation Procedure 8 140 Replacement of the HDD in DRFM 8 141 Manpower 8 141 Tools 8 141 Preparations 8 141 Removal Procedure 8 141 Installation Procedure 8 141 Replacement of the DBM64 8 142 Manpower 8 142 Tools 8 142 Preparations 8 142 Removal Procedure 8 142 Installation Procedure 8 143 Replacement of the T DPI 8 144 Manpower 8 144 Tools 8 144 Preparations 8 144 Removal Procedure 8 14...

Страница 36: ...ocedure 8 154 Replacement of AN KEY TOP 8 155 Manpower 8 155 Tools 8 155 Preparations 8 155 Key Caps Removal Procedure 8 155 Replacing optional Peripherals 8 156 Manpower 8 156 Tools 8 156 Printer Installation BW and Color 8 157 BW Printer SONY UP D898 Color Printer SONY UP D25MD 8 157 Color Printer Assembly 8 163 HORIZONTAL TV PROBE HOLDER Installation 8 170 VERTICAL TV PROBE HOLDER Installation ...

Страница 37: ... of ECG Kit assy 8 180 Manpower 8 180 Tools 8 180 Preparations 8 180 Removal Procedure 8 180 Installa tion Procedure 8 181 Replacement of Battery Pack Assy 8 184 Manpower 8 184 Tools 8 184 Preparation 8 184 Removal procedure 8 184 Installation Procedure 8 185 Replacement of WIRELESS LAN 8 186 Manpower 8 186 Tools 8 186 Preparation 8 186 Removal Procedure 8 186 Installation Procedure 8 189 Set up W...

Страница 38: ...ers Front Sides Rear 9 4 LCD Monitor 9 6 OPIO 9 7 Nest Box Parts 9 10 Mechanical Parts 9 13 Options 9 15 Optional Peripherals and Accessories 9 16 Printers and options 9 16 System Manuals 9 18 Probes 9 19 2D Probes Curved Array Transducers 9 19 2D Probes Linear Array Transducer 9 20 Real Time 4D Volume Probes 9 21 2D Probes Phased Array Transducers 9 22 Probe Compatibility 9 23 Power Cord 9 24 Bio...

Страница 39: ...re Maintenance 10 7 System Maintenance 10 8 Preliminary Checks 10 8 Functional Checks 10 9 System Checks 10 9 Peripheral Option Checks 10 10 Input Power 10 10 Mains Cable Inspection 10 10 Cleaning 10 10 General Cleaning 10 10 Physical Inspection 10 11 Optional Diagnostic Checks 10 11 Probe Maintenance 10 12 Probe Related Checks 10 12 Basic Probe Care 10 12 Basic Probe Cleaning and or Disinfection ...

Страница 40: ...akage Current Test Probe 10 26 Definition 10 26 Generic Procedure on Leakage Current 10 27 Type CF Patient Auxiliary current ECG Lead to Probe 10 27 Type BF Patient Leakage current Mains to Probe 10 28 Type CF Patient Leakage current Mains to ECG lead 10 28 Total Patient leakage current Type CF probe to ground Only for Europe 10 29 Definition 10 29 Generic Procedure for measurement 10 29 Total Pat...

Страница 41: ...Checks Contains functional checks that are recommended as part of the installation or as required during servicing and periodic maintenance 5 Chapter 5 Components and Functions Theory Contains block diagrams and functional explanations of the electronics 6 Chapter 6 Service Adjustments Contains instructions on how to make available adjustments 7 Chapter 7 Diagnostics Troubleshooting Provides proce...

Страница 42: ... Personnel installation maintenance etc Hospital s Service Personnel Contractors Some parts of Chapter 2 Pre Installation 1 1 4 Models Covered by this Manual Table 1 2 Voluson P8 Voluson P6 Model Designations Description Part Number BT16 VolusonTM P6 BT16 Console 5670023 VolusonTM P6 BT16 Console High 5670024 VolusonTM P8 BT16 Console 5669999 VolusonTM P8 BT16 Console High 5670021 ...

Страница 43: ...und on the equipment and in the service information The different levels of concern are identified by a flag word that precedes the precautionary message Known or potential hazards are labeled in one of following ways NOTE Notes provide important information about an item or a procedure Information contained in a NOTE can often save you time or effort DANGER DANGER INDICATES THE PRESENCE OF A HAZA...

Страница 44: ... symbols may be used to make you aware of specific types of hazards that could cause harm Even if a symbol isn t used in this manual it is included for your reference Other hazard icons make you aware of specific procedures that should be followed Table 1 3 Standard Hazard Icons ELECTRICAL MECHANICAL RADIATION LASER HEAT PINCH Table 1 4 Standard Icons Indicating a Special Procedure be Used AVOID S...

Страница 45: ...harmful ingress of water for foot switch Probes and Footswitch CAUTION This unit weighs Special care must be used to avoid This precaution is intended to prevent injury that may result if one person attempt to move the unit considerable distances or on an incline due to the weight of the unit Used in the Service and User Manual which should be adjacent to equipment at all times for quick reference...

Страница 46: ... representative of the manufacturer for information concerning the decommissioning of your equipment Rear side of the unit On the plug of each probe These symbols indicate that at least one of the six hazardous substances of the China RoHS Labelling Standard is above the RoHS limitation The number inside the circle is referred to as the Environmental Friendly Use Period EFUP It indicates the numbe...

Страница 47: ...ed roads excessive speeds and erratic stops or starts CAUTION CAUTION The Voluson P8 Voluson P6 weighs approximately 70 kg depending on installed peripherals 154 lbs or more when ready for use Care must be used when moving it or replacing its parts Failure to follow the precautions listed could result in injury uncontrolled motion and costly damage ALWAYS Use the handle to move the system Be sure ...

Страница 48: ...replacement Avoid pulling pinching or kinking the probe cable since a damaged cable may compromise the electrical safety of the probe To avoid the risk of a probe accidentally falling do not allow the probe cables to become entangled or to be caught in the machine s wheels Never immerse the probe connector or adapter into any liquid NOTE For detailed information on handling probes refer to the Vol...

Страница 49: ...on procedures see Chapter 3 Connection of Auxiliary Devices on page 3 12 NOTICE NOTICE The system integrator any person connecting the medical device to other devices is responsible that the connections are safe If in doubt consult the technical service department or your local representative CAUTION CAUTION The leakage current of the entire system including any all auxiliary equipment must not ex...

Страница 50: ...1 10 Section 1 3 Safety Considerations 1 3 6 Labels Locations The Voluson P8 Voluson P6 comes equipped with product labels and icons These labels and icons represent pertinent information regarding the operation of the unit Figure 1 1 Monitor caution label on rear of the Voluson P8 Voluson P6 system ...

Страница 51: ...ear of the Voluson P8 Voluson P6 system Figure 1 2 Main Label located on left rear of Voluson P8 Voluson P6 1 Manufacturer 7 Product name 2 Weight 8 Console Part No 3 Classification 9 Brand and model name 4 System Voltage Setting 10 Manufacturing date 5 Frequency 11 System Serial Number 6 Power Consumption nominal 12 Bar Code 1 2 3 4 10 5 6 7 8 9 11 12 5 6 7 8 9 1 2 3 4 10 11 12 ...

Страница 52: ...OLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 1 12 Section 1 3 Safety Considerations 1 3 6 2 TUV mark TUV mark is located on the rear of the system Figure 1 3 TUV mark ...

Страница 53: ...n Label The Multi caution label is located on the rear of the Voluson P8 Voluson P6 System Figure 1 4 Multi Caution Label 1 Symbol Caution 2 Symbol Follow instructions for use 3 Symbol Do not use the following devices near this system 4 Symbol Pushing prohibited 5 China RoHS mark 6 e IFU label 7 CE mark 8 Rx Only mark 9 WEEE mark 1 2 3 4 5 6 7 8 9 ...

Страница 54: ...or shipped from a clinic or site for example body coils or and ultrasound probe The purpose of the regulation is to protect employees in the transportation industry as well as the people who will receive or open this package DANGER DANGER DANGEROUS VOLTAGES CAPABLE OF CAUSING DEATH ARE PRESENT IN THIS EQUIPMENT USE EXTREME CAUTION WHEN HANDLING TESTING AND ADJUSTING WARNING WARNING WARNING EXPLOSI...

Страница 55: ...and other heat sources DO NOT charge the battery near a heat source such as a fire or heater DO NOT leave the battery in direct sunlight DO NOT pierce the battery with a sharp object hit it or step on it DO NOT use a damaged battery DO NOT apply solder to a battery DO NOT connect the battery to an electrical power outlet DO NOT immerse the battery in water or allow it to get wet DO NOT place the b...

Страница 56: ...son P6 1 4 2 Compliance The Voluson P8 Voluson P6 unit conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge radiated and conducted RF fields magnetic fields and power line transient requirements as mentioned in IEC 60601 1 2 NOTE For CE Compliance it is critical that all covers screws shielding gaskets mesh clamps are in good condition inst...

Страница 57: ...does not work as indicated in this service manual or in the Basic User Manual or if you require additional assistance please contact the local distributor or appropriate support resource as listed below NOTE Prepare vital system information see Section 7 2 on page 7 2 before you call System Type System Serial number also visible on label on back of the system Application Software version Backup ve...

Страница 58: ...on 1 5 Customer Assistance additional information about installed software 1 5 2 System Manufacturer Figure 1 5 Phone Number for Customer Assistance Table 1 6 System Manufacturer Manufacturer FAX Number GE Ultrasound Korea Ltd 9 Sunhwan ro 214beon gil Jungwon gu Seongnam si Gyeonggi do Korea 82 0 31 740 6436 ...

Страница 59: ...of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a Voluson P8 Voluson P6 ultrasound unit Included are descriptions of the facility and electrical needs to be met by the purchaser Table 2 1 Contents in Chapter Voluson P8 Voluson P6 Section Description Page Number 2 1 Overview 2 1 2 2 General Console Requirements 2 2 2 3 Facility Needs 2 6 ...

Страница 60: ...ble interference 2 2 2 Electrical Requirements NOTE Voluson P8 Voluson P6 requires a dedicated power and ground for the proper operation of its Ultrasound equipment This dedicated power shall originate at the last distribution panel before the system The dedicated line shall consist of one phase a neutral not shared with any other circuit and a full size Ground wire from the distribution panel to ...

Страница 61: ...upport this unit must also be present within 1 m 3 2 ft of the unit Electrical installation must meet all current local state and national electrical codes 2 2 2 5 Main Power Plug The Voluson P8 Voluson P6 is supplied with a main power plug as standard In the event that the unit arrives without a power plug or with the wrong plug contact your GE dealer When necessary the installation engineer will...

Страница 62: ...ls Ground the unit Poor grounding is the most likely reason a unit will have noisy images Check grounding of the power cord and power outlet Replace and or reassemble all screws RF gaskets covers and cores After you finish repairing or updating the system replace all covers and tighten all screws Any cable with an external connection requires a magnet wrap at each end Install the shield over the f...

Страница 63: ... on control console and monitor positioning For more details refer to Chapter 2 System Specifications on page 2 5 2 2 6 2 Acoustic Noise Output max 55 dB A 2 2 6 3 Electrical Specifications Please refer to Section 2 2 2 1 Voluson P8 Voluson P6 Power Requirements on page 2 3 NOTICE NOTICE SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF 10 TO 50 o C WHEN EXPOSED TO LARGE TEMPE...

Страница 64: ... The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment The desire to use a non listed or customer provided product or to place an approved product further from the system than the interface kit allows presents challenges to the installation team To avoid delays during installation such variances should be made known to the individuals ...

Страница 65: ...oluson P8 Voluson P6 has four outlets inside the unit One is for the monitor and three for on board peripherals In case of network option An active network outlet in the vicinity of the ultrasound unit A network cable of appropriate length regular Pin to Pin network cable An IT administrator who will assist in configuring the unit to work with your local network A fixed IP address is required Refe...

Страница 66: ...mended Ultrasound Room Layout Figure 2 1 below shows a floor plan illustrating the recommended layout of the Ultrasound Room and depicting the minimal room layout requirements Figure 2 1 Recommended Floor Plan 4 3m x 5 2m 14ft x 17ft 35 5 IN 92 CM Dedicated Power Outlets Hospital Network Cabinet for Software and Manuals optional VOLUSON P6 P8 ...

Страница 67: ... up the on board monitor and peripherals enabling viewing to be done while scanning continues With DICOM images can be archived stored and retrieved faster easier and at a lower cost 2 3 4 4 DICOM Option Pre installation Requirements To configure the Voluson P8 Voluson P6 ultrasound unit to work with other network connections the network administrator must provide some necessary information Use th...

Страница 68: ...GE RAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 2 10 Section 2 3 Facility Needs This page was intentionally left blank ...

Страница 69: ...at all the required information is available prior to setup the system Table 3 1 Contents in Chapter Voluson P8 Voluson P6 Section Description Page Number 3 1 Overview 3 1 3 2 Set Up Reminders 3 2 3 3 Receiving and Unpacking the Equipment 3 6 3 4 Preparing for Set Up 3 10 3 5 Connection of Auxiliary Devices 3 12 3 6 Completing the Set Up 3 25 3 7 Printer Installation 3 32 3 8 Cellular Modem Instal...

Страница 70: ...nce the site has been prepared the average installation time required is shown in Table 3 2 below Table 3 2 Average Installation Time Description Average Installation Time Comments Unpacking the ultrasound unit 0 5 hours Installing the ultrasound unit options printers 0 5 to 1 5 hours depends on required configuration DICOM Option connectivity 0 5 to 1 5 hours depends on configuration amount ...

Страница 71: ...old or hot It requires one hour for each 2 5 C increment if it s temperature is below 10 C or above 40 C Figure 3 1 moving or lifting the system CAUTION CAUTION Equipment damage possibility Turning the system on without acclimation after arriving at site may cause the system to be damaged Table 3 3 Acclimation Time C 60 55 50 45 40 35 30 25 20 15 10 5 0 5 10 15 20 25 30 35 40 F 140 131 122 113 104...

Страница 72: ...Y 23 2018 BASIC SERVICE MANUAL 3 4 Section 3 2 Set Up Reminders 3 2 2 3 Brake Pedal Operation WARNING WARNING WARNING REMEMBER If the front wheel brakes are engaged for transportation release brake pedals brakes on front wheels under the foot rest to disengage the lock ...

Страница 73: ...e to two prong adapter This defeats safety grounding CAUTION CAUTION DO NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is present CAUTION CAUTION DO NOT use a 20 Amp to 15 Amp adapter on the 120 Vac unit s power cord This unit requires a dedicated 16 A circuit CAUTION CAUTION DO NOT operate this unit unless all board covers and frame panels are securely in plac...

Страница 74: ...o people available to unpack the Voluson P8 Voluson P6 Attempts to move the unit considerable distances or on an incline by one person alone could result in personal injury and or damage to the system Table 3 4 Shipping Carton Dimensions and Weight Description Height Width Depth Weight Voluson P8 Voluson P6 incl accessories 1490 mm 870 mm 650 mm 100kg including the system Figure 3 3 Shock Indicato...

Страница 75: ...iving and Unpacking the Equipment cont Figure 3 4 envelope at front panel of the crate Table 3 5 Unpacking Procedure Step Task 1 Unbend the Packing box by cutting Check whether delivery is complete according to packing list and check visual damage The envelope with delivery address packing list and invoice is located on the front panel of the crate ...

Страница 76: ... MAY 23 2018 BASIC SERVICE MANUAL 3 8 Section 3 3 Receiving and Unpacking the Equipment 2 Remove the top cover and top plate 3 Remove the three Plastic Joints from the outer sleeve 4 Remove the outer sleeve and the monitor pad Table 3 5 Unpacking Procedure Step Task ...

Страница 77: ...p Instructions 3 9 5 Cut the tape to unfold the package and remove the antistatic vinyl cover 6 Disengage the brakes grasp the system at the handle then slowly move unit down the ramp Note Packing crate and material should be stored for future use Table 3 5 Unpacking Procedure Step Task Rear panel is used for ramp ...

Страница 78: ...y that the system is switched OFF and unplugged Clean the console and control panel 3 Control Console Physically inspect the control console for missing or damaged items After switching on the system verify the proper illumination of all the control panel buttons 4 Probes Check all probes for wear and tear on the lens cable and connector Look for bent or damaged pins on the connector and in the co...

Страница 79: ...f wood plaster or concrete which help prevent EMI Shield the system when operating it in the vicinity of radio broadcast equipment if necessary Do not operate mobile phones or other EMI emitting devices in the ultrasound room Verify that all EMI rules listed in the following table are followed The Voluson P8 Voluson P6 ultrasound unit is approved for use in hospitals clinics and other environmenta...

Страница 80: ...the Black White Printer 3 14 3 5 3 Connecting the Color Printer 3 16 3 5 4 Connecting the DeskJet Color Printer 3 18 3 5 5 Connecting the Wireless Network Adapter 3 20 3 5 6 Connecting the Footswitch 3 21 3 5 7 Connecting the USB Flash Memory Stick 3 22 3 5 8 Connecting the external USB Hard disk Handydrive 3 22 3 5 9 General Remarks and Hints when using external USB Devices 3 23 WARNING WARNING W...

Страница 81: ...DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 3 Setup Instructions 3 13 3 5 1 Connecting the LCD Monitor NOTE The LCD monitor comes pre installed with the system Figure 3 5 Connection Scheme LCD Monitor ...

Страница 82: ... the cables are connected press the Power ON switch on the Black White printer 4 Power ON Boot up the Voluson P8 Voluson P6 system as described in Section 3 6 2 on page 3 25 All software drivers are pre installed for the designated Black White printer only 5 After physical connection to the Voluson P8 Voluson P6 system assign the printer to a remote key P1 P2 and P3 as described in Section 3 7 8 R...

Страница 83: ...E DRAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 3 Setup Instructions 3 15 3 5 2 1 Connection Scheme B W Printer Figure 3 6 B W Printer connection ...

Страница 84: ...e cables are connected press the Power ON switch on the Color printer 4 Power ON Boot up the Voluson P8 Voluson P6 system as described in Section 3 6 2 on page 3 25 All software drivers are pre installed for the designated Color printer only 5 After physical connection to the Voluson P8 Voluson P6 system assign the printer to a remote key P1 P2 and P3 as described in Section 3 7 8 Remote Control S...

Страница 85: ...DRAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 3 Setup Instructions 3 17 3 5 3 1 Connection Scheme Color Printer Figure 3 7 Color Printer connection ...

Страница 86: ...nection via Bluetooth Adapter The DeskJet Color Printer can be connected to an external non isolated power source The bluetooth adapter should be directly connected to the indicated USB 3 on rear panel Figure 3 8 HP Officejet H470 DeskJet Printer connection via Bluetooth Adapter Power Cable connected to the rear side Bluetooth Printer Adapter of the printer USB Bluetooth Adapter connected to USB p...

Страница 87: ...sembly has to be located outside of the patient environment according to IEC 60601 1 UL 60601 1 CAUTION CAUTION The printer being used may not be a medical device The Bluetooth Printer Set and the Power Supply of the Bluetooth Printer Adapter is also not a medical device The equipment meets the requirements of the EN 60950 Standard NOTICE NOTICE Please use the proper Bluetooth Printer Connection s...

Страница 88: ...WLAN Adapter is always the same 1 Turn ON the power of the system and wait till the system has booted 2 Plug the Wireless Network adapter into an accessible USB port of the Voluson P8 Voluson P6 All software drivers are pre installed for the designated Wireless Network adapter only NOTE After physical connection of the WLAN adapter to the Voluson P8 Voluson P6 system follow the procedure described...

Страница 89: ...switch should be directly connected to any accessible USB port on the Voluson P8 Voluson P6 e g on rear of the system NOTE Connection of the Footswitch is always the same After physical connection adjust the Footswitch Left Right as described in Section 3 9 1 7 How to adjust function of the Footswitch Left Middle Right on page 3 67 Figure 3 10 Footswitch connection scheme ...

Страница 90: ... once the system is powered ON or after shutdown The Voluson P8 Voluson P6 Windows detects the device and automatically installs a driver During this process several dialogs may pop up starting with the Found New Hardware dialog NOTE Memory drives or sticks may be sensitive to EMC interference This may affect system performance and or image quality NOTICE NOTICE Before connecting an USB device ple...

Страница 91: ... By pressing the DRIVES in the Sytem Setup a dialog window see Figure 3 11 below is displayed The Connect USB and Network Drives window shows all USB and Network drives connected to the system Using this dialog the USB devices can be stopped before they are physically disconnected WARNING WARNING WARNING Do not connect or disconnect any external USB devices to or from the system while scanning a p...

Страница 92: ...018 BASIC SERVICE MANUAL 3 24 Section 3 5 Connection of Auxiliary Devices To stop the external device select it and then click the STOP DEVICE button Confirm the Stop Device dialog with OK the Connect USB and Network Drives window Figure 3 12 Device can now be safely removed ...

Страница 93: ...hat it is intact and of hospital grade Products equipped with a power source wall outlet plug should be connected to the fixed power socket that has a protective grounding conductor Never use an adapter or converter to connect with a power source plug for example a three prong to two prong converter WARNING WARNING WARNING The unit s power must be supplied from a separate properly rated outlet to ...

Страница 94: ...IRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 3 26 Section 3 6 Completing the Set Up 3 Switch ON the Circuit at the rear of the system Figure 3 13 Circuit and main power cable at rear of system Circuit Main Power ...

Страница 95: ... the control panel is pressed the System including the Back end Processor starts and the operating system is loaded which then leads the application software to activate the Ultrasound unit The system automatically performs an initialization sequence which includes the following Loading the operating system Running a quick diagnostic check of the system Detecting connected probes As soon as the so...

Страница 96: ...onitor 2 Afterwards the Boot Screen is displayed Voluson is highlighted Figure 3 15 below 3 Back End Processor is turned ON and starts to load the software 4 The Start Screen Voluson P8 Voluson P6 is displayed on the monitor 5 Start up progress bars indicating software loading procedures are displayed on the monitor as shown in Figure 3 16 below 6 The software initiates and sets up the Front End e...

Страница 97: ...4 on the control panel Following dialog appears 3 Select the SHUTDOWN button The system performs an automatic full shutdown sequence NOTE Full shut down is also performed when pressing the ON OFF button on the control panel twice 4 Switch OFF the Circuit at the rear of the system NOTICE NOTICE After turning off a system wait at least 10 seconds before turning it on again The system may not be able...

Страница 98: ...le from the system or unplug it from the AC wall outlet socket 6 Press on the brakes to block the front caster wheels 7 Disconnect probes Turn the probe locking handle counterclockwise and then pull the connector straight out of the probe port WARNING WARNING WARNING Disconnection of the Main Power Cable is necessary For Example When repairing the system Figure 3 18 Circuit and main power at rear ...

Страница 99: ... 19 3 Insert the connector on the receptacle and move lever to the right position to lock probe 4 Carefully position the probe cord so that it is free to move and is not resting on the floor NOTE Prior to connecting or disconnecting a probe freeze the image It is not necessary to turn OFF power to connect or disconnect a transducer Figure 3 19 Transducer Connection CAUTION CAUTION Do not bend the ...

Страница 100: ...directly via an USB cable use the AC mains power outlet privided by the Voluson P series system auxiliary output This ensures medical grade separation from AC mains 3 7 1 Installing the Digital Black White Printer 1 Power off Shutdown the system as described in Section 3 6 3 on page 3 29 2 Connect the printer as described on page 3 14 3 Verify correct printer settings Section 3 7 7 Adjustment of P...

Страница 101: ...Section 3 6 2 on page 3 25 4 Press the Power On button on the printer 5 Install ink tanks and load paper 6 Wait until the printer displays Ready 7 By means of the Printer Controls Left Right and OK set the TCP IP address to Automatic 8 Press the Back button until the Printer displays Ready again 9 By means of the Printer Controls Left Right and OK print out the Config Report Figure 3 20 printer co...

Страница 102: ...t has to be entered during Printer Installation Procedure 11 Download the appropreate printer driver from GE folders http libraries ge com foldersIndex do entity_id 52794606101 sid 101 SF 1 and extract files to CD or USB disk 12 Connect the USB stick to the Voluson P series system or insert the CD into the drive 13 Press the Utilities key on the control console 14 In the Utilities menu press the S...

Страница 103: ...yed message carefully and click Yes if you have skills to do this 17 In the displayed window click on Add a network wireless or Bluetooth printer 18 In the displayed window click on The printer that I want isn t listed and then clik Next 19 Check mark Connect to this printer 20 Enter http flowed by the previously print out printer IPv4 Address e g 3 429 68 25 see Section Figure 3 23 on page 3 34 a...

Страница 104: ...7 Printer Installation NOTE This example show fictional numbers 21 Click on Have Disk and browse for E where E stands for the USB stick resp the CD drive search for the printer driver folder and open the file hpcu175u inf Finally click OK Figure 3 26 enter printers IP Address Figure 3 27 search for printer driver ...

Страница 105: ...Instructions 3 37 22 Select HP Universal Printing PCL6 and then Click OK to start the installtion 23 Confirm the following screeen with Next 24 Remove the check mark at Set as the default printer and then click on Print a test page NOTE Please DO NOT yet click on Finish Figure 3 28 Figure 3 29 installation succesful ...

Страница 106: ...25 In the following window check mark No do not send this info to HP and do not ask me again and then click OK 26 In the following screen click on Settings 27 Remove all check marks to avoid further setup notifications and then click OK Figure 3 30 No do not send info the HP Figure 3 31 click Settings Figure 3 32 remove all check marks ...

Страница 107: ...ee Section 3 7 8 Remote Control Selection on page 3 52 6 Assign the printer as Report Printer see 3 7 5 Installing the DeskJet Color Printer via Bluetooth connection NOTE Installation of the DeskJet Color Printer has to be performed in sequential order Content in this section 3 7 5 1 Installing the Bluetooth Adapter and Software 1 Turn ON the system and wait till the system has booted 2 Connect th...

Страница 108: ...ee it in the Device Manager see In this case skip installation of the Bluetooth Adapter close all open windows and proceed with Section 3 7 5 2 on page 3 43 8 Using the Arrow keys on the keyboard select System Tools Device Manager Other Devices Select the designated Bluetooth device click the upper trackball key and select Properties 9 In the displayed window click on Update Driver Figure 3 34 sel...

Страница 109: ...apter 3 Setup Instructions 3 41 10 Select Browse my computer for driver software 11 Click on Browse and navigate to F where F stands for the DVD drive select Win7 and click OK 12 Verify the path and click Next to start installation Figure 3 36 Browse for driver Figure 3 37 Win7 Figure 3 38 verify path ...

Страница 110: ... Printer Installation 13 When installation is finished click on Close 14 Now you should see the installed Bluetooth device in the Device Manager Click the upper trackball key and verify its device status This device is working properly Figure 3 39 installation finished Figure 3 40 Bluetooth device is up and running ...

Страница 111: ...the printer software driver 1 Close all open windows 2 Press CTRL ESC keys simultaneously and open the Windows Start menu 3 Using the Arrow keys on the keyboard select Voluson Application 2 to start the Ultrasound Application 4 Press the Utilities key on the control console 5 In the Utilities menu press the Setup button to invoke the setup desktop on the screen 6 On the right side of the screen se...

Страница 112: ... MANUAL 3 44 Section 3 7 Printer Installation 8 Select Add a network wireless or Bluetooth printer 9 Select your printer and click on Next 10 You have successfully added the printer Click Next Figure 3 43 add Bluetooth printer Figure 3 44 select printer Figure 3 45 successfully added printer ...

Страница 113: ...ter 3 Setup Instructions 3 45 11 Check box desired printer sharing option and then click Next 12 Check box Set as the default printer and click Print a test page 13 When test page is printed click on Close and Finish Figure 3 46 printer sharing Figure 3 47 print test page Figure 3 48 finish installation ...

Страница 114: ...s off unexpectedly after being idle for a period of time Issue A print copy or scan job fails becuase the printer is turned off You might have completed a job a few hours ago and you are sure that no one turned off the printer Solution Disable Auto off when Bluetooth is turned on Procedure 1 Press the Power 3 button to turn on the printer 2 Simultaneously press and bold the Bluetooth 1 and Cancel ...

Страница 115: ... tab 4 Click the ADD PRINTER button Please read the displayed message carefully and click YES if you have skills to do this 5 Click the NEXT button to start the Add Printer Wizard 6 After installation close all open windows select Save Exit and restart the system turn off and on the system 7 Verify correct printer settings see 8 Assign the printer to the remote keys P1 P2 and or P3 see Section 3 7...

Страница 116: ...screen 3 On the right side of the screen select CONNECTIVITY and then click on the PERIPHERALS tab 4 Select the desired printer from the Printer Settings pull down menu and click the EDIT button To adjust the UP D898 printer see Section 3 7 7 1 UP D897 UP D898MD Printer Settings To adjust the UP D25MD Printer see Section 3 7 7 2 UP D25MD Printer Settings Figure 3 50 Select the desired printer WARN...

Страница 117: ...lect the LAYOUT page and select Paper 960x1280 Orientation Portrait Interpolation Method Bilinear 3 Select the DENSITY ADJUST page and select Gamma TONE2 Advanced 0 Dark 0 Light 0 Sharpness 2 4 Save the adjusted printer settings click APPLY and then OK Close window and exit System Setup with SAVE EXIT 5 Assign the Printer to the remote keys P1 P2 and or P3 with the save button see Section 3 7 8 Re...

Страница 118: ...has to be used Otherwise you will get an error message at printing 3 Select the Graphics page From the Color Adjust pull down menu selet 4 Color Balance Cyan 0 Magenta 0 Yellow 0 5 Lightness Sharpness 7 Dark 0 Gamma 6 Gamma Curve Curve 1 NOTICE The appropriate printer settings can be automatically loaded by means of a Registry file Therefore Download the SonyUp D25MD reg from GE folders http libra...

Страница 119: ...e Color Correction 7 Gray Balance X 50 Y 50 8 Adjustment by Color Color region Magenta Red 9 Save the adjusted printer settings with Apply and OK Close window and exit System Setup 10 Assign the printer to the remote keys P1 P2 and or P3 see Section 3 7 8 1 on page 3 53 Figure 3 53 Graphic page lightness Color Correction Figure 3 54 Graphic page Gray Balance Adjustment by Color ...

Страница 120: ...ttons Select the desired remote control button B Select Printer Check mark and select the desired Printer for the remote control button C Recorder Control Check mark this item to use the selected P key for Recorder Control D Check mark Use P key to send Email and adjust Settings NOTE Optionally the Printer Remote Control can be done by the Foot switch refer to Figure 3 49 System Setup Connectivity...

Страница 121: ...ructions 3 53 3 7 8 1 Report Printer Selection 1 Click on the PERIPHERALS tab refer to Figure 3 49 System Setup Connectivity PERIPHERALS page on page 3 47 2 Select the desired Report Printer from the pull down menu NOTICE NOTICE The selected Report Printer is usually used for printing reports and images from the Archive ...

Страница 122: ...on Section 3 8 Cellular Modem Installation 3 8 1 Overview This section describes the installation procedure of Cellular Modem Kit 3 8 2 Tools Required Standard Field Service Tool Kit 3 8 3 Parts Required Table 3 1 H48691WL Cellular Modem option kit No Part No Part Name Q ty 1 KTZ304025 Cellular Modem 1 Figure 3 56 Cellular Modem option kit ...

Страница 123: ... 4 Removal procedure 3 8 4 1 Removal procedure for Cellular Modem installation 1 Remove the power cord and REAR CABINET BEXEL ASSY BT16 2 Unscrew 7 screws and disassemble REAR EMI COVER Figure 3 57 Removing the POWER CORD and REAR CABINET BEZEL ASSY BT16 Figure 3 58 Unscrewing 7 screws POWER CORD FIX Bracket REAR CABINET BEZEL ASSY BT16 ...

Страница 124: ...n 3 8 Cellular Modem Installation 3 8 5 Installation procedure 1 To refer to below picture Unplug power cable for cellular modem and cut cable tie 2 Slide up the upper cover of the Cellular Modem Pos 1 to get access to the SIM card holder Figure 3 59 Unpluging power cable Figure 3 60 Sliding up the upper cover ...

Страница 125: ...ow fix the SIM card with the mouted clips and put back on the upper cover 4 Combine loop fastener and hook fastener and detach sticker from hook fastener like ebelow picture make two pair 5 Attach fastener pair to cellular modem like below picture Figure 3 61 Putting in the SIM card like figure above Figure 3 62 Combining loop fastener and hook fastener Figure 3 63 Attaching fastener pair ...

Страница 126: ...Cellular Modem Installation 6 Refer to cellular modem guide and attach Cellular modem to REAR CABINET BEZEL ASSY BT16 like below 7 Plug power cable and USB cable to port 3 for cellular modem like below picture Figure 3 64 attaching Cellur modem to REAR CABINET BEZEL ASSY BT16 Figure 3 65 Pluging Power Cable and USB ...

Страница 127: ...MANUAL Chapter 3 Setup Instructions 3 59 8 Refer to below picture insert two cables USB and power for cellular modem through hole of REAR EMI COVER 9 Assemble REAR EMI COVER and screw 7 screws Figure 3 66 Inserting two cables through hole of REAR EMI COVER Figure 3 67 Assembling REAR EMI COVER ...

Страница 128: ...s from REAR EMI COVER First pull out power cable within the limits of the possible Therefore pull out USB cable like below picture To align cable edge pull out USB cable as much as power cable 11 Connect Power Cable and USB Cable to Cellular modem Figure 3 68 Pulling out 2 cabels from REAR EMI COVER Figure 3 69 Connecting Power Cable and USB Cable to Celular modem ...

Страница 129: ...REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 3 Setup Instructions 3 61 12 Close REAR CABINET BEZEL ASSY BT16 and connect POWER COARD FIX BRACKET Figure 3 70 Closing REAR CORD FIX BRACKET POWER CORD FIX BRACKET REAR CABINET BEZEL ASSY BT16 ...

Страница 130: ...r settings etc Device Setup to set up all DICOM Archive and Network configuration nodes Button Configuration to adjust assignment of Remote keys P1 P2 and P3 e g Printer selection Drives USB and Network drives stop devices map network drive Erase CD System Configuration Save Load Settings only Save Load Delete Full System Configuration Image Archive to save or load Image Archive Presets to save Us...

Страница 131: ...TEM SETUP to activate the setup desktop screen 3 Select the corresponding major group from the left side of the screen and then click the desired tab Trackball mouse position positions the pointing device arrow on the desktop left trackball key left mouse button sets fixates markers and activates pages buttons etc marked by the pointing device upper trackball key right mouse button no function in ...

Страница 132: ...ETTINGS tab 3 Click the CHANGE DATE AND TIME button to activate a sub dialog window to enter date time and time zone 4 Select the DATE FORMAT button only one can be active 5 Select the Time Format only one can be active 6 Close the Service page with Save Exit Figure 3 71 System Setup General Settings GENERALSETTINGS page 1 Date and Time 6 Display options 2 Change date and time 7 Diverse options 3 ...

Страница 133: ...he text box to enter a new Clinic Name with the keyboard 4 Close the Service page with Save Exit The clinic name will be copied into the Hospital ID in the information header 3 9 1 4 How to change Language and or EUM Language 1 Invoke System Setup as described in Section 3 9 1 1 on page 3 63 2 On the left side of the screen select GENERAL SETTINGS and then click on the GENERAL SETTINGS tab 3 Selec...

Страница 134: ... letter characters 0 9 or c Retype new Password and then click SAVE EXIT to save new screen lock password 4 Click SAVE EXIT to save Settings and exit System Setup NOTE If screen is locked you have to enter the password to get full system control If password is unknown click EMERGENCY This enables standard but limited operation 5 Click SAVE EXIT to save Settings and exit System Setup A Service Dong...

Страница 135: ... Middle Right 1 Invoke System Setup as described in Section 3 8 1 1 on page 3 55 2 On the left side of the screen select CONNECTIVITY and then click on the PERIPHERALS tab 3 Select desired function of the Footswitch Left Middle and Right 4 Click SAVE EXIT to save Settings and exit System Setup 3 9 1 8 How to change the Keyboard Layout see Chapter 6 Modification of Keyboard Layout on page 6 3 3 9 1...

Страница 136: ...cted to these outlets must not exceed 400VA Table 3 2 Approved Peripherals Device Manufacturer Model Connection Comment Digital B W printer SONY UP D898 USB Port Digital Color Printer SONY UP D25MD USB Port UVC S200 Kretz Cellular Modem Kretz Isolation Transformer GEUK Isolation USB connector GEUK Isolation 1Gb Ethernet connection GEUK ECG Module Norav Marlin ECGUSB1D USB Port Footswitch Whanam El...

Страница 137: ... Description 1 HDMI Out Table 3 4 HDMI Connector for external Monitor 2 NETWORK Table 3 6 DICOM input output twisted pair RJ 45 10 100 megabit s 3 USB Table 3 7 USB 2 0 port 4 RGB Out Table 3 5 RGB Connector for external Monitor 5 Audio Out Table 3 8 Audio out for external speaker Table 3 4 HDMI out Connector 23 Pin Pin No Output Signal Description 1 12 TMDS clock data Shield TMDS clock data Shiel...

Страница 138: ...DC_DATA DATA 15 DDC_CLK CLK 5 6 7 8 10 GND Ground 4 9 11 NC No Connection Table 3 6 Network Connector RJ45 Modular 8 Pin Pin No Output Signal Description 1 ETHER TD Ethernet RD 2 ETHER TD Ethernet RD 3 ETHER RD Ethernet TD 6 ETHER RD Ethernet TD Others NC No connection Table 3 7 USB 2 0 Connectors Pin No Output Signal Description 1 VCC USB Power Supply 2 Data USB Data 3 Data USB Data 4 GND USB Pow...

Страница 139: ...n page 9 19 for part numbers to be used when ordering new or replacement service probes Section 3 11 Software Option Configuration For description refer to Section 3 9 1 System Setup on page 3 62 NOTICE NOTICE More detailed information pertaining System Setup and Measure Setup adjustments is found in the Voluson P8 Voluson P6 Basic User Manual which is available in different languages ...

Страница 140: ...ove data because no network is available see 3 12 1 4 on page 3 74 Voluson P8 Voluson P6 and DICOM server in a network see Section 3 12 1 5 on page 3 74 3 12 1 1 The Dataflow Concept Communication between the Voluson P8 Voluson P6 ultrasound unit and other information providers on the network takes the form of data flows Each dataflow defines the transfer of patient information from either an inpu...

Страница 141: ... archive The local database is used for patient archiving Images are stored to internal hard drive The local database is used for patient archiving Afterwards images are stored to a DVD CD or external USB device etc Search in the DICOM Modality Worklist the patient found is copied into local database The patient information and the examination results are stored to the local database Images are st...

Страница 142: ... Voluson P8 Voluson P6 Physical Connection No network connection needed 3 12 1 5 Connection between Voluson P8 Voluson P6 and DICOM Server In this configuration the Voluson P8 Voluson P6 is configured to work with a DICOM server in a network environment Usually this will be the hospital network Images are first saved on the local image buffer on the System At the end of the examination the images ...

Страница 143: ... for the Voluson P8 Voluson P6 The IP addresses for the default gateway and other routers at the site for ROUTING INFORMATION Only if necessary e g for Internet access 1 Press the UTILITIES key on the control console 2 In the Utilities menu select SETUP button to invoke the setup desktop on the screen 3 On the left side of the screen select CONNECTIVITY and then click the DEVICE SETUP tab 4 Click ...

Страница 144: ...onnect the Wireless Network adapter as described in Section 3 5 5 on page 3 20 2 Press the Utilities key on the control console 3 In the Utilities menu press the Setup button to invoke the setup desktop on the screen 4 On the right side of the screen select Connectivity and then click the Device Setup tab 5 Click the WLAN Configuration button see Section Figure 3 76 on page 3 75 Figure 3 77 Intern...

Страница 145: ...e control console 2 On the right side of the screen select Connectivity and then click the Device Setup tab 3 Click the WLAN Configuration button see Section Figure 3 76 on page 3 75 4 The Wireless Network Configuration tool with available Wireless Networks appear 5 Select the WLAN you are connected to and then click Disconnected 3 13 2 3 Adding a WLAN Profile 1 Press the Utilities key on the cont...

Страница 146: ...ion button see Section Figure 3 76 on page 3 75 4 The Wireless Network Configuration tool with available Wireless Networks appear 5 Click Refresh 3 13 2 5 Setting a WLAN Network as Non Preferable When you make a WLAN non preferable you disconnect the network from the system and delete all connection settings from the system Afterwards the system WILL NOT try to reconnect to this WLAN automatically...

Страница 147: ... 3 13 2 7 Customizing an existing WLAN Profile 1 Press the Utilities key on the control console 2 On the right side of the screen select Connectivity and then click the Device Setup tab 3 Click the WLAN Configuration button see Section Figure 3 76 on page 3 75 4 The Wireless Network Configuration tool with available Wireless Networks appear 5 Select the Security tab and then click Customize 6 Cust...

Страница 148: ...the Device Setup tab 3 Click the WLAN Configuration button see Section Figure 3 76 on page 3 75 4 The Wireless Network Configuration tool with available Wireless Networks appear 5 Select the Properties tab 6 Click Available Channels 3 13 2 9 Monitoring the WLAN If there are wireless network communication issues you can monitor the wireless connection to see if it is dropping out and recorvering pe...

Страница 149: ... TCP IP protocol The diagnostics would showw if there are errors when performing the full two way communication For example if the IP address starts with 197 or 169 loopback addresses then something is wrong 1 Press the Utilities key on the control console 2 On the right side of the screen select Connectivity and then click the Device Setup tab 3 Click the WLAN Configuration button see Section Fig...

Страница 150: ...Voluson P6 1 Press the Utilities key on the control console 2 On the right side of the screen select Connectivity and then click the Device Setup tab 3 Click the WLAN Configuration button see Section Figure 3 76 on page 3 75 4 The Wireless Network Configuration tool with available Wireless Networks appear 5 Select the Diagnostics tab and then Click Repair NOTE DO NOT cancel the Repair operation af...

Страница 151: ...is available Used SIM card in the Smartphone should use same provider as SIM card for Cellular Modem Hardware Setup To use the feature the peripheral Cellular Modem needs to be installed and a SIM card from your local service provider needs to be ordered SIM card requirements Data capable Standard size prepaid or post paid NOTE Since the Voluson P8 Voluson P6 does not enable browser access the ser...

Страница 152: ...or connection to the service provider 12 Please wait while busy symbol is on screen If the connection is established Tooltip of the status icon provides signal strength in percentage If the connection is not established Tooltip of the status icon provides an error message 13 Depending on the region and or provider it might be possible that you need one time registration of your SIM card therefore ...

Страница 153: ...In the Utilities menu select the System Setup button to invoke the setup desktop on the screen 3 On the left side of the screen select Connectivity and then click the Device Setup tab 4 Click the Email Configuration button see Figure 3 76 on page 3 75 The Email Configuration screen appear 5 Enter Your name and Email Address in the Identity group As an option you can enter a different Reply to Addr...

Страница 154: ...son P6 Series sends an E mail with an attached image to the E mail to MMS service The MMS phone number is encapsulated in the recipient E mail address MMS Options 1 Press the Utilities H key on the control console 2 In the Utilities menu select the System Setup button to invoke the setup desktop on the screen 3 On the left side of the screen select Connectivity and then click the Device Setup tab ...

Страница 155: ...y pressing the Patient ID key on the control console 2 Click the icon 1 The Patient s Email address and Phone number window appear 3 Enter patient s information and adjust Email MMS Options 2 4 Store your configuration with click on OK 3 NOTE Since the E mail to MMS service provider is maybe not located in the same area as your MMS receptions you need to add the area code e g 49 or 0049 for German...

Страница 156: ...cted to a shared network drive of another server 5 Enter the name of the shared network folder in the Network Folder Name field 6 Supply a valid user name and a password for this folder NOTE If you check the Automatic Reconnect box the system tries to establish the connection again when starting up Otherwise the connection must be re established manually after a shutdown or reboot Figure 3 89 Syst...

Страница 157: ...op CSD NOTE There are different possiblilities to access the Common Service Desktop and its available features see NOTICE NOTICE If there is an error during the connection a warning message appears inside the dialog In this case please verify the data in the dialog NOTICE NOTICE If there already is a connection to the remote server the CONNECT button is grayed To change the existing connection fir...

Страница 158: ...and Country from the pull down menu Enter City State Postal Code and Institution where the system resides 4 Check settings in the Advanced Configuration section C Enterprise Server PRODUCT Service Center EURO Log Level WARN Further fields should be pre populated and should not be modified However please ensure correct setting Enterprise Server URL pre populated URL Enterprise Tunnel URL pre popula...

Страница 159: ...en close the page 7 Reboot 8 Reenter the Common Service Desktop CSD 9 Confirm that Service Connectivity is Configured and Checked Out in the Home page If the system indicated Checked Out you can be confident that the system has registered correctly if required verify further connectivity e g remote connectivity using your account 10 Close the page and check create InSite permanent user setting see...

Страница 160: ...screen click Add user 3 5 Fill in the required information and confirm with Add User 4 6 Check mark the appropriate user and then click Set Default Machine Contact 5 Color of the selected user turns from BLACK to RED 6 The permanent user is created 7 Close the window Figure 3 93 Add User Figure 3 94 Confirm with Add User Figure 3 95 Set User as Default ...

Страница 161: ... Destination Devices TCP IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Dept Voluson P8 Voluson P6 Type Floor Room REV Comments CONTACT INFORMATION Name Title Phone E Mail Address Site System IP Settings IP Address Subnet Mask Default Gateway Server Name Remote DB User Name Name AE Title Remote Archive Setup IP Address Subnet Mask Default Gateway Name...

Страница 162: ... Figure 3 96 Worksheet for DICOM Network Information Voluson P8 Voluson P6 Host Name AE Title Local Port IP Address Net Mask ROUTING INFORMATION ROUTER2 ROUTER3 ROUTER1 Destination IP Addresses GATEWAY IP Addresses Default DICOM APPLICATION INFORMATION NAME Store 2 Store 3D_1 Store 1 MAKE REVISION IP ADDRESSES PORT AE TITLE Print Worklist Store 3D_2 Storage Commit MPPS Structured Reporting ...

Страница 163: ...t all relevant safety and user information is available during the operation and service of the complete system 3 15 1 Product Locator Installation NOTE The Product Locator Installation Card shown may not be same as the provided Product Locator card 3 15 2 User Manual s Check that the correct User Manual s for the system and software revision is included with the installation Specific language ver...

Страница 164: ...GE RAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 3 96 Section 3 15 Paperwork This page was intentionally left blank ...

Страница 165: ...vice stick or hard disk drive At least one transducer See Chapter 9 Probes on page 9 19 for an overview normally you should check all the transducers used on the system Table 4 1 Contents in Chapter 4 Section Description Page Number 4 1 Overview 4 1 4 2 Required Equipment 4 1 4 3 General Procedure 4 2 4 4 Functional Checks 4 8 4 5 Backup and Restore Database Preset Configurations and Images 4 11 4...

Страница 166: ...EM REQUIRES ALL COVERS Operate this unit only when all board covers and frame panels are securely in place The covers are required for safe operation good system performance and cooling purposes NOTICE NOTICE Lockout Tagout Requirements For USA only Follow OSHA Lockout Tagout requirements by ensuring you are in total control of the Power Cable on the system NOTICE NOTICE After turning off a system...

Страница 167: ...ON position before starting the system However be aware some auxiliary equipment may switch itself to standby mode e g Color video printer and must therefore be switched on separately As soon as the software has been loaded the system enters 2D Mode with the probe and application that were used before the system was shut down Total time used for start up is about 2 minutes NOTE The power for on bo...

Страница 168: ...so performed when pressing the ON OFF button on the control panel twice 4 Switch OFF the Circuit Breaker at the rear of the system NOTE The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the ON OFF button So the auxiliary equipment need not to be switched ON OFF separately NOTICE NOTICE After turning off a system wait at least 10 seconds before turning it ...

Страница 169: ...r cable from the system or unplug it from the AC wall outlet socket 6 Press on the brakes to block the front caster wheels 7 Disconnect probes Figure 4 4 Circuit Breaker pull out protection and main power cable at rear of system CAUTION CAUTION DO NOT disconnect a probe while running Live Scan Write mode A software error may occur In this case switch the unit OFF perform a reset Pull out Protectio...

Страница 170: ...System Features 4 3 3 1 Control Panel Figure 4 5 Control Panel Tour 1 Loudspeaker position 7 Menu window flipswitch controls 2 Menu window digipot controls 8 Mode keys digipot controls 3 TGC Slider controls 9 Toggle switch controls 4 Trackball 10 Hard keys 5 Alphanumeric Keyboard 11 Trackbal buttons 6 Probeholders 1 2 3 4 9 7 5 10 11 6 8 ...

Страница 171: ...e Last First Middle Name 16 Status bar area 3 Patient ID number GA Gestational Age 17 P buttons 4 Probe Application 18 Run Freeze Cine 5 Depth Frame rate 19 Trackball 6 Mechanical Index 20 7 Thermal Index 21 8 Sonographer s Name 22 Rotary button and Paddle Key 9 Hospital Name Identification 23 On Screen Menu 10 Date 24 Bodymarks 11 Time 25 Depth scale markers 12 Image Info 26 Gray scale wedge 13 T...

Страница 172: ...e system s different modes refer to the Voluson P8 Voluson P6 Basic User Manual NOTE Some software may be considered standard depending upon system configuration If any Modes or Options are not part of the system configuration the check can be omitted NOTE Different menus are displayed depending on which Mode is selected Some function keys only appear if they are available for the selected probe ...

Страница 173: ...ia the DATA TRANSFER button it is possible to send images over the DICOM network print exams images export exams images import exams images etc Image History Image History gives you access to all the US pictures and exams of a particular patient Exam Review Exam Review allows you to view all exams of a particular patient Additionally it is possible to view image properties input comments and voice...

Страница 174: ...select CONNECTIVITY and then click the DRIVES tab 4 Click the ERASE CD button to displays the Erase DVD CD window as shown in Figure 4 9 5 Select the Erase Mode and click the OK button to start the process 6 When erasing is finished select EXIT to return to the Scan Mode Figure 4 8 System Setup Connectivity DRIVES page NOTICE NOTICE It is highly recommended to use the complete erase mode to avoid ...

Страница 175: ...ystem Configuration Full Backup on page 4 16 Section 4 5 4 Load Full System Configuration Full Backup on page 4 18 Section 4 5 5 Delete Full System Configuration Full Backup on page 4 20 The Image Settings and or Full System Configuration can be saved to the following destinations D partition of internal hard disk DVD CD R RW Mapped Network Drive see Section 3 13 2 Wireless Network Configuration o...

Страница 176: ...iopsy Lines 1 Insert a DVD CD R RW into the drive or connect an external USB device 2 Press the F2 key on the control panel 3 In the Utilities menu select SYSTEM SETUP to invoke the setup desktop on the screen 4 On the left side of the screen select BACKUP and then click the SYSTEM CONFIGURATION tab 5 Click the SAVE button of the Small Backup Scan Settings group 6 Choose the media and click the SA...

Страница 177: ...kup Scan Settings group 6 Choose the media and click LOAD NOTE If it is desired to load image settings from media Internal HDD click on the CHANGE FOLDER button browse for the folder D usersettings FactoryDefault Tetris and then click the LOAD button 7 Select the appropriate file and click OK 8 Select the desired loading procedure Load Complete Backup on page 4 14 Load only parts of the Complete B...

Страница 178: ... user prefers factory default settings which are adapted for the installed Application Software version 1 Perform Section 4 5 2 1 Preparations on page 4 13 2 Select the Complete Backup marked blue see Figure 4 13 below and click the button to copy the Complete Backup into the Load Data field 3 Click the LOAD button to start the loading procedure of the complete backup into the system The US Applic...

Страница 179: ...e version e g 2D 3D 4D Programs for new probe etc 1 Perform Preparations on page 4 13 2 Click the sign next to Complete Backup to open the content tree 3 For example Click the sign and copy the desired content by clicking the button and so on NOTE To return selected items from the Load Data field to Backup Data field select the button 4 Confirm selection with the LOAD button Settings will be loade...

Страница 180: ...ICOM settings e g DICOM servers AE Title Station Name etc Network Configuration network settings including the computer name Service Platform state of the Service Software NOTE It is recommended to Full Backup system configuration data before upgrading the software and or image settings presets This ensures that if settings need to be reloaded will be the same ones the customer was using prior to ...

Страница 181: ...ure 4 16 directory structure of full backup data Figure 4 17 Other drive drop down menu NOTICE NOTICE When the backup is saved to an external USB device the system has to be informed about the removal of the hardware For this purpose every last dialog of Full Backup Save and Full Backup Delete has a STOP USB DEVICES button see Figure 4 18 Figure 4 18 Please stop USB Devices before unplugging The s...

Страница 182: ...e are circumstances where it is not possible to load restore all the data The following rules specify these restrictions 1 Generally only restoring data from an older to a newer software version is possible Loading a backup into a system that has a lower software version than the system the backup was created on is prohibited 2 Options can only be restored on the same Voluson P8 Voluson P6 system ...

Страница 183: ...the SYSTEM CONFIGURATION tab 4 Click the LOAD button of the Full System Configuration group 5 Choose the Destination 6 Click on the backup to be restored additional information is displayed in the table 7 Select the NEXT button The following window will be displayed 8 Select the data to be restored to the Voluson P8 Voluson P6 system NOTE For description of the check box names refer to Save Full S...

Страница 184: ...tly 4 5 5 Delete Full System Configuration Full Backup 1 Press the F2 key on the control panel 2 In the Utilities menu select SYSTEM SETUP to invoke the setup desktop on the screen 3 On the left side of the screen select BACKUP and then click the SYSTEM CONFIGURATION tab 4 Click the DELETE button of the Full System Configuration group 5 Choose the Destination 6 Click on the backup to be deleted ad...

Страница 185: ...ontrol panel 2 In the Utilities menu select SYSTEM SETUP to invoke the setup desktop on the screen 3 On the left side of the screen select BACKUP and then click the IMAGE ARCHIVE tab 4 Click the SAVE button CAUTION CAUTION It is recommended to Save Image Archive Section 4 5 6 1 on page 4 21 and Backup the Full System Configuration Section 4 5 3 on page 4 16 once a week Figure 4 23 System Setup Bac...

Страница 186: ... Destination 10 Enter the description of the image backup 11 Select the NEXT button 12 To start the backup process click YES NOTICE NOTICE If for example All images older than 1 Day is chosen see Figure 4 24 on page 4 21 images of the current day will not be archived However if you click the Advanced button you can put this right Figure 4 25 Image Archive Save choose destination CAUTION CAUTION Vo...

Страница 187: ...click the IMAGE ARCHIVE tab 4 Click the LOAD button see Figure 4 23 on page 4 21 5 Choose the Destination 6 Click on the backup to be restored additional information is displayed in the table 7 Select the NEXT button The following window will be displayed 8 Select check mark the image archive data to be restored to the Voluson P8 Voluson P6 system 9 Click the NEXT button and then select YES to sta...

Страница 188: ...Check EUM Language settings desired EUM Language is displayed 5 User Setting Check all the User Settings settings assigned as desired by the customer 6 Patient Info Display Check all settings settings assigned as desired by the customer 7 Scan Assistant Check settings settings assigned as desired by the customer Table 4 4 System Setup Checks ADMINISTRATION Step Page Task Expected Result s 1 Option...

Страница 189: ...r when creating a new parameter in the MEASURE CALC page it is very important that all items are chosen correctly and that the relevant item is highlighted For further information refer to Chapter 18 in the Basic User Manual of Voluson P8 Voluson P6 Table 4 6 Measurement Setup Checks Step Task Expected Result s 1 Measure Calc Check all settings for all applications setting assigned as desired by t...

Страница 190: ...on 2 Press the remote key P1 P2 P3 which is assigned to the BW printer The image displayed on the screen is printed on the Black White printer 3 Press the remote key P1 P2 P3 which is assigned to the color printer The image displayed on the screen is printed on the Color printer 4 Press the remote key P1 P2 P3 which is assigned to recorder control Recording starts stops Step Task Expected Result s...

Страница 191: ... can be adjusted to a fixed height 910 mm with an Extender Kit H46712CE 4 7 2 1 Assembly procedure for converting Voluson P8 Voluson P6 from Low Version to High Version 1 Remove 2 RUBBER CAPS on SPEAKER COVER LEFT and RIGHT Figure 4 28 Removing the RUBBER CAPS 2 Unscrew 2 screws to remove the SPEAKER COVER LEFT and RIGHT Figure 4 29 Unscrewing 2 screws to remove SPEAKER LEFT and RIGHT ...

Страница 192: ... MONITOR CABLE COVER And unscrew 5 screws to disconnect the LCD cables Figure 4 30 Removing the MONITOR CABLE COVER and cables 4 Unscrew 1 screw to remove MONITOR ARM COVER BT16 Figure 4 31 Unscrewing 1 screw to remove the MONITOR ARM COVER BT16 5 Unscrew 1 screw to remove MONITOR CABLE FIX BRACKET Figure 4 32 Unscrewing 1 screw to remove MONITOR CABLE FIX BRACKET ...

Страница 193: ... DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 4 Functional Checks 4 29 6 Unscrew 9 screws under the keyboard base Figure 4 33 Unscrewing 9 screws under the keyboard base 7 Remove 5 Probe holders from the keyboard assembly Figure 4 34 Removing 5 probe holders ...

Страница 194: ...BASIC SERVICE MANUAL 4 30 Section 4 7 Peripheral Checks 8 Remove keyboard assembly Figure 4 35 Removing lower keyboard assembly 9 Disconnect the OPIO cable Power cable USB cables from keyboard assembly Figure 4 36 Disconnecting the cables from keyboard assembly Pull it downwards ...

Страница 195: ...tional Checks 4 31 10 Unscrew 4 screws to remove upper keyboard assembly Figure 4 37 Unscrewing 4 screws to remove upper keyboard assembly 11 Disconnect the OPIO cable power cable USB cables from upper keyboard assembly Figure 4 38 Disconnecting the OPIO cable Power cable and USB cable from upper keyboard assembly ...

Страница 196: ...2 Section 4 7 Peripheral Checks 12 Remove the cable tie Figure 4 39 Removing cable tie 13 Pull the LCD cables through the LCD ARM and MONITOR HOLDER Figure 4 40 Pulling LCD cables through the LCD ARM 14 Disconnect Speaker cables Figure 4 41 Disconnecting speaker cables Cable tie 2160955 speaker cables ...

Страница 197: ...o disassemble the keyboard base from the pedestal Refer to figure below 17 Hold the cast base firmly and lift it up as shown in the images below Figure 4 44 Holding the KEYBOARD BASE firmly and lift it up Figure 4 43 Unscrewing 4 screws to disassemble keyboard base ground cables remove 2 screws Hold the cast KEYBOARD BASE firmly and lift it Lift the base and pull the LCD cables OPIO cable and Spea...

Страница 198: ...nd place it on the extender as shown in images below Figure 4 45 Assembling Extender Table 4 8 Part list of Extender GEPN Description Qty 5432893 Pedestal Extender Cast 1 2327604 Flat Washer ISO7089 M8 Steel 140HV Zinc Plated 4 5412573 M8 Helical spring lock washers with flat end 4 5172296 Hexagon Socket Head Cap Screw M8 1 25 X25mm 4 Figure 4 46 Holding the KEYBOARD BASE and place in on the exten...

Страница 199: ...ing the keyboard base 21 Re assemble LCD cables through the LCD ARM Refer to figure below Figure 4 48 Re assembling LCD cables through the LCD ARM 22 Assemble HDMI cables for right direction monitor power cable monitor USB cable for left direction to align the second colored point as shown Then screw 1 screw to fix the cables using MONITOR CABLE FIX BRACKET Figure 4 49 Routing the cables and screw...

Страница 200: ... 36 Section 4 7 Peripheral Checks 23 Screw 1 screw to assemble MONITOR ARM COVER BT16 Then screw 5 screws to connect the LCD cables Figure 4 50 Screwing 1 screw to assemble LCD ARM COVER then connect the LCD cables 24 Connect the speaker cables Figure 4 51 Connecting the speaker cables speaker cables ...

Страница 201: ... Connnecting the ground cables 26 Connect the power cable and USB cables Figure 4 53 Connecting the power and USB cables 27 Reassemble the KEYBOARD UPPER COVER 28 Route the cables as shown in the figure below and tie cable tie Figure 4 54 Routing the cables ground cables Screw 2 screws power cable USB cables Screw 2 screws and assemble the KEYBOARD UPPER COVER Cable tie ...

Страница 202: ...Screw 2 screws and assemble the upper keyboard assembly Figure 4 55 Screwing 2 screws and assembling the upper keyboard assembly 30 Connect OPIO power and USB cables Figure 4 56 Connecting OPIO power and USB cables 31 Reassemble keyboard assembly Figure 4 57 Assembling the keyboard assembly OP cable power cable USB cable Push it upwards ...

Страница 203: ...ctional Checks 4 39 32 Screw 9 screws under the KEYBOARD BASE 33 Reassemble 5 probe holders Figure 4 59 Assembling 5 probe holders 34 Screw 2 screws to assemble SPEAKER COVER LEFT AND RIGHT Figure 4 58 Screwing 9 screws under the KEYBOARD BASE Figure 4 60 Screwing 2 screws to assemble SPEAKER COVER LEFT AND RIGHT ...

Страница 204: ...IC SERVICE MANUAL 4 40 Section 4 7 Peripheral Checks 35 Assemble 2 RUBBER CAP on SPEAKER COVER LEFT RIGHT 36 Screw 2 screws to assemble MONITOR CABLE COVER Figure 4 61 Assembling 2 Rubber cap on speaker cover left right Figure 4 62 Screwing 2 screws to assemble MONITOR CABLE COVER ...

Страница 205: ...apter 4 Functional Checks 4 41 4 7 3 Brakes and Direction Swivel Locks Check the brakes and swivel locks function as described below Table 4 9 Brakes and Direction Swivel Lock Step Task Expected Result s 1 Swivel lock engaged 2 Brakes and swivel lock released 3 Brakes and swivel lock engaged full lock ...

Страница 206: ... 3 Press B key on the operator panel to enter only B mode 4 Press CF key on the operator panel 5 Verify that B CF onde scan starts 4 8 4 PW Mode 1 Make sure that the probe is connected properly 2 Select the probe on the screen menu 3 Press PW key on the operator panel 4 Verify that PW mode scan starts 4 8 5 CW Optional NOTE This test is only applied if CWD hardware option has been installed 1 Make...

Страница 207: ...TION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 4 Functional Checks 4 43 Section 4 9 Site Log Table 4 10 Voluson P8 Voluson P6 Site Log Paper Documentation Voluson P8 Voluson P6 Date Service person Problem Comments ...

Страница 208: ...GE RAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 4 44 Section 4 9 Site Log ...

Страница 209: ...ribution scheme and probes Table 5 1 Contents in Chapter Voluson P8 Voluson P6 Section Description Page Number 5 1 Overview 5 1 5 2 General Information 5 2 5 3 FrontEnd Processor 5 22 5 4 BackEnd Processor 5 27 5 5 Internal I O 5 30 5 6 OPIO User Interface 5 31 5 7 Monitor 5 37 5 8 Monitor setup 5 37 5 9 External I O 5 38 5 10 Peripherals 5 39 5 11 Power Distribution 5 40 5 12 Service Platform InS...

Страница 210: ...trasound imaging system The system can be used for 2D Mode Imaging and additional Operating Modes XTD View Color Doppler Imaging CFM PD and HD Flow M Mode MCF Imaging Doppler PW CW 3D Mode and Real Time 4D Imaging Different combinations of the above modes Figure 5 1 Voluson P8 Voluson P6 Major Components 1 Monitor 2 Loudspeaker 3 Probe Holder 4 DVD Drive 5 Place for B W and or Color printer 6 Lock...

Страница 211: ...xing Signal flow from the Probe Connector Panel to the FrontEnd FE Electronics to the BackEnd Processor and finally is displayed on the LCD monitor and peripherals Voluson P8 Voluson P6 internal electronics are divided into three Front End FE Processor Back End Processor Power Supply Unit Interconnecting signals from FrontEnd BackEnd keyboard monitor and power distribution sub systems are routed i...

Страница 212: ...SON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 5 4 Section 5 2 General Information Section 5 2 General Information cont Figure 5 2 Basic Block diagram of Voluson P8 Voluson P6 ...

Страница 213: ...arious clinical applications Any 3 probes may be connected at the same time Medical application fields include Obstetrics Gynecology and Fertility Radiology Neurology Oncology Urology Pediatrics Musculoskeletal The system is designed for follow up expansion In addition to the initial operational settings for each transducer pre programmed in the system user customized parameter settings for each t...

Страница 214: ...natomy Coded Harmonics may improve the B Mode image quality without introducing a contrast agent 5 2 1 1 2 XTD View XTD View Extended View provides the ability to construct and view a static 2D image which is wider than the field of view of a given transducer This feature allows viewing and measurement of anatomy that is larger than what would fit in a single image XTD View constructs the extended...

Страница 215: ...rounding tissue although additional parameters may also be used to enhance the discrimination Color flow can be used in combination with 2D and Spectral Doppler modes as well as with 3D mode 5 2 1 3 2 Power Doppler A real time two dimensional cross section of blood flow is displayed The 2D cross section is presented as a full color display with various colors being used to represent the power in b...

Страница 216: ... rendering 5 2 1 6 3D Data Collection and Reconstruction 2D gray scale images including Color Flow or Power Doppler information may be reconstructed The acquisition of volume data sets is performed by sweeping 2D scans with special transducers called 3D transducers designed for the 2D scans and the 3D sweep 2D ultrasound imaging modes are used to view a two dimensional cross sections of parts of t...

Страница 217: ...ualize certain 3D structures of a scanned volume by means of a 2D image The gray value for each pixel of the 2D image is calculated from the voxels along the corresponding projection path analyzing beam through the volume The render calculation algorithm surface or transparent mode determines how 3D structures are visualized Figure 5 3 Principle Volume Rendering PIXEL VOXEL Y X Z PROJECTION WAY 2D...

Страница 218: ...USON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 5 10 Section 5 2 General Information 5 2 2 Block diagram Voluson P8 Voluson P6 Figure 5 4 Voluson P8 Voluson P6 Block diagram ...

Страница 219: ...agram For further details see Front End Processor Block diagram on page 5 22 see T DPI Blockdiagram on page 5 23 see Beamformer Frontend on page 5 24 see DRFM Block diagram on page 5 25 5 2 2 2 Back End Block diagram For further details see BackEnd Processor Block diagram on page 5 27 5 2 2 3 OPIO Block diagram For further details see Main OPIO Block Diagram on page 5 31 ...

Страница 220: ...tly combine Nearfield Frame with Farfield Frame in FFC Mode see Section 5 2 3 1 1 Special B Mode Techniques on page 5 13 Multibeam DeInterleave means Incoming Pixel order shot1pix1 shot2pix1 shot3pix1 shot4pix1 shot1pix2 shot2pix2 shot3pix2 shot4pix2 is converted to the new order shot1pix1 shot1pix2 shot1pix3 shot2pix1 shot2pix2 shot2pix3 After DC cancelling the signal is mixed with RX Frequency a...

Страница 221: ...erent direction steered images PC calculated 5 2 3 2 M Mode 1 DRFM see Section 5 2 3 1 B Mode on page 5 12 a DMA section B mode Data is written via SP0 into SDRAM Fifo Buffer memory DMA Controller 1 transfers the data into SOM System On Module main memory where scan conversion is performed per software i e the sweep image is generated scaling and interpolation between lines CineMode CineMode Memor...

Страница 222: ...ignal is mixed with RX Frequency and brought to LF Spectrum where the LowPassFilter cuts HF Mixer and Magnitude Calculator arrange Complex Demodulation a DMA section D mode Data is written via SP0 into SDRAM Fifo Buffer memory PC calculates Autotrace Curve from D Mode data Cine Mode with Autotrace Cine Mode Memory for the Autotrace Curve is inside SOM Memory Software has to take care that D Spectr...

Страница 223: ... with RX Frequency and brought to LF Spectrum where the LowPassFilter cuts HF Mixer and Magnitude Calculator arrange Complex Demodulation 5 2 3 6 3D Mode Freezes after 1 volume sweep see Section 5 2 3 1 B Mode on page 5 12 5 2 3 7 Real Time 4D Mode nonstop volume rendering see Section 5 2 3 1 B Mode on page 5 12 5 2 3 8 XBeam CRI Mode CrossBeam Compound Resolution Imaging see Section 5 2 3 1 B Mod...

Страница 224: ...and then click the OPTIONS tab NOTE Option a 4D Realtime Static 3D SonoRenderlive Option b 3D 4D Activation TUI SingleView Option c Can be used with 3D 4D Activation Table 5 2 Software Options Options VP6 VP8 3D 4D Activation Option a Option a Advanced 3D 4D Package Option b Option b Advanced VCI N A Option c SingleView Option c Option c SonoAVC Follicle Pro Option c Option c SonoL D Option Option...

Страница 225: ...nal to the parallel slices shows which parts of the volume are displayed in the parallel planes This method of visulaization is consistent with the way other medical systems such as CT or MRI present the data to the user The distance between the different planes can be adjusted to the requirements of the given data set In addition it is possible to set the number of planes The planes and the overv...

Страница 226: ...be changed to red and the measurement will show only one result 5 2 4 7 SonoL D Enhances the labor and delivery process with quantitative progression data 5 2 4 8 TUI TUI is a new visualization mode for 3D and 4D data sets The data is presented as slices through the data set which are parallel to each other An overview image which is orthogonal to the parallel slices shows which parts of the volum...

Страница 227: ...lected to undergo surgery The LR2 model is based on published literature and has been tested only on the stated population The literature states that the LR2 Model can help to estimate the probability of an adnexal mass of being malignant The IOTA group has evaluated other ways including an LR2 model in additional to the LR2 model 5 2 4 14 DICOM Software package providing following DICOM functiona...

Страница 228: ...mformer CW this beam is steerable by the operator and is done by adjusting the location of the M line The CW Doppler beam or M mode line can be steered allowing interrogation along an operator selected line within the image NOTE This feature is available only on Voluson P8TM systems 5 2 5 2 ECG Module For details see Section 5 10 3 ECG Module on page 5 39 5 2 5 3 Wireless Network Adapter WLAN Wire...

Страница 229: ...ndows settings IP address Network Name etc US Application Software UISAPP Global Service Platform Software Software Options D User partition User Presets Backup database Images Archive Patient ID s and Reports database Service database System settings database R Rescue partition Factory Images of C Partition for System recovery after HDD Windows crash Printer Drivers LINUX partition not visible in...

Страница 230: ...le 5 4 below outlines the Voluson P8 Voluson P6 Front End components described in the sub sections Figure 5 6 Front End Processor Block diagram Table 5 4 Voluson P8 Voluson P6 Front End components Sub section Description Page Number 5 3 1 T DPI Probe Interface board 5 23 5 3 2 DBM64 Beam former board 5 24 5 3 3 DRFM Radio Frequency interface and System controller 5 25 ...

Страница 231: ...L Chapter 5 Components and Functions Theory 5 23 5 3 1 T DPI Probe Interface board Switches the Probe Connectors 3 RS Connectors and recognizes Probes 3 RS Probe Connectors Probe Select Relays Probe Recognition For Block diagrams refer to Figure 5 7 T DPI Blockdiagram Figure 5 7 T DPI Blockdiagram ...

Страница 232: ...rd has 64 TX RX channels respectively beamformer board is a complete 64 channel ultrasound Front end that includes the following features and functions TX pulse genetration TX Apodization TX Focus RX Focus Delay and summation Full receiving channel with integrated AFE IC TGC Anti Aliasing Filter ADC High Voltage Multiplexer 192 x 64 channels DBM64 Figure 5 8 Beamformer Frontend ...

Страница 233: ... P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 5 Components and Functions Theory 5 25 5 3 3 DRFM Radio Frequency interface and System controller Figure 5 9 DRFM Block diagram ...

Страница 234: ...cessing Raw ultrasound data is coming in from the Interface FPGA It is pre processed Further processing will be done by the PC The pre processed data is transferred back to the Interface FPGA E System control PRF Pulse Repetition Frequency Generator Line Number Lateral Position Control of TX Power DACs generation of TX POWER reference voltage Controls discharge of capacity on DPS board during the ...

Страница 235: ...VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 5 Components and Functions Theory 5 27 Section 5 4 BackEnd Processor Figure 5 10 BackEnd Processor Block diagram ...

Страница 236: ...led by the CPU via signals from SATA is divided into 4 partitions For further details see Section 5 2 6 Data Location on page 5 21 5 4 2 SOM System On Module 5 4 2 1 Congatec SOM Module for VS6 S8 S10 On Board HDMI LAN USB 2 0 Sound CPU Intel CoreTM i5 3610ME Processor 2 7GHz SOM Module for VP6 P8 On Board HDMI LAN USB 2 0 Sound CPU Intel CoreTM i3 3120ME Processor 2 4GHz Major Tasks System Contro...

Страница 237: ...SON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 5 Components and Functions Theory 5 29 Control DVD drive SATA User Interface USB DVR USB Figure 5 11 SOM System On Module ...

Страница 238: ...GE RAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 5 30 Section 5 5 Internal I O Section 5 5 Internal I O Figure 5 12 Voluson P8 Voluson P6 Internal I O ...

Страница 239: ...ders with integrated push buttons USB Trackball 2 with dedicated buttons to emulate standard three button mouse USB standard alphanumeric keyboard LED Indicators with wide range dimming Figure 5 13 Main OPIO Block Diagram Micro Controller USB USB HUB Down Stream 2 Power LED 4 Function S W 9 LED Control Backlight Highlight LED 174 Encoder 10 WHEEL 1 LED Control Atmega Function S W 40 Trackball S W ...

Страница 240: ...yout Figure 5 14 Voluson P8 Voluson P6 OPIO Layout 5 6 3 Button Description 1 2 3 4 9 7 5 10 11 6 8 1 Loudspeaker position 2 Menu window digipot controls 3 TGC Slider controls 4 Trackball 5 Alphanumeric Keyboard 6 Probeholders 7 Menu window flip switch controls 8 Mode keys digipot controls 9 Toggle switch controls 10 Hard keys 11 Trackball buttons ...

Страница 241: ...GE DRAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 5 Components and Functions Theory 5 33 ...

Страница 242: ...GE RAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 5 34 Section 5 6 OPIO User Interface P3 Programmable key ...

Страница 243: ...r 5 Components and Functions Theory 5 35 In general operations at diverse dialog pages and windows on the system desktop e g Patient Data Entry Usage of EUM System Setup Measurement Setup etc are executed by trackball and the trackball keys mouse emulation Function of the Trackball at Diverse Dialog Pages ...

Страница 244: ...GE RAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 5 36 Section 5 6 OPIO User Interface SW DVR Go to SW DVR menu ...

Страница 245: ...luson P6 system it s accessible via not monitor adjustment but monitor menu on application Please refer to Section 5 8 Section 5 8 Monitor setup NOTE Changing the monitor settings through the utility page it takes 1 second to apply the adjusted value to the monitor 1 Press the UTILITIES key on the control panel On the left side of the screen select the MONITOR Figure 5 15 Monitor settings ...

Страница 246: ...External I O Connectors on Rear of System GES Item Connector Name Description 1 USB USB 2 0 port 2 HDMI OUT Connector for external Monitor TMDS signal with DDC_Control HDMI or DVI Resolution SXGA 1280 X 1024 3 NETWORK DICOM input output twisted pair RJ 45 10 100 megabit s 4 RGB RBG Analog Output 5 Audio out Audio out for external speaker Figure 5 17 External I O Connectors on the TOP OPIO Item Con...

Страница 247: ...d other service utility operations The DVD R RW drive can read write CDs and DVDs And also it is one of unit of DVR Recorder when DVR option is installed 5 10 3 ECG Module The ECG Module is used for acquiring an ECG signal to be displayed with the ultrasound image This optional peripheral serves for gaining an ECG signal to mark the systolic and end diastolic moments in M Mode and Doppler evaluati...

Страница 248: ...SIC SERVICE MANUAL 5 40 Section 5 11 Power Distribution Section 5 11 Power Distribution 5 11 1 ACFE Primary Power Module Figure 5 18 Primary Power Module Curcuit breaker Equipotential Connection Main Voltage Indicator Shows input voltage level Pull out Protection Main power cable ...

Страница 249: ...GE DRAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 5 Components and Functions Theory 5 41 5 11 1 1 DPS Block Diagram Figure 5 19 DPS Block Diagram ...

Страница 250: ...ON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 5 42 Section 5 11 Power Distribution 5 11 1 2 Power Distribution Block Diagram Figure 5 20 Power Distribution Block Diagram T DPI ...

Страница 251: ...e ON button on the Operator Panel DRFM power up 5 11 1 3 1 Power Up Sequence Description 1 DRFM powered by 5Vstb detects contact of Power On button event 2 DRFM sends PSON_N low to DPS 3 DPS powers up 24V 12V 5V 4 DPS provides LV_OK as soon as 12V is within specification 5 DPS provides PC voltages from 12V 5V 6 DRFM distribute enable signals for 3 3V 2 5V 1 2V 1 1 to itself and DBM64 7 DRFM provid...

Страница 252: ... 5VSB 5 11 1 4 Mechanical Concept and Overview The AC Power s main tasks are to supply the internal subsystems with AC power To reduce inrush current an inrush current limiter is implemented Voltage to peripherals are provided from input without configuration From the Mains Power Input module the AC power is routed via an Inrush Current Limiter to an internal ACDC module 5 11 1 5 Functions of ACFE...

Страница 253: ...Supply 5 11 2 1 Function of ACDC With the AC input voltage from the ACFE the ACDC module generates 12Vdc 5 11 3 DCDC main DC Power Supply With the 12Vdc from ACDC the DCDC module generates all DC power sorces Front end voltages Standby voltage Tx voltages DVD supply 4D module supply Fan supply 5 11 4 4D Module In addition the AC DC device contains the digital motor amplifier ...

Страница 254: ... by GE technicians through Common Service Desktop Images and log files download Virtual Network computing 5 12 2 Access Security The Service Platform has different access and security user levels Each user is only granted access to the tools that are authorized for their use Local Access via System Setup Administration SERVICE page Remote Access This offers the possibility to view the entire custo...

Страница 255: ...e the dongle 5 Enter the Standard GE Revolving password changes every six months and then click ENABLE 6 On the right side of the screen select ADMINISTRATION and then click on the SERVICE tab 7 Click the CSD button 8 As soon as the GEHC Service Home Page appears select GE Service from the pull down enter the password password changes every six months and then click Okay The Common Service Desktop...

Страница 256: ...n P8 Voluson P6 System requires permission and customer before a GE service technician or OLC can access the customer s system remotely Every access request whether successful or not as well as actions and changes will be logged into a service access log that is viewable to authorized users Disruptive Mode can be selected by the customer directly on the Voluson P8 Voluson P6 as described below or ...

Страница 257: ...ystem Remote access to the Voluson P Series system requires permission and customer input before a GE service technician or OLC can access the customer s ultrasound remotely 3 Local Extended Access by Windows Start Menu All Programs Voluson GE Field Engineer Common Service Desktop NOTE A Service Dongle and the Standard GE Revolving password password changes every six months is necessary when perfo...

Страница 258: ...screen 3 Connect the USB standard GE service dongle to the USB connector s on the rear of the system As soon as the dongle is detected the login window pops up Remove the dongle 4 Enter the Standard GE Revolving password changes every six months and then click ENABLE 5 On the left side of the screen select ADMINISTRATION and then click the SERVICE tab 6 Click the SERVICE TOOLS button to display th...

Страница 259: ... table which describes the procedure to upgrade the current master image by the new one on the DVD This table is sored as CSV text on the System DVD It logs the development history of the master images via a successive Svn nr At the same time the steps necessary for upgrading an old image with a new one are defined NOTICE NOTICE For intermittent problems this file can be requested from the Service...

Страница 260: ...grams and Auto Text remain unaffected Therefore it is not necessary to perform any readout preparations 5 14 3 3 TCP IP Buffersize The TCP IP Buffersize selects the amount of buffer memory used for DICOM transfers both directions 5 14 3 4 Show Factory Presets Select the Show Factory Presets button to display User and Factory Presets probe settings 5 14 3 5 Common Service Desktop CSD Access to the ...

Страница 261: ...ton to check control console function Operation see 5 14 3 10 Keyboard Layout To change the keyboard layout to different languages Operation see Section 6 4 Modification of Keyboard Layout on page 6 3 NOTE Reconfigure the layout of the keyboard is only useful by changing the concerned keys also see Section 8 51 Replacement of AN KEY TOP on page 8 155 5 14 3 11 Monitor Test Select the MONITOR TEST ...

Страница 262: ... desired file and then click OK Load procedure starts immediately including a reboot of the system 4 If warning message is displayed confirm it with OK and then load appropriate Application Settings perform loading procedure as described in steps above NOTICE NOTICE When reloading these Application Settings any existing User Programs 3D 4D Programs and Auto Text remain unaffected Figure 5 29 Appli...

Страница 263: ...nd or choose correct probe name from the pull down menu 1 3 Check and or enter probes Serial Number in the appropriate SNO filed 2 4 Select Today from the Start date pull down menu 3 5 Click the OK Global EEPROM button 4 NOTE DO NOT disconnect the probe during the re programming process Probe ID and coding is assigned to the probe that is connected to the selected prot 6 Confirm with OK CAUTION Do...

Страница 264: ...E Service The Voluson Maintenance System functions are secured by service password class C 1 Turn system OFF and then back ON 2 As soon as the Boot Screen appears see Figure 5 31 above press the PG DN Arrow down key on the keyboard until the GE Service item is highlighted then press ENTER 3 Enter the Password Rudi case sensitive and confirm with ENTER The system boots into LINUX Figure 5 31 Boot s...

Страница 265: ...n appears see Figure 5 31 on page 5 56 press the PG DN Arrow down key on the keyboard until the Rollback item is highlighted then press ENTER 3 When the following WARNING message appears press the Arrow left button to highlight OK and then press ENTER After performing the rollback the system reboots The Ultrasound Application is started and finally the 2D screen is displayed on the monitor CAUTION...

Страница 266: ...cally and will take about 2 5 hours If there are errors they will be listed NOTE After one cycle 2 5 hours the memory check starts again To interrupt the test press the Ctrl Alt Del key If you don t interrupt the memory test it will perform never ending cycles of memory checks NOTE If after one cycle about 2 5 hours no error messages are listed it can be assumed that the Back End Processor includi...

Страница 267: ...he dongle 3 Enter the Standard GE Revolving password changes every six months and then click Enable 4 Press Ctrl ESC keys simultaneously and open the Windows Start menu Description of Sub folder GE Field Engineer Figure 5 34 Windows Start Menu Table 5 6 Sub folder GE Field Engineer Windows Tool Description Command Prompt Performs text based command line functions Common Service Desktop Access to E...

Страница 268: ...in th System Setup Service page to update the System Software Operation See Start Linux auto System boots into Start Restore Activates Restore C from recovery partition R operation see Start Rollback Activates Rollback operation see Taskmanager Provides information about programs and processes running on the computer it also displays the most commonly used performance measures for processes WinExp...

Страница 269: ...m Electronic User Manual EUM Drivers for Peripherals e g printer Scripts 5 17 1 1 General Operation and Selection in LINUX and BIOS menu 5 17 2 Access Security There are different possibilities to access the LINUX Maintenance System and its available functions System is still responsive starts up without problems application software is running as usual via Boot Screen Functions GE Service functio...

Страница 270: ...figuration group Select destination enter description and then click on NEXT For more details about Full Backup see Section 4 5 3 on page 4 37 Click the IMAGE ARCHIVE tab and then click on SAVE Select destination enter the description and then click on NEXT For more details about saving the Patient Image Archive see Section 4 4 1 Patient Archive Image Management on page 4 9 CAUTION If the system i...

Страница 271: ...CONFIGURATION button and print out the internet Prodocol TCP IP Properties dialog page or write down the IP settings 7 On the left side of the screen select GENERAL SETTINGS Click the GENERAL SETTINGS tab and print out or write down the Time format setting 8 Disconnect all external USB devices such as printers hubs bluetooth memory devices as this might interfere with the recovery installation pro...

Страница 272: ... it is necessary to create a New Hard disk which contains factory default partition images Linux C D and R booting from System DVD is required To change Boot Priority order in BIOS and setup new HDD see NewDisk on 5 69 5 17 4 2 Advanced This item switches to the Advanced Functions Menu see 5 17 6 Advanced Functions Menu on page 5 75 that contains the functions Restore and Initialize 5 17 4 3 About...

Страница 273: ... the System DVD to replace the existing image of the C partition Additionally the R Rescue image of C is replaced by the DVD image The System C image contains Windows 7 with patches Ultrasound Application Software UIS Image Application Settings Backup Electronic User Manual EUM Operation NOTICE The Media Menu is only accessible if the System DVD see Table 9 13 is inserted into the system s DVD CD ...

Страница 274: ... specific Permanent key is required a Enter the appropriate Permanent Key calculated in OKOS http 3 187 187 9 OKOS select OK and confirm with Enter If the entered Permanent key is correct following window appears confirm with Enter 2 Only if you are sure you have saved the Image Archive and Full System Configuration select Yes in the following dialog box and confirm with Enter NOTE If the message ...

Страница 275: ...e System DVD from the DVD CD R RW Drive drive 6 Please wait until all processes are finished and the final DOS window Figure 5 42 below appear 7 Press A key to activate the Ultrasound Application Software UIS Figure 5 40 Restore C done Figure 5 41 Exit Exit Reboot NOTICE The system boots with windows An automated process was developed to check disk assign an unique computer name and matching drive...

Страница 276: ...NUAL 5 68 Section 5 17 LINUX Voluson Maintenance System 8 If prompted enter the proper Permanent key and then click Submit NOTE Options will be recovered at a later step using Full Backup process 9 Customize the system as described in Section 5 17 8 on page 5 80 Figure 5 43 enter Permanent key ...

Страница 277: ...IRE HARD DISK Also data on D User partition e g Patient Archive images Service Database Measurement results Patient data and Reports will be lost and can not be carried over to the new hard disk NOTICE However if system functionality allows it save the patient images and corresponding patient data via System Setup Backup IMAGE ARCHIVE as described in Section 4 5 6 on page 4 21 gather full system c...

Страница 278: ...tem DVD into the DVD Drive 2 Turn system off and then back on 3 During startup press the DEL key immediately to enter the BIOS Setup 4 Enter the BIOS password polaris case sensitive and then press Enter 5 Use the Arrow keys or on the alphanumeric keyboard to select the Boot tab 6 Using the keys PgUp Arrow up and PgDn Arrow down on the alphanumeric keyboard select Set Boot Priority and then press E...

Страница 279: ... Press the key on the keyboard to change the Boot Priority order 1st Boot from Hard Disk SATA 2 Drive to DVD Drive SATA 0 Drive 2nd Boot from DVD Drive SATA 0 Drive to Hard Disk SATA 2 Drive 8 Press the F10 key on the alphanumeric keyboard 9 Confirm the message with Yes and Enter to reboot the system Proceed with Figure 5 44 Save Exit Set Priority ...

Страница 280: ...for automatic boot into the Voluson Maintenance System 2 In the displayed menus select Media NewDisk and then press ENTER 3 The following warning message will be displayed 4 ONLY if you are absolutely sure select YES and then confirm with ENTER 5 Creating a new disk will take some time Figure 5 45 Media NewDisk Figure 5 46 WARNING WARNING WARNING NOTICE If you select YES all data on the entire har...

Страница 281: ...ng the Arrow keys or on the alphanumeric keyboard select the Save Exit tab 10 Select Load GE BIOS Setting xxxx xxxx is the installed BIOS version and then press Enter This changes the Boot Priority order back to 1 Boot Hard Disk 11 Press the F10 key on the alphanumeric keyboard to save configuration changes 12 Confirm the message with Yes and then press Enter NOTE The system boots with windows An ...

Страница 282: ... wait until all processes are finished and the final DOC window see figure below appear 14 Press A key to activate the Ultrasound Application Software UIS 15 If prompted enter the Permanent key and then click Submit 16 Customize the system as described in Section 5 17 8 on page 5 80 Figure 5 49 press the A key to start UlSApp Figure 5 50 enter Permanent key ...

Страница 283: ...rface for network setup The Network function is not intended for use in the field 5 17 6 3 SSHD By selecting this item a SSH daemon is started on the system The SSHD function is reserved for extended troubleshooting Not intended for use in the field 5 17 6 4 Restore C from recovery partition R This image stored on the Rescue R partition is used to replace the existing image of the C partition NOTE...

Страница 284: ...e image on the on the Rescue R partition of the hard disk is used for restoring C WARNING WARNING All data on D User partition e g Patient Archive images Measurement results Patient Data and Reports will be deleted and re initialized NOTICE The Initialize function should only be performed if NO System DVD is available If you have a System DVD it is highly recommended to use the New Disk function s...

Страница 285: ...vanced 3 In the Advanced Functions menu Figure 5 51 on page 5 75 select Service Tools The Service Tools Menu with following functions appear NOTE Use the keys PgUp Arrow up and PgDn Arrow down on the alphanumeric keyboard to scroll up down 5 17 7 1 Exit Return to the Advanced Functions Menu 5 17 7 2 Shell By selecting this item a Bash shell is opened Enter exit to return to the main menu The Shell...

Страница 286: ... about the hard disk s performance and operations detects and reports various indicators of reliability in the hope of anticipating failures NOTE This test takes about 1 3 hours depending on the HDD size 5 17 7 5 HDDWipeTool By means of this function all data e g patient data archive images measurement results and reports on the hard disk will be removed and overwritten with zeros Figure 5 53 USB ...

Страница 287: ... Backup Process Logger files Auto tester files SMART logs dump files etc 1 Connect an USB stick 2 In the Service Tools Menu select CollectLogs and then press Enter 3 In the displaye screen select your USB stick and then press the Space key An asterisk appears between teh brackets 4 Click OK to save the log files 5 17 7 7 CheckLnkSpeed This checks link speed communication between Beamformer and Bac...

Страница 288: ... as well as step 13 to step 14 in this section can be omitted 8 On the left side of the screen select ADMINISTRATION click the OPTIONS tab and enter the codes for the options as described in Section Section 8 7 Replacement or Activation of Options on page 8 23 Clickthe Service tab enter the password SHE and then click ACCEPT In the displayed Service Tools window click the keyboard Layout button to...

Страница 289: ...of specification Although some tests may be optional they should only be performed by qualified personnel Section 6 2 Regulatory Verify where applicable that any regulatory information or tests required by national law are present and accounted for and any regulatory tests required by national law are performed and documented Table 6 1 Chapter Voluson P8 Voluson P6 Contents Section Description Pag...

Страница 290: ... 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 6 2 Section 6 3 LCD Monitor Adjustment Section 6 3 LCD Monitor Adjustment NOTE From Voluson P8 Voluson P6 system it s accessible via not monitor adjustment but monitor menu on application Please refer to Section 5 8 ...

Страница 291: ...ress the UTILITIES key on the control console 2 In the Utilities menu select SETUP button to invoke the setup desktop on the screen 3 Connect the USB Standard GE service dongle to the USB connector s on the rear of the system As soon as the dongle is detected the login window pops up Remove the dongle 4 Enter the Standard GE Revolving password changes every six months and then click Enable 5 On th...

Страница 292: ...8 If not already done select the default input language from the drop down menu see Figure 6 1 The corresponding keyboard layout is changed automatically 9 Click on APPLY and then close the window with OK 10 Close the Service page with the SAVE EXIT and restart the system 11 Reenter Keyboard Layout by repeating step 1 to step 6 This time remove unused language s 12 Test the Keyboard function Press...

Страница 293: ...mote diagnostics or to be sent back to the manufacturer for analysis There are different options to acquire this data that would give different results Table 7 1 Contents in Chapter 7 Section Description Page Number 7 1 Overview 7 1 7 2 Collect Vital System Information 7 2 7 2 1 Request for Service DRFM 7 4 7 2 2 Shortcuts List 7 6 7 3 Check Points Voltages 7 7 7 4 Screen Captures and Logs 7 8 7 5...

Страница 294: ...lect ADMINISTRATION and then click the SYSTEM INFO tab The following information is necessary in order to properly analyze data or images being reported as a malfunction or being returned to the manufacturer System Type System Serial number also visible on label on the back of the system Application Software version Backup Version File Name Date of Factory Settings Tune version etc additional info...

Страница 295: ...RVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 3 Section 7 2 Collect Vital System Information cont Figure 7 1 System Setup Administration SYSTEM INFO page e g VP8 Move the scroll bars downwards to view additional information about installed software hardware e g Console information ...

Страница 296: ...MINISTRATION and then click the SERVICE tab 4 Enter the password SHE and click the ACCEPT button to display the Service Tools window 5 Click the DRFM button and fill out the displayed form Enter detailed Problem description 4 Click the SEND button to send the problem description to GE Service Application representatives NOTICE NOTICE Service Connectivity has to be checked out once before you can r...

Страница 297: ...equest confirmation screen is displayed 5 Write down and keep the Reference Number for follow up procedures then click OK NOTE If the service platform is not configured an Error message is displayed The request is NOT sent The request is saved in QUEUE for reviewing and or resending Figure 7 3 Request submitted Figure 7 4 Contact GE Queue write down and keep the Reference Number ...

Страница 298: ...tion Ctrl C copy text Alt D export system state Full Backup and Dump files Ctrl H shortcuts list see Figure 7 5 above Alt P stores screenshot on D export Ctrl V paste insert text from clipboard Ctrl X cut text Ctrl Z undo typing Alt F5 create process logger snapshot Alt F6 burn process logger snapshot to CD Alt F7 start stop process logger Shift F11 start remote socket control Alt F11 stop remote ...

Страница 299: ...cs Troubleshooting 7 7 Section 7 3 Check Points Voltages 7 3 1 How to check power 7 3 1 1 Dachstein Power Supply DPS Turn on mains switch and check green LED of P5V N5V LIVE5V and 12V on backplane NOTE If LED is on DPS has full function all voltages are within the specified range Figure 7 6 Check green LED on backplane ...

Страница 300: ...g is the generic process to capture any screen from the System 1 Navigate to and display the image screen volume to be captured 2 Press the P1 P2 or P3 key depending on system configuration on the control panel and store the image onto the clipboard frame on left side of the screen NOTE A short summary of P1 P2 and P3 key s configuration is shown in the status area on the screen 3 Select the store...

Страница 301: ...s a DVR it will show up in the Interface list NOTE If you are working with a Wireless connection you must uncheck the Capture packets in promiscuous mode A The wireless option does not support this mode 4 If desired select a Capture Filter from the pull down menu 2 see 5 Adjust Display Options 3 for best performance to avoid packet loss on high throughput 6 Select Start 4 7 Minimize network sniffe...

Страница 302: ...for the troubleshooting 10 Select Capture Stop or select the icon in the task bar to stop the capture A Packnet List the packet list pane displays all the packets in the current capture file Each line in the packet list corresponds to one packet in the capture file B Packet Details pane shows the current packet selected in the Packet List pane in a more detailed form This pane shows the protocols ...

Страница 303: ...rhe filltered values rather than the entire capture 12 Click Save 6 NOTE Wireshark uses lipcap pcap cap file format as the default format to save captured packets If you need to open this file with D Trace or DVTK the capture canb eresaved in NA Sniffer Windows format 13 Exit the sniffer application by clicking on the X in the upper right corner 14 If you have already performed Save as you can con...

Страница 304: ...ter name A For example DICOM port 104 4 Enter the Filter string with the following syntax port xxx and host yy yy yy yy where xxx is the number of My computer in the System and yy yy yy yy is the IP address of the device under test PACS server etc 5 Click OK 2 by IP only 1 Click on CAPTURE FILTER see Figure 7 12 on page 7 12 2 Select IP only B 3 Use the IP address of the System or the one from the...

Страница 305: ... addition to the pre capture filter use the Filter tool in the screen to filter what is displayed Filter the DICOM packets since they are the most probable for the troubleshooting 1 Type dicom in the Filter field and click on APPLY The display should filter all DICOM packets filtering out image data 2 To deactivate the filter click on CLEAR Figure 7 13 Filter dicom Filter Apply ...

Страница 306: ...blem should be investigated by R D It is not intended to replace or enhance the existing Full Backup functionality NOTE The Full Backup created by Alt D was pressed The main use is when R D or OLC need detailed information about the system e g when experiencing strange behaviour or when the problem shoud be investigated by R D It is intended to replace or enhance the existing Full Backup functiona...

Страница 307: ...RVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 15 3 Select the destination of the System state backup 4 Select the NEXT button to start the backup process After saving the data the Voluson P8 Voluson P6 reboots and the application starts again Figure 7 15 System Configuration Save ...

Страница 308: ... screen select ADMINISTRATION and then click the SERVICE tab 6 Click the SERVICE TOOLS button on the Service window to get access to the Service Tools page 7 Click on the EXPORT SYSTEM DATA button to Full Backup the System State This includes dump files see Section 7 4 3 2 1 and text files full Service Database informations about probes boards Software Options and the Event Log File 7 4 3 2 1 Dump...

Страница 309: ... the right trackball key right click 3 Select Connect Clinical Lifeline see Figure 7 17 below This activates Disruptive Mode and VCO for the application OLC to quickly assist the customer 4 If you select Cancel Disruptive Mode and VCO is turned OFF NOTE Visual information about the GE network status is shown in the status area on the bottom of the screen NOTE If the customer does not wish to have ...

Страница 310: ...ng VNC 1 In the Total Access Session window see Figure 7 26 on page 7 24 find the item labeled UL_VNC and click on the Connect button Within 1 minute the Proxy Config window opens 2 If available and required enter the correct proxy server address and then click OK NOTICE NOTICE The system may require a correct proxy configuration Please ask the customer Figure 7 18 Proxy Config ...

Страница 311: ...d on your remote workstaticcreen The tool bar on top of the virtual monitor screen contains useful options Figure 7 19 virtual Monitor Screen of Voluson P8 Voluson P6 Figure 7 20 Tool bar on top of virtual screen 1 Send Ctrl Alt Del Taskmanager to host 2 Toggle Fullscreen Mode On Off 3 Show Connection Options 4 Refresh Screen 5 Send Start Ctrl ESC Windows Start Menu to host 6 Send Custom Key 7 Sho...

Страница 312: ...ing the Virtual Network Computing VNC feature TIP Frequently used shortcut functions to release the mouse cursor press the shortcut CTRL M Mode Control On Off Toggle function to display the system setup press the shortcut ALT SHIFT S to display the Console Emulator see Figure 7 31 on page 7 27 press the shortcut ALT S NOTE For complete shortcut list refer to Section 7 2 2 Shortcuts List on page 7 ...

Страница 313: ... to display the Console Emulator NOTE For description of keys and their functions refer to Section 5 6 2 OPIO layout on page 5 32 NOTE To regain free sight to the monitor image the virtual Voluson P8 Voluson P6 console disappears as soon as you select a button It is simply hidden in the background Press the shortcut ALT S twice to show it again Figure 7 22 Voluson P8 Voluson P6 Console Emulator ...

Страница 314: ...the Proxy Config window opens NOTE The system may require a correnct proxy configuration Please ask the customer 2 If available and required enter the correct proxy server address from your location and then click OK 3 The next window shows the command to be issued click on OK to grant the request 4 Select GE Service from the pull down menu enter the password changes every six months and then clic...

Страница 315: ... system As soon as the dongle is detected the login window pops up 4 Enter the Standard GE Revolving password changes every six months and then click ENABLE 5 On the left side of the screen select ADMINISTRATION and then click the SERVICE tab 6 Click the SERVICE TOOLS button on the Service window to get access to the Service Tools page 7 Activate the Auto Tester program by clicking START The follo...

Страница 316: ...AutotesterScripts 13 Press Alt D to export system data and the needed logs Operation see NOTE he standalone recorded Auto Tester file makes only sense with the Alt D log To analyze the workflow the exact date and time of occurrence needs to be documented 14 If desired save the script file also on DVD CD Therefore Insert an empty DVD CD R RW in the Drive and select the Burn on CD icon Enter a Filen...

Страница 317: ... or the field engineer Which probe is in use and which probes are connected Which Mode is activated Which peripherals are connected Dicom Printer etc Collected information from all steps above exported system data and needed logs Alt D can be sent to the Make Center With this information the Make Center can see how the customer is using the system and reproduce potential failures NOTE Providing al...

Страница 318: ...system As soon as the dongle is detected the login window pops up 4 Enter the Standard GE Revolving password changes every six months and the click ENABLE 5 On the left side of the screen select ADMINISTRATION and then click the SERVICE tab 6 Click the SERVICE TOOLS button on the Service window to get access to the Service Tools page 7 Click the HARDWARE TEST button 8 Confirm the displayed Warning...

Страница 319: ...IRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 27 Automatic Not Automatic diagnostic tests are executed from this System Diagnostics Tool Figure 7 30 System Diagnostics Tool ...

Страница 320: ...ection 7 7 Hardware Tests 7 7 1 Audio System Interface Test Test of the Audio path from system to Speaker and AudioOutput can be performed 1 Select the AUDIO SYSTEM INTERFACE TEST button in the Diagnostics Tool window 2 Check if there is hearable audio sound from Speaker and AudioOutput port on rear panel ...

Страница 321: ...wn as dark Ultrasound Lines NOTE To distinguish if it is a board problem or if the probe is defective it would be helpful to have an additional probe on site 7 7 2 1 Check for missing channels or elements 1 Connect a probe and then press the FREEZE key to run the live scan Write Mode 2 Select the APERTUR LATERAL MIN button in the Diagnostics Tool window Missing channels or elements can easily be d...

Страница 322: ...gnostics Tool its execution log message will be displayed in Log Window and saved also in log file located in D Export diagnosis txt with execution time stamp Overall Test like DRFM Overall Test means that it executes sub diagnostic test items one by one automatically Table 7 1 below outlines Voluson P8 Voluson P6 Diagnostics Tool described in the sub sections 7 7 3 1 System Overall Test System Ov...

Страница 323: ...and displays the part of VPD information and the test result as PASSED FAILED 7 7 3 2 4 DPS It verifies VPD information embedded in EEPROM on DPS board and displays the part of VPD information and the test result as PASSED FAILED 7 7 3 2 5 DC4D It verifies VPD information embedded in EEPROM on DC4D and displays the part of VPD information and the test result as PASSED FAILED 7 7 3 2 6 MDM It verif...

Страница 324: ...e measure on DRFM and displays the current temperature status and the test result as PASSED FAILED 7 7 3 3 3 IIC DRFM Interface It verifies I2C path named I2C_Local by using EEPROM on DRFM board and displays the test result as PASSED FAILED 7 7 3 3 4 HW ID Config It verifies HW ID configuration on DRFM board and displays the current HW ID and the test result as PASSED FAILED 7 7 3 3 5 FPGA DRFM In...

Страница 325: ... DRFM and DBM and displays the current passed GL units and the test result as PASSED FAILED 7 7 3 4 5 ADC Analog Digital Converter It verifies ADC Interface between DBM and AFE on DBM board and displays the current passed total ADC number and the test result as PASSED FAILED 7 7 3 4 6 SignalGen SignalGen is called by ASP Analog signal Path It verifies from AFE to DRFM by generating test wave signa...

Страница 326: ...escribed in the sub sections 7 7 3 6 1 Voltage DPS It verifies Each Power Voltage Levels on DPS board and displays the current voltage levels and the test result as PASSED FAILED 7 7 3 6 2 ADM Interface It verifies the interface to ADM1062 chip on DPS board by reading writing test data and displays the test result as PASSED FAILED 7 7 3 6 3 HV Linearity It verifies the linearity of HVA and HVB vol...

Страница 327: ...iagnostic test performs the self diagnosis of battery and battery control board status Table 7 11 below outlines Voluson P8 Voluson P6 Diagnostics Tool described in the sub sections 7 7 3 8 1 Battery Interface It veriftes whether installed Battery and verifies I2C path by using EEPROM on Battery Pack and the test result as PASSED FAILED 7 7 3 8 2 Battery Status It verifies Battery Status Temperatu...

Страница 328: ...on testing 7 7 3 9 1 Aperture Lateral std It verifies system displays single aperture scan image on off 7 7 3 9 2 Audio Interface This diagnostics test performs if there is hearable audio sounding from speaker when executing this test 7 7 3 9 3 USB Loopback This diagnostics test performs check usb port status by plugged usb loopback 7 7 3 9 4 UVC HDMI Test It verifies HDMI connection between UVC d...

Страница 329: ...g startup the application can be started even if the ultrasound hardware is broken The hardware is not initialized the application simply does not get ultrasound data into process However one can use the Image Archive function to save patient archive and images and to gather system information by saving a Full Backup The remote service capabilities are also present Before that some BIOS value shou...

Страница 330: ...N 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 7 38 Section 7 8 Minimum Configuration to Boot Scan 4 Set to Disable In case of Skylake Com Express Move to Advanced PCI Express Configuration PCI Express Port 0 3 5 Set to Disable ...

Страница 331: ...me Ivybridge There is no Codename field NOTE BIOS value to be changed again after simulation mode for normal scan mode as following procedure 1 Turn the system on and press Del key to enter the BIOS Menu 2 Enter Password polaris and Press Enter key 3 Press F9 and select Yes 4 Press F10 and select Yes to save change Operation 1 Power On Boot Up the system as described in Section 3 6 2 Power On Boot...

Страница 332: ...und information 6 Backup all images see Section 4 5 6 Archiving Images on page 4 21 7 Save a Full Backup as described in Section 4 5 3 Save Full System Configuration Full Backup on page 4 16 or press the ALT D key to gather detailed information and data from the system see Section 7 4 3 1 Export System Data by pressing the ALT D key on page 7 14 8 Press the F2 key on the control panel and select S...

Страница 333: ...ow pops up 2 Enter the Standard GE Revolving password password changes every six months and then click ENABLE 3 Start a probe 4 Press the F2 key on the control panel 5 In the Utilities menu select SYSTEM SETUP to invoke the setup desktop on the screen 6 On the left side of the screen select ADMINISTRATION and then click the SERVICE tab 7 In the displayed window select the SERVICE TOOLS button The ...

Страница 334: ... is shown in the Status window on the screen LED TEST Select the LED TEST button All LEDs will be switched on off First click All LEDs changes to green Second click All LEDs go off Third click All LEDs changes to orange Forth click All LEDs go off The CLEAR LED S button sets all LED s off The illumination of all hardkeys is switched off Hardkey Test Press each hardkey Indication key code number of...

Страница 335: ...8 Voluson P6 troubleshooting trees and instructions shown in the sub sections Table 7 8 Troubleshooting Trees and Instructions Sub section Description Page Number 7 10 1 System does not boot up 7 50 7 10 2 Noise disturbs the Image 7 51 7 10 3 Trackball Impaired Sensitivity 7 52 7 10 4 Printer Malfunction 7 53 7 10 5 Monitor Troubleshooting 7 54 7 10 6 DVD CD R RW Drive Test 7 55 7 10 7 Network Tro...

Страница 336: ...by Switch illuminated Press Standby Switch till it changes to green System starts booting Is Standby Switch orange Is there any picture on the monitor Other error Please contack the service engineer Is there any picture on the monitor Power Cord is connected to Mains Power Switch is turned on on the backside of the system Verify that Press Standby Switch for approximately 5 sec untill system is do...

Страница 337: ... Chapter 7 Diagnostics Troubleshooting 7 45 7 10 2 Noise disturbs the Image Figure 7 39 Noise disturbs the Image Troubleshooting Noise is shown in image YES NO Is there any electrical device that affect the System Keep the System away from those electrical devices use another wallsocket Check probes ...

Страница 338: ...ball has low sensitivity NO YES Is it difficult to roll the trackball Remove the trackball and clean it from dust Remove the trackball and clean it from dust Mount the outer fixing ring by turning it clockwise Close Return to normal scanning Contact GE service representative to replace the trackball module Set up the appropriate trackball response time via Trackball Speed button in the System Setu...

Страница 339: ...GE DRAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 47 7 10 4 Printer Malfunction Figure 7 41 Printer Malfunction ...

Страница 340: ...ooting For further details refer to Section 6 3 LCD Monitor Adjustment on page 6 2 Fault symptom Check these items No image No signal Check the HDMI and Power cable are properly connected to monitor Check the HDMI and Power cable are properly connected to DRFM Check no pins of the HDMI and Power cable are bent Video test pattern are not clear bright parallel or square Replace the monitor ...

Страница 341: ... DVD R RW and CD R W medias are available 1 Insert an empty DVD CD R RW into the Drive 2 Enter Patient Archive by pressing the PATIENT key on the control panel 3 On the left side of the screen select ARCHIVE as shown in Figure 7 51 above 4 If not already selected choose Source Local Archive Drive from the pop up menu 5 Select an Exam with images Figure 7 42 Patient Archive ARCHIVE ...

Страница 342: ...rt images of the selected exam to DVD CD R RW a Click the EXPORT button b If not already selected choose DVD CD drive from pop up menu c Enter File name d Select any Voluson Format 4dv from the pop up menu e Click the SAVE button 7 After successful export choose Source 4DV from the pop up menu 8 The images which you have chosen to export should be visible Figure 7 43 Export images to DVD CD as 4dv...

Страница 343: ...mote service connection The customer gets visual information about GE remote service status shown in the status area on the right bottom of the screen Gray Cable disconnected or no network signal on a connected cable Green Cable connected to a network Does not imply proper network settings Gray Idle State Remote access is inactive Yellow Access pending Remote Access connection is active but Disrup...

Страница 344: ... The Open File window appears 3 Select Drive in the Look In pull down menu 4 In the File of type field change file type to V730 Format v730 mdb 5 Browse for the folder where SonoView Backup was stored 6 Select file v730 mdb and then click OPEN Table 7 9 Technical Tips Sub section Description Page Number 7 10 8 1 Storing SonoView images to Voluson P8 Voluson P6 7 58 Issue Storing SonoView images fr...

Страница 345: ...ages NOTICE NOTICE If the problem error message still exists after performing the described actions call technical support Index Error Message Reason Action s 01 When exceeds 15watt Protection limit 1 See error message 2 Activated probe deselect 02 Restart the system 1 See error message 2 Stop scanning 03 Restart the system 1 See error message 2 Restart the system ...

Страница 346: ...ERVICE MANUAL 7 54 Section 7 11 Error Messages 04 Restart the system 1 See error message 2 Stop scanning 05 Restart the system 1 See error message 2 Shut down the system within 10 minutes 06 Restart the system 1 See error message 2 Restart the system Index Error Message Reason Action s ...

Страница 347: ...GE DRAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 55 This page was intentionally left blank ...

Страница 348: ...GE RAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 7 56 Section 7 11 Error Messages ...

Страница 349: ...pment and 5 years for the probes may be expected The Waste of Electrical and Electronic Equipment WEEE must not be disposed as unsorted municipal waste and must be collected separately Please contact the manufacturer or other authorized disposal company for information concerning the decommission of your equipment Table 8 1 Chapter 8 Contents Section Description Page Number 8 1 Overview 8 1 8 2 Sy...

Страница 350: ... 8 28 Replacement of the NEST 8 93 8 29 Replacement of the AIR FILTER 8 98 8 30 Replacement of the Probe Holder KIT 8 100 8 31 Replacement of SIDE COVER L ASSY BT16 8 101 8 32 Replacement of SIDE COVER R ASSY 8 103 8 33 Replacement of the Trackball 8 106 8 34 Replacement of the TGC BOARD 8 111 8 35 Replacement of AN KEYBOARD ASSY 8 113 8 36 Replacement of the Caster Cap BT16 8 115 8 37 Replacement...

Страница 351: ...om a clinic or site for example body coils or and ultrasound probe The purpose of the regulation is to protect employees in the transportation industry as well as the people who will receive or open this package The US Department of Transportation DOT has ruled that items what were saturated and or dripping with human blood that are now caked with dried blood or which were used or intended for use...

Страница 352: ...Ultrasound Application Software Service Software EUM MS patches etc and installation time depend on contents of the SW update package PRECONDITION InSite connection has to be configured and checked out Manpower One person 1 hour depends on contents of System DVD resp SW update package peripherals etc Tools System DVD Under the top cabinet cover inside the Voluson P8 Voluson P6 or active InSite con...

Страница 353: ...image Receiver Frequency Use the FREQUENCY control to switch the Receiver Frequency range penet norm resol no disturbances in the 2D image during changing the Receiver Frequency M Mode Start the Abdomen Program adjust the M Cursor vessel and activate the M Mode Adjust the SPEED key to change the M Mode sweep speed After FREEZE move the TRACKBALL to recall the stored sequence the M Cursor agree wit...

Страница 354: ...he screen 3 On the left side of the screen select ADMINISTRATION and then click the SYSTEM INFO tab 4 Check the currently installed Software Hardware version of the Voluson P8 Voluson P6 system 5 Select the Option tab to see which options are currently installed Please print out the options page or write down down the state of the options P permanent I inactive Figure 8 1 Version check System Setu...

Страница 355: ...n and the system is updated via the FMI from DVD button a new software specific Permanent key is required Calculated in OKOS Option Key Order System http 3 187 187 9 OKOS For detailed description see Section 8 2 1 3 on page 8 7 Figure 8 2 System Setup Administration Option page 1 D Demo Option is activated for demo and expires at date shown in the Valid column I Inactive Option is not activated P ...

Страница 356: ...OTE For more detailed information about FMI from DVD refer to 5 14 3 2 1 on page 5 51 1 Perform Preparations as described in 8 2 1 on page 8 4 2 If not already done disconnect all external USB devices except DVD CD R RW drive 3 Insert the System DVD into the drive 4 Restart the system Turn system OFF and then back ON 5 Press the Utilities key on the control console 6 In the Utilities menu select S...

Страница 357: ...e from C is stored on R After executing all LINUX commands the system reboots again NOTE If the currently installed software has to be upgraded to a newer version a new software specific Permanent key is required 12 Enter the appropriate Permanent Key calculated in OKOS select OK and confirm with ENTER If the entered Permanent key is correct following window appears confirm with ENTER Figure 8 4 V...

Страница 358: ...ICE MANUAL 8 10 Section 8 2 System Software Installation Upgrade Procedure 13 Check disk is performed automatically restart 14 3 dots one after the other appear on the screen see Figure 8 8 below 15 Booting auto Figure 8 7 Check disk is performed automatically Figure 8 8 Boot screen auto ...

Страница 359: ...OTE An automated process was developed to install the required software parts perform check disk remap driveletters and match settings During this process the system might reboot several times 19 Please wait until all processes are finished Finally the 2D screen is displayed on the monitor 20 Remove the System DVD from the DVD CD R RW Drive drive NOTE If the BT warning dialog appears enter the pro...

Страница 360: ...r Remote Control selection in the System Setup Connectivity BUTTON CONFIGURATION page 24 Confirm that the date and time are set correctly and that the Windows automatic DST feature is ON 25 Perform Software and Functional checks as described in Section 8 3 on page 8 20 8 2 3 Software DVD Storage 1 Disassemble the Top Cabinet Cover 2 Put Software DVD in the jacket 3 Re assemble the Top Cabinet Cove...

Страница 361: ... User Programs 3D 4D Programs and Auto Text remain unaffected Afterwards the new software is configured such that it is integrated again in its environment 8 2 4 2 Update Procedure 1 Remove any CD DVD from the DVD drive 2 Perform Preparations as described in Section 8 2 1 1 3 If not already done disconnect all external USB devices except DVD drive 4 Press the ON OFF button on the control console 5...

Страница 362: ...ware Installation Upgrade Procedure 7 When downloading is finished click Install NOTE Installation may take up to one hour DO NOT interrupt the installation If you want to install the new software later click Exit Installation can be resumed by clicking Install in the Shutdown window Figure 8 14 Download finished click Install ...

Страница 363: ...cedure starts The system saves Full Backup in silent mode on R then it reboots into LINUX 9 Check disk is performed automatically restart A silent Rollback image from C is stored on R After executing all LINUX commands the system reboots again 10 3 dots one after the other appear on the screen Figure 8 15 Installation in progress Figure 8 16 Check disk is performed automatically ...

Страница 364: ... RAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 8 16 Section 8 2 System Software Installation Upgrade Procedure 11 Booting auto Figure 8 17 Boot screen auto ...

Страница 365: ...018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 17 12 Updating will take some time First the image is saved to R Saving NTSF to image then the image is restored from R Restoring NTFS from image 13 Please wait until all processes are finished 100 percent completed ...

Страница 366: ...e Installation Upgrade Procedure 14 The system is rebooting into windows Boot screen Voluson NOTE An automated process was developed to install the required software parts perform check disk remap drive letters and match settings During this process the system might reboot several times Figure 8 18 Automatic processes are running ...

Страница 367: ...Remote Control selection in the System Setup Connectivity Button Configuration page 18 Confirm date and time setting in the System Setup General Settings General Settings page 19 Perform a check of all modes and features listed Move the cursor over the feature name to get information how to check Section 3 7 Printer Installation on page 3 32 20 When all features are OK enter your signature and the...

Страница 368: ... the Application Software version 7 Check that the Tune version of the Application presets match the Application Software version 8 Perform basic functional checks to ensure system is functioning normally 9 Check Service Connectivity if required perform InSite Checkout i e Service Platform has to be configured properly Figure 8 20 System Setup Administration SYSTEM INFO page NOTICE NOTICE It is ne...

Страница 369: ... ___________________________ ___ Passed ___ Failed Demo images were successfully imported Optional ___ Passed ___ Failed ___ N A Function Check A Power ON Boot up ___ Passed ___ Failed Mode Transitions ___ Passed ___ Failed B Flow Optional ___ Passed ___ Failed ___ N A CW Optional ___ Passed ___ Failed ___ N A 4D Optional ___ Passed ___ Failed ___ N A Power Shut Down ___ Passed ___ Failed System S...

Страница 370: ...3D 4D Programs Auto Text entries gray curves and complete System settings such as language time date format etc Measure Configuration user specific measure setup settings Patient Archive database containing patient demographic data and measurements no images Options Permanent Key that is specific for enabled software options and Demo Key Image Transfer Configuration DICOM settings e g DICOM server...

Страница 371: ...Can be used with 3D 4D Activation Table 8 4 Software options Options VP6 VP8 3D 4D Activation Option a Option a Advanced 3D 4D Package Option b Option b HDliveTM 3D Not Available Option c Advanced VCI Not Available Option c SingleView Option c Option c SonoAVC follicle Pro Option c Option c SonoL D Option Option TUI Option c Option c CFMM CFM Option Standard XTD Not Available Option Anatomical M M...

Страница 372: ... the left side of the screen select ADMINISTRATION and then click the OPTIONS tab see Figure 8 2 on page 8 7 8 7 1 1 Operation for installing a Demo Key or a Permanent Key 1 Position the cursor inside the input field desired and press the right left trackball key 2 If one exists clear edit the current key code 3 Enter the encrypted serial code with the keyboard and click on SUBMIT Code will be che...

Страница 373: ...f FRONT CABINET BEZEL ASSY BT16 8 8 1 Manpower 1 person 1 minute 8 8 2 Tools Not required 8 8 3 Disassembly Procedure 1 Remove the FRONT CABINET BEZEL ASSY BT16 Refer to the figure below 8 8 4 Installation Procedure Install the new parts in the reverse order of removal procedure Figure 8 21 Removing the FRONT CABINET BEZEL ASSY BT16 FRONT CABINET BEZEL ASSY BT16 ...

Страница 374: ...ent of ASSY BEZEL FRONT DVD PLASTIC BT16 8 9 1 Manpower 1 people 1 minute 8 9 2 Tools Not required 8 9 3 Disassembly Procedure 1 Remove the ASSY BEZEL FRONT DVD PLASTIC BT16 Refer to the figure below 8 9 4 Installation Procedure Install the new parts in the reverse order of removal procedure Figure 8 22 Removing the ASSY BEZEL FRONT DVD PLASTIC BT16 ASSY BEZEL FRONT DVD PLASTIC BT16 ...

Страница 375: ...driver 8 10 3 Disassembly Procedure 1 Remove the FRONT PRINTER BEZEL ASSY BT16 Refer to the figure below 2 Unscrew 2 screws and remove the PRINTER DUMMY COVER ASSY BT16 Refer to the figure below 8 10 4 Installation Procedure Install the new parts in the reverse order of removal procedure Figure 8 23 Removing the FRONT PRINTER BEZEL ASSY BT16 Figure 8 24 Removing the Printer Dummy Cover Assy BT16 F...

Страница 376: ...Plastic 8 11 1 Manpower 1 people 3 minutes 8 11 2 Tools Not required 8 11 3 Disassembly Procedure 1 Remove the FRONT CABINET BEZEL ASSY BT16 Refer to the figure below 2 Remove the FRONT PRINTER BEZEL ASSY BT16 Refer to the figure below Figure 8 25 Removing the FRONT CABINET BEZEL ASSY BT16 Figure 8 26 Removing the FRONT PRINTER BEZEL ASSY BT16 FRONT CABINET BEZEL ASSY BT16 FRONT PRINTER BEZEL ASSY...

Страница 377: ...IC SERVICE MANUAL Chapter 8 Replacement Procedures 8 29 3 Remove the TOP CABINET COVER BT16 Refer to the figure below 8 11 4 Installation Procedure Install the new parts in the reverse order of removal procedure Figure 8 27 Removing the TOP CABINET COVER BT16 TOP CABINET COVER BT16 ...

Страница 378: ...STAL COVER BT16 8 12 1 Manpower 1 people 5 minutes 8 12 2 Tools Philips screw driver 8 12 3 Disassembly Procedure 1 Remove the FRONT PRINTER BEZEL ASSY BT16 Refer to the figure below 2 Remove the TOP CABINET COVER 16 Refer to the figure below Figure 8 28 Removing the FRONT PRINTER BEZEL ASSY BT16 Figure 8 29 Removing theTOP CABINET COVER BT16 FRONT PRINTER BEZEL ASSY BT16 TOP CABINET COVER BT16 ...

Страница 379: ...VER BT16 Refer to the figure below 4 Unscrew 2 screws and remove the Cover Front Pedestal Refer to the figure below 8 12 4 Installation Procedure Install the new parts in the reverse order of removal procedure Figure 8 30 Removing the BOTTOM PEDESTAL COVER BT16 Figure 8 31 Removing the FRONT PEDESTAL COVER BT16 BOTTOM PEDESTAL COVER BT16 FRONT PEDESTAL COVER BT16 ...

Страница 380: ...ABINET PLASTIC BT16 8 13 1 Manpower 1 people 5 minutes 8 13 2 Tools Not required 8 13 3 Disassembly Procedure 1 Remove the TRAY PEDESTAL CABINET PLASTIC BT16 Refer to the figure below 2 Remove the two Self Adhesive Loops Refer to the figure below Figure 8 32 Removing the TRAY PEDESTAL CABINET PLASTIC BT16 Figure 8 33 Removing the two Self Adhesive Loops TRAY PEDESTAL CABINET PLASTIC BT16 Self Adhe...

Страница 381: ...emove the Paper and place the TRAY PEDESTAL CABINET PLASTIC on the System Pedestal Refer to the figure below Figure 8 34 Installing the Self Adhesive Hook and Loop set Figure 8 35 Removing the Paper and placing the TRAY PEDESTAL CABINET PLASTIC BT16 Stick the self adhesive hook and loop set on the TRAY PEDESTAL CABINET PLASTIC BT16 Align it with the locator pillars and stick at center Place the TR...

Страница 382: ...eplacement of REAR CABINET BEZEL ASSY BT16 8 14 1 Manpower 1 person 1 minute 8 14 2 Tools Not required 8 14 3 Disassembly Procedure 1 Remove the Rear Cabinet BEZEL ASSY BT16 Refer to the figure below 8 14 4 Installation Procedure Install the new parts in the reverse order of removal procedure Figure 8 36 Removing the REAR CABINET BEZEL ASSY BT16 REAR CABINET BEZEL ASSY BT16 ...

Страница 383: ...tion 8 15 Replacement of ENCODER set 8 15 1 Manpower One person 40 minutes 8 15 2 Tools Philips screw driver spanner 8 15 3 Removal procedure 1 Unscrew 9 screws at the bottom of the keyboard body Refer to the figures below 2 Remove 5 probe holders Refer to the figure below Figure 8 37 Unscrewing 9 screws Figure 8 38 Removing 5 probe holders ...

Страница 384: ...RAFT MAY 23 2018 BASIC SERVICE MANUAL 8 36 Section 8 15 Replacement of ENCODER set 3 Lift up main keyboard assy Refer to the figure below 4 Disconnect 4 cables Refer to the figure below Figure 8 39 Lifting up main keyboard assy Figure 8 40 Disconnecting 4 cables ...

Страница 385: ...NUAL Chapter 8 Replacement Procedures 8 37 5 Pull 10 encoder key caps out Refer to the figure below 6 Unscrew 2 screws on trackball and disconnect the cable to remove the trackball Refer to the figures below Figure 8 41 Pulling 10 encoder key caps out Figure 8 42 Unscrewing 2 screws on trackball ...

Страница 386: ... 38 Section 8 15 Replacement of ENCODER set 7 After removing trackball unscrew 21 screws and disconnect TGC cable And lift up the main keyboard circuit board Refer to the figure below Figure 8 43 Lifting up the main keyboard circuit board after unscrew 21 screws and disconnecting TGC cable TCG cable ...

Страница 387: ...coder to be replaced And replace the encoder Refer to the figures below 9 Pull the encoder out from the main keyboard circuit board Refer to the figure below 8 15 4 Installation Procedure Install a new part in the reverse order of removal procedure Figure 8 44 Unscrewing the nut of encoder to be replaced Figure 8 45 Pulling the encoder out from the main keyboard circuit board ...

Страница 388: ... Replacement of Power ON OFF key PCB Assembly Section 8 16 Replacement of Power ON OFF key PCB Assembly 8 16 1 Manpower One person 30 minutes 8 16 2 Tools Philips screw driver 8 16 3 Disassembly procedure 1 Unscrew 9 screws at the bottom of the keyboard body Refer to the figures below Figure 8 46 Unscrewing 9 screws ...

Страница 389: ...DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 41 2 Remove 5 probe holders Refer to the figure below 3 Lift up main keyboard assy Refer to the figure below Figure 8 47 Removing 5 probe holders Figure 8 48 Lifting up main keyboard assy ...

Страница 390: ...NUAL 8 42 Section 8 16 Replacement of Power ON OFF key PCB Assembly 4 Disconnect 4 cables under the main keyboard assy Refer to the figure below 5 Unscrew 2 screws on the system body Refer to the figure below Figure 8 49 Disconnecting 4 cables Figure 8 50 Unscrewing 2 screws on the system body ...

Страница 391: ...pter 8 Replacement Procedures 8 43 6 Remove 2 RUBBER CAPS on SPEAKER COVER LEFT RIGHT 7 Unscrew 2 screws and then remove the SPEAKER COVER LEFT RIGHT Refer to figure below Figure 8 51 Assembling 2 RUBBER CAPS on SPEAKER COVER LEFT RIGHT Figure 8 52 Unscrewing 2 screws to remove SPEAKER COVER LEFT RIGHT ...

Страница 392: ...UAL 8 44 Section 8 16 Replacement of Power ON OFF key PCB Assembly 8 Unscrew 2 screws and remove the UPPER KEYBOARD ASSY 9 Unscrew 7 screws then remove metal cover and USB connection circuit board Refer to the figure below Figure 8 53 Removing Upper keyboard Assy Figure 8 54 Unscrewing 7 screws ...

Страница 393: ... MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 45 10 Remove the On Off key circuit Refer to the figures below 8 16 4 Installation procedure Install a new part in the reverse order of removal procedure Figure 8 55 Removing the On Off key circuit ...

Страница 394: ...ction 8 17 Replacement of Keyboard Assembly TOP Section 8 17 Replacement of Keyboard Assembly TOP 8 17 1 Manpower One person 25 minutes 8 17 2 Tools Philips screwdriver 8 17 3 Disassembly procedure 1 Unscrew 9 screws at the bottom of the keyboard body Refer to the figures below Figure 8 56 Unscrew 9 screws ...

Страница 395: ...DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 47 2 Remove 5 probe holders Refer to the figure below 3 Lift up main keyboard assy Refer to the figure below Figure 8 57 Removing 5 probe holders Figure 8 58 Lifting up main keyboard assy ...

Страница 396: ...8 17 Replacement of Keyboard Assembly TOP 4 Disconnect 4 cables under the main keyboard assy Refer to the figure below 5 Unscrew 2 screws and remove cable tie on the system body Refer to the figure below Figure 8 59 Disconnecting 4 cables Figure 8 60 Unscrewing 2 screws and removing cable tie on the system body ...

Страница 397: ...cement Procedures 8 49 6 Remove 2 RUBBER CAPS on SPEAKER COVER LEFT RIGHT Refer to the figure below 7 Unscrew 2 screws and then remove the SPEAKER COVER LEFT RIGHT Refer to figure below Figure 8 61 Removing 2 RUBBER CAPs on SPEAKER COVER LEFT RIGHT Figure 8 62 Unscrewing 2 screws to remove SPEAKER COVER LEFT RIGHT ...

Страница 398: ...ASIC SERVICE MANUAL 8 50 Section 8 17 Replacement of Keyboard Assembly TOP 8 Unscrew 2 screws and pick up the sub keyboard Refer to the figures below 8 17 4 Installation procedure Install a new part in the reverse order of removal procedure Figure 8 63 Picking up the sub keyboard ...

Страница 399: ...NUAL Chapter 8 Replacement Procedures 8 51 Section 8 18 Replacement of OPIO BUTTON SET 8 18 1 Manpower One person 40 minutes 8 18 2 Tools Philips screwdriver 8 18 3 Removal Procedure 1 Unscrew 9 screws at the bottom of the keyboard body Refer to the figures below Figure 8 64 Unscrewing 9 screws ...

Страница 400: ...MAY 23 2018 BASIC SERVICE MANUAL 8 52 Section 8 18 Replacement of OPIO BUTTON SET 2 Remove 5 probe holders Refer to the figure below 3 Lift up main keyboard assy Refer to the figure below Figure 8 65 Removing 5 probe holders Figure 8 66 Lifting up main keyboard assy ...

Страница 401: ...VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 53 4 Disconnect 4 cables Refer to the figure below Figure 8 67 Disconnecting 4 cables ...

Страница 402: ...8 54 Section 8 18 Replacement of OPIO BUTTON SET 5 Pull 10 encoder key caps out Refer to the figure below 6 Unscrew 2 screws on trackball and disconnect the cable to remove the trackball Refer to the figures below Figure 8 68 Pulling 10 encoder key caps out Figure 8 69 Unscrew 2 screws on trackball ...

Страница 403: ...d disconnect TGC calbe And lift up the main keyboard circuit board Refer to the figure below 8 Remove the button to be replaced Refer to the figure below 8 18 4 Installation Procedure Install a new part in the reverse order of removal procedure Figure 8 70 Lifting up the main keyboard circuit board after unscrewing 21 screws and disconnecting TGC cable Figure 8 71 Removing the button ...

Страница 404: ... Replacement of OPIO Plastic lower cover assy Section 8 19 Replacement of OPIO Plastic lower cover assy 8 19 1 Manpower One person 40 minutes 8 19 2 Tools Philips screwdriver 8 19 3 Disassembly procedure 1 Unscrew 9 screws at the bottom of the keyboard body Refer to the figures below Figure 8 72 Unscrewing 9 screws ...

Страница 405: ...DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 57 2 Remove 5 probe holders Refer to the figure below 3 Lift up main keyboard assy Refer to the figure below Figure 8 73 Removing 5 probe holders Figure 8 74 Lifting up main keyboard assy ...

Страница 406: ...LUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 8 58 Section 8 19 Replacement of OPIO Plastic lower cover assy 4 Disconnect 4 cables Refer to the figure below Figure 8 75 Disconnecting 4 cables ...

Страница 407: ...t Procedures 8 59 5 Pull 10 encoder key caps and 8 TGC control key caps out Refer to the figure below 6 Disconnect A N key cable and unscrew 10 screws of A N key Refer to the figures below Figure 8 76 Pulling 10 encoder key caps and 8 TGC control key caps out Figure 8 77 Disconnecting A N key cable and unscrew 10 screws ...

Страница 408: ...er disconnecting the cable Refer to the figures below 8 Pull out the circuit board Refer to the figure below 9 Unscrew 2 screws on trackball and disconnect the cable to remove the trackball Then remove trackball Refer to the figure below Figure 8 78 Unscrewing 4 screws of TGC control circuit board Figure 8 79 Pulling out the circuit board Figure 8 80 Unscrewing 2 screws on trackball and disconnect...

Страница 409: ...E MANUAL Chapter 8 Replacement Procedures 8 61 10 Unscrew 21 screws from main keyboard Then pull the main circuit board out Refer to the figure below 8 19 4 Installation procedure Install a new part in the reverse order of removal procedure Figure 8 81 Unscrewing 21 screws from main keyboard ...

Страница 410: ... Replacement of OPIO Plastic upper cover assy Section 8 20 Replacement of OPIO Plastic upper cover assy 8 20 1 Manpower One person 30 minutes 8 20 2 Tools Philips screwdriver 8 20 3 Disassembly procedure 1 Unscrew 9 screws at the bottom of the keyboard body Refer to the figures below Figure 8 82 Unscrewing 9 screws ...

Страница 411: ...DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 63 2 Remove 5 probe holders Refer to the figure below 3 Lift up main keyboard assy Refer to the figure below Figure 8 83 Removing 5 probe holders Figure 8 84 Lifting up main keyboard assy ...

Страница 412: ...NUAL 8 64 Section 8 20 Replacement of OPIO Plastic upper cover assy 4 Disconnect 4 cables under the main keyboard assy Refer to the figure below 5 Unscrew 2 screws on the system body Refer to the figure below Figure 8 85 Disconnecting 4 cables Figure 8 86 Unscrewing 2 screws on the system body ...

Страница 413: ... 65 6 Remove 2 RUBBER CAPS on SPEAKER COVER LEFT RIGHT 7 Unscrew 2 screws and then remove the SPEAKER COVER LEFT RIGHT Refer to the figures below 8 Unscrew 2 screws and remove the upper keyboard assy Figure 8 87 Removing the RUBBER CAPS Figure 8 88 Unscrewing 2 screws to remove SPEAKER COVER LEFT RIGHT Figure 8 89 Unscrewing 2 screws ...

Страница 414: ...ction 8 20 Replacement of OPIO Plastic upper cover assy 9 Unscrew 7 screws then remove metal cover power On Off key assembly and USB connection circuir board Refer to the figure below 8 20 4 Installation procedure Install a new part in the reverse order of removal procedure Figure 8 90 Unscrewing 7 screws ...

Страница 415: ...res 8 67 Section 8 21 Replacement of Monitor 8 21 1 Manpower 2 people 10 minutes 8 21 2 Tools Philips screwdriver Allen Wrench 8 21 3 Removal Procedure 1 Unscrew 2 screws to remove the Monitor REAR CABINET BEZEL ASSY BT16 Refer to the figure below Figure 8 91 Unscrewing 2 screws to remove the Monitor REAR CABINET BEZEL ASSY BT16 ...

Страница 416: ...e MONITOR the other person should unscrew the 2 hexagonal socket screws and then separate the MONITOR for the LCD ARM Refer to the figure below 8 21 4 Installation Procedure 1 Install the new parts in the reverse order of removal 2 Perform FRU8 8 Replacement of Monitor Functional Tests Figure 8 92 Unscrewing 5 screws to disconnect the LCD cables Figure 8 93 Unscrewing 2 hexagonal screws to separat...

Страница 417: ...0 minutes 8 22 2 Tools Philips screwdriver Nipper Allen Wrench 8 22 3 Removal Procedure 1 Disassemble the LCD MONITOR Refer to the Section 8 21 Replacement of Monitor 2 Unscrew 1 screw to remove Monitor arm cover Refer to the figure below 3 Removing 2 RUBBER CAPS on SPEAKER COVER LFET RIGHT Figure 8 94 Removing the LCD Arm Cover Figure 8 95 Removing 2 RUBBER CAPS on SPEAKER COVER LEFT RIGHT MONITO...

Страница 418: ...hen remove SPEAKER COVER LEFT RIGHT Refer to the figure below 5 Unscrew 1 screw to remove Monitor Cable Fix bracket 6 Pull the LCD Cables through the LCD ARM and MONITOR HOLDER Figure 8 96 Unscrewing 2 screws to remove SPEAKER COVER LEFT RIGHT Figure 8 97 Removing the MONITOR CABLE FIX BRACKET Figure 8 98 Pulling out the LCD Cables from LCD ARM MONITOR CABLE FIX BRACKET ...

Страница 419: ...D ARM STOPPER SCREW Refer to figure below 8 Pull out the LCD ARM from the LCD ARM HOLDER 8 22 4 Installation Procedure 1 Install the new parts in the reverse order of removal 2 Perform FRU8 9 Replacement of the MONITOR STAND ARM Functional Tests Figure 8 99 Unscrewing LCD ARM STOPPER SCREW Figure 8 100 Pulling out the LCD ARM from the LCD ARM HOLDER ...

Страница 420: ...5 minutes 8 23 2 Tools Philips screw driver Flat blade screwdriver wrench Pincette Cutter 8 23 3 Disassembly Procedure 1 Unscrew 2 screws to remove the MONITOR REAR CABINET BEZEL ASSY BT16 Refer to the figure below 2 Unscrew 5 screws to disconnect the LCD Cables Figure 8 101 Unscrewing 2 screws to remove the MONITOR REAR CABINET BEZEL ASSY BT16 Figure 8 102 Unscrewing 5 screws to disconnect the LC...

Страница 421: ...ment Procedures 8 73 3 Unscrew 1 screw to remove MONITOR ARM COVER Refer to the figure below 4 Unscrew 1 screw to remove MONITOR CABLE FIX BRKT Refer to the figure below Figure 8 103 Unscrewing 1 screw to remove Monitor arm cover Figure 8 104 Unscrewing 1 screw to remove MONITOR CABLE FIX BRKT MONITOR CABLE FIX BRKT ...

Страница 422: ...LEFT RIGHT Refer to the figure below 6 Unscrew 2 screws and then remove SPEAKER COVER LEFT RIGHT Refer to the figure below 7 Unscrew 9 screws at the bottom of the OPIO Refer to the figure below Figure 8 105 Removing 2 RUBBER CAPS on SPEAKER COVER LEFT RIGHT Figure 8 106 Unscrewing 3 screw to remove SPEAKER COVER LEFT RIGHT Figure 8 107 Unscrewing 9 screws at the bottom of the OPIO ...

Страница 423: ...hapter 8 Replacement Procedures 8 75 8 Lift up the OPIO and disconnect Power OPIO 2 USB cables from the OPIO HUB PWA Refer to the figure below Figure 8 108 Disconnecting Power OPIO 2 USB cables KEYBOARD ASSY Insert power cable Host interface 10 pin connector OPIO cable USB connector 1 USB connector 2 ...

Страница 424: ...Remove the FRONT PRINTER BEZEL ASSY BT16 Refer to the figure below 10 Remove the Bottom Pedestal Cover BT16 Figure 8 109 Removing the FRONT PRINTER BEZEL ASSY BT16 Figure 8 110 Removing the TOP CABINET COVER BT16 Figure 8 111 Removing the Bottom Pedestal Cover BT16 FRONT PRINTER BEZEL ASSY BT16 TOP CABINET COVER BT16 Bottom Pedestal Cover BT16 ...

Страница 425: ...crew 2 screws and remove FRONT PEDESTAL COVER BT16 12 Untie the cable ties on keyboard base Refer to the figure below 13 Untie the cable clamps underneath the pedestal Refer to the figure below Figure 8 112 Removing the FRONT PEDESTAL COVER BT16 Figure 8 113 Untiing the cable ties Figure 8 114 Untiing the cable clamps FRONT PEDESTAL COVER BT16 ...

Страница 426: ...acement of LCD Cable 14 Unscrew 2 screws to remove POWER CORD FIX BRACKET Refer to the figure below 15 Disassemble power cord and REAR CABINET BEZEL ASSY BT16 Refer to the figure below Figure 8 115 Unscrewing 2 screws to remove POWER CORD FIX BRACKET Figure 8 116 Disassembling power cord and REAR CABINET BEZEL ASSY BT16 ...

Страница 427: ...DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 79 16 Unscrew 7 screws and disassemble REAR EMI COVER 17 Pull out the LCD Cables through the hole Refer to the figure below Figure 8 117 Unscrewing 7 screwing to disassemble REAR EMI COVER ...

Страница 428: ...SION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 8 80 Section 8 23 Replacement of LCD Cable 18 Disconnect Monitor cables 8 23 4 Assembly Procedure Install the new parts in the reverse order of disassembly procedure Figure 8 118 Disconnecting Monitor cables ...

Страница 429: ...npower One person 13 minutes 8 24 2 Tools Philips screwdriver 8 24 3 Disassembly Procedure 1 Remove the FRONT CABINET BEZEL ASSY BT16 Refer to the figure below 2 Remove the FRONT PRINTER BEZEL ASSY BT16 Refer to the figure below Figure 8 119 Removing the FRONT CABINET BEZEL ASSY BT16 Figure 8 120 Removing the FRONT PRINTER BEZEL ASSY BT16 FRONT PRINTER BEZEL ASSY BT16 FRONT PRINTER BEZEL ASSY BT16...

Страница 430: ... Replacement of ASSY Bumper Base Rubber 3 Remove the TOP CABINET COVER BT16 Refer to the figure below 4 Remove the REAR CABINET BEZEL ASSY BT16 Refer to the figure below Figure 8 121 Remove the TOP CABINET COVER BT16 Figure 8 122 Remove the REAR CABINET BEZEL ASSY TOP CABINET COVER BT16 REAR CABINET BEZEL ASSY BT16 ...

Страница 431: ... 8 83 5 Unscrew 3 screws and remove the RS Knobs Refer to the figure below 6 Unscrew 4 screws and remove the TRAY SIDE CABINET BT16 Refer to the figure below Figure 8 123 Removing the RS Knob Figure 8 124 Removing the TRAY SIDE CABINET BT16 Removing fisrt RS knob Removing second RS Knob Removing thrid RS Knob TRAY SIDE CABINET BT16 ...

Страница 432: ...t of ASSY Bumper Base Rubber 7 Unscrew 2 screws and remove the ASSY Left Side plastic Cover Refer to the figure below 8 Unscrew 2 screws and remove the ASSY RIGHT SIDE PLASTIC COVER BT16 Refer to the figure below Figure 8 125 Removing the ASSY Left Side Plastic Cover Figure 8 126 Removing the ASSY RIGHT SIDE PLASTIC COVER BT16 ...

Страница 433: ...IRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 85 9 Remove the COVER CASTOR NARROW L R Plastic Refer to the figure below Figure 8 127 Removing the COVER CASTOR NARROW L R Plastic ...

Страница 434: ...8 BASIC SERVICE MANUAL 8 86 Section 8 24 Replacement of ASSY Bumper Base Rubber 10 Remove the Bumper Base Rubber Refer to the figure below 8 24 4 Installation Procedure Install the new parts in the reverse order of removal procedure Figure 8 128 Removing the Bumper Base Rubber ...

Страница 435: ...ips screwdriver Nipper Allen Wrench 8 25 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 25 4 Removal Procedure 1 Remove 2 RUBBER CAPs on SPEAKER COVER LEFT RIGHT Refer to the figure below 2 Unscrew 2 screws and then remove SPEAKER Cover L R Refer to the figure below Figure 8 129 Removing 2 RUBBER CAPS on SPEAKER COVER LEFT RIGHT Figure 8 130 Unscrewing 2 screws ...

Страница 436: ...CE MANUAL 8 88 Section 8 25 Replacement of OPIO 3 Remove the 5 probe holders Refer to the figure below 4 Unscrew 9 screws located in the bottom of the OPIO Refer to the figure below Figure 8 131 Removing the 5 probe holders Figure 8 132 Unscrewing 9 screws located on the bottom of the OPIO ...

Страница 437: ...5 Lift up the OPIO and disconnect Power OPIO 2 USB cables from the OPIO HUB PWA Refer to the figure below 6 Cut the CABLE TIE by nipper Figure 8 133 Disconnecting Power OPIO 2 USB cables Figure 8 134 Cutting the CABLE TIE by nipper Insert power cable KEYBOARD ASSY Host interface 10 pin connector OPIO cable USB connector 1 USB connector 2 ...

Страница 438: ... 8 90 Section 8 25 Replacement of OPIO 7 Unscrew 4 screws to remove upper keyboard assembly 8 25 5 Installation Procedure 1 Install the new parts in the reverse order of removal 2 Perform FRU8 10 Replacement of OPIO Functional Tests Figure 8 135 Unscrewing 4 screws to remove upper keyboard assembly ...

Страница 439: ...4 3 2 on page 4 4 8 26 4 Removal Procedure 1 Remove the MAIN OPIO and SUB OPIO Refer to the Section 8 25 on page 8 87 OPIO 2 Disconnect the 2 SPEAKER CONNECTORS Refer to the figure below 3 Unscrew each 3 screws of SPEAKER ASSY L R and remove Refer to the figure below 8 26 5 Installation Procedure 1 Install the new parts in the reverse order of removal 2 Perform FRU8 11 Replacement of the SPEAKER F...

Страница 440: ...Remove the PEDESTAL BOTTOM COVER BT16 8 27 1 Manpower One person 2 minutes 8 27 2 Tools N A 8 27 3 Preparations N A 8 27 4 Removal Procedure 1 Remove PEDESTAL BOTTOM COVER BT16 Refer to the figure below 8 27 5 Installation Procedure 1 Install the new parts in the reverse order of removal Figure 8 138 Removing the PEDESTAL BOTTOM COVER BT16 PEDESTAL BOTTOM COVER BT16 ...

Страница 441: ... of the NEST 8 28 1 Manpower One person 10 minutes 8 28 2 Tools Philips screwdriver Allen Wrench 8 28 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 28 4 Removal Procedure 1 Remove the FRONT CABINET BEZEL ASSY BT16 Refer to the figure below Figure 8 139 Removing the FRONT CABINET BEZEL ASSY BT16 FRONT CABINET BEZEL ASSY BT16 ...

Страница 442: ...on 8 28 Replacement of the NEST 2 Remove the REAR CABINET BEZEL ASSY BT16 Refer to the figure below 3 Unscrew 7 screws and remove the REAR EMI COVER BRKT Refer to the figure below Figure 8 140 Removing the Rear Cabinet Bezel Assy BT16 Figure 8 141 Removing the REAR EMI COVER BRKT REAR CABINET BEZEL ASSY BT16 ...

Страница 443: ...GE DRAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 95 4 Disconnect all connectors Figure 8 142 Disconnection all cables ...

Страница 444: ...AL 8 96 Section 8 28 Replacement of the NEST 5 Set the 3 RS KNOBS to unlock position to separate from the SIDE COVER R and unscrew one screw from the each 3 RS KNOB Refer to the figure below Figure 8 143 Removing the 3 RS KNOBS Removing first RS Knob Removing second RS Knob Removing third RS Knob ...

Страница 445: ...er 8 Replacement Procedures 8 97 6 Unscrew 4 screws and remove the Nest box Refer to the figure below 8 28 5 Installation Procedure 1 Install the new parts in the reverse order of removal 2 Perform FRU8 13 Replacement of the NEST Functional Tests Figure 8 144 Unscrewing 4 screws and removing the NEST BOX ...

Страница 446: ...LTER 8 29 1 Manpower One person 1 minutes 8 29 2 Tools None 8 29 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 29 4 Removal Procedure 1 Pull out the FRONT CABINET BEZEL ASSY BT16 Refer to the figure below 2 Pull out the AIR FILTER Refer to the figure below Figure 8 145 Removing the FRONT CABINET BEZEL ASSY BT16 Figure 8 146 Removing the AIR FILTER ...

Страница 447: ...OLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 99 8 29 5 Installation Procedure 1 Install the new parts in the reverse order of removal ...

Страница 448: ...of the Probe Holder KIT Section 8 30 Replacement of the Probe Holder KIT 8 30 1 Manpower One person 1 minutes 8 30 2 Tools None 8 30 3 Removal Procedure 1 Remove 5 probe holders Refer to the figure below 8 30 4 Installation Procedure 1 Install the new parts in the reverse order of removal Figure 8 147 Removing the 5 PROBE HOLDERs ...

Страница 449: ... 31 2 Tools Philips Screwdriver 8 31 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 31 4 Removal Procedure 1 Unscrew 2 screws and remove the Tray Side Cabinet BT16 Refer to the figure below 2 Remove the FRONT PRINTER BEZEL ASSY BT16 Refer to the figure below Figure 8 148 Removing the Tray Side Cabinet BT16 Figure 8 149 Removing the FRONT PRINTER BEZEL ASSY BT16 ...

Страница 450: ...COVER BT16 Refer to the figure below 4 Unscrew 2 screws and remove the ASSY LEFT SIDE PLASTIC COVER BT16 Refer to the figure below 8 31 5 Installation Procedure 1 Install the new parts in the reverse order of removal Figure 8 150 Removing the TOP CABINET COVER BT16 Figure 8 151 Removing the ASSY LEFT SIDE PLASTIC COVER BT16 TOP CABINET COVER BT16 ASSY LEFT SIDE PLASTIC COVER BT16 ...

Страница 451: ... 8 32 2 Tools Philips Screwdriver 8 32 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 32 4 Removal Procedure 1 Unscrew 2 screws and remove the Tray Side Cabinet Refer to the figure below 2 Remove the FRONT PRINTER BEZEL ASSY BT16 Refer to the figure below Figure 8 152 Removing the Tray Side Cabinet Figure 8 153 Removing the FRONT PRINTER BEZEL ASSY BT16 FRONT PR...

Страница 452: ...VER R ASSY 3 Remove the COVER TOP CABINET PLASTIC BT16 Refer to the figure below 4 Unscrew 3 screws and remove the RS Knobs Refer to the figure below Figure 8 154 Removing the COVER TOP CABINET PLASTIC BT16 Figure 8 155 Removing the RS Knob COVER TOP CABINET PLASTIC BT16 Removing first RS knob Removing second RS Knob Removing third KR Knob ...

Страница 453: ... 8 Replacement Procedures 8 105 5 Unscrew 2 screws and remove the ASSY RIGHT SIDE PLASTIC COVER BT16 Refer to the figure below 8 32 5 Installation Procedure 1 Install the new parts in the reverse order of removal Figure 8 156 Removing the ASSY RIGHT SIDE PLASTIC COVER BT16 ASSY RIGHT SIDE PLASTIC COVER BT16 ...

Страница 454: ...8 106 Section 8 33 Replacement of the Trackball Section 8 33 Replacement of the Trackball 8 33 1 Manpower One person 20 minutes 8 33 2 Tools Philips Screwdriver 8 33 3 Removal procedure 1 Unscrew 9 screws at the bottom of the keyboard body Refer to the figures below Figure 8 157 Unscrewing 9 screws ...

Страница 455: ...RAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 107 2 Remove 5 probe holders Refer to the figure below 3 Lift up main keyboard assy Refer to the figure below Figure 8 158 Removing 5 probe holders Figure 8 159 Lifting up main keyboard assy ...

Страница 456: ...N P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 8 108 Section 8 33 Replacement of the Trackball 4 Disconnect 4 cables Refer to the figure below Figure 8 160 Disconnecting 4 cables ...

Страница 457: ...w 2 screws on trackball and disconnect the cable side of the trackball Refer to the figures below 8 33 4 Installation Procedure 1 Put the trackball guide ring Refer to the figure below If the trackball guide ring is installed already please skip step 1 Figure 8 161 Unscrewing 2 screws on trackball Figure 8 162 Installing the trackball guide ring ...

Страница 458: ...main PWA and screw the trackball with 2 screws Refer to the figure below NOTE Please check the position of the label as the below figure to connect in the right position 3 Install the OPIO ASSY like as reverse order of removal procedure from step 4 of 8 33 3 Removal procedure Figure 8 163 Connecting the cable to the connector and screwing the trackball CP Main PWA side Ball Trackball side ...

Страница 459: ...8 34 2 Tools Philips Screwdriver 8 34 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 34 4 Removal Procedure 1 Follow the removal procedure of Section 8 25 Replacement of OPIO 2 Remove 8 TGC control key caps refer to Section 8 19 Replacement of OPIO Plastic lower cover assy on page 8 56 3 Disconnect TGC BOARD CABLE Refer to the figure below Figure 8 164 Disconnec...

Страница 460: ...4 Replacement of the TGC BOARD 4 Unscrew 4 screws then pull out the TGC BOARD Refer to the figure below 8 34 5 Installation Procedure 1 Install the new parts in the reverse order of removal 2 Perform FRU8 19 Replacement of the TGC BOARD Functional Tests Figure 8 165 Unscrewing 4 screws and pulling out the TGC BOARD ...

Страница 461: ...er 8 35 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 35 4 Removal Procedure 1 Follow the removal procedure of Section 8 25 Replacement of OPIO 2 Disconnect the HUB TO MAIN OPIO CABLE Refer to the figure below 3 Unscrew 10 screws and pull out the AN KEYBOARD and OPIO HUB PWA ASSY Refer to the figure below Figure 8 166 Disconnecting the HUB TO MAIN OPIO CABLE Fi...

Страница 462: ...VISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 8 114 Section 8 35 Replacement of AN KEYBOARD ASSY 8 35 5 Installation Procedure 1 Install the new parts in the reverse order of removal 2 Perform FRU8 20 Replacement of AN KEYBOARD ASSY Functional Tests ...

Страница 463: ...Caster Cap BT16 8 36 1 Manpower One person 2 minutes 8 36 2 Tools None 8 36 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 NOTE Lock Caster before removing it 8 36 4 Removal Procedure 1 Remove the FRONT CABINET BEZEL ASSY BT16 Refer to the figure below Figure 8 168 Removing the FRONT CABINET BEZEL ASSY BT16 FRONT CABINET BEZEL ASSY BT16 ...

Страница 464: ...placement of the Caster Cap BT16 2 Remove the FRONT PRINTER BEZEL ASSY BT16 Refer to the figure below 3 Remove the Top Cabinet Cover BT16 Refer to the figure below Figure 8 169 Removing the FRONT PRINTER BEZEL ASSY BT16 Figure 8 170 Removing the Top Cabinet Cover BT16 FRONT PRINTER BEZEL ASSY BT16 TOP CABINET COVER BT16 ...

Страница 465: ...243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 117 4 Remove the Rear Cabinet Bezel Assy BT16 Refer to the figure below v Figure 8 171 Removing the Rear Cabinet Bezel Assy BT16 Rear Cabinet Bezel Assy BT16 ...

Страница 466: ...Caster Cap BT16 5 Unscrew 3 screws and remove the RS Knobs Refer to the figure below 6 Unscrew 4 screws and remove the SIDE CABINET TRAY BT16 Refer to the figure below Figure 8 172 Removing the RS Knob Figure 8 173 Removing the SIDE CABINET TRAY BT16 Remove first RS Knob Remove second RS Knob Remove third RS Knob SIDE CABINET TRAY BT16 ...

Страница 467: ...Procedures 8 119 7 Unscrew 2 screws and remove the SIDE COVER L ASSY BT16 Refer to the figure below 8 Unscrew 2 screws and remove the SIDE COVER R ASSYBT16 Refer to the figure below Figure 8 174 Removing the SIDE COVER L ASSYBT16 Figure 8 175 Removing the SIDE COVER R ASSYBT16 SIDE COVER L ASSY BT16 SIDE COVER R ASSY BT16 ...

Страница 468: ...UAL 8 120 Section 8 36 Replacement of the Caster Cap BT16 9 Remove the COVER CASTOR NARROW L R BT16 Refer to the figure below 8 36 5 Installation Procedure 1 Install the new parts in the reverse order of removal Figure 8 176 Removing the COVER CASTOR NARROW L R BT16 COVER CASTOR NARROW L R BT16 ...

Страница 469: ...ower One person 10 minutes 8 37 2 Tools Monkey Wrench 8 37 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 NOTE Lock casters before removing them 8 37 4 Removal Procedure 1 Follow the removal procedure of Section 8 24 Replacement of ASSY Bumper Base Rubber 2 Unscrew the NUT And pull out the CASTER Refer to the figure below Figure 8 177 Unscrewing the NUT ...

Страница 470: ... DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 8 122 Section 8 37 Replacement of the CASTER Figure 8 178 Unscrewing the nut 8 37 5 Installation Procedure 1 Install the new parts in the reverse order of removal ...

Страница 471: ... Philips screwdriver 8 38 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 38 4 Removal Procedure 1 Unscrew 2 screws and remove the SIDE CABINET TRAY BT16 Refer to the figure below 8 38 5 Installation Procedure 1 Install the new parts in the reverse order of removal 2 Perform FRU8 25 Replacement of the Side Cabinet Tray BT16 Functional Tests Figure 8 179 Unscrewin...

Страница 472: ...EAR HANDLE 8 39 1 Manpower One person 5 minutes 8 39 2 Tools Philips screwdriver Allen Wrench 8 39 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 39 4 Removal Procedure 1 Remove the ASSY BEZEL FRONT PRINTER PLASTIC BT16 Refer to the figure below Figure 8 180 Removing the ASSY BEZEL FRONT PRINTER PLASTIC BT16 ASSY BEZEL FRONT PRINTER PLASTIC BT16 ...

Страница 473: ...igure below 3 Remove BOTTOM PEDESTAL COVER BT16 Refer to the figure below 4 Unscrew 2 screws and remove the FRONT PEDESTAL COVER BT16 Refer to the figure below Figure 8 181 Removing the TOP CABINET COVER BT16 Figure 8 182 Removing the BOTTOM PEDESTAL COVER BT16 Figure 8 183 Removing the FRONT PEDESTAL COVER BT16 TOP CABINET COVER BT16 BOTTOM PEDESTAL COVER BT16 FRONT PEDESTAL COVER BT16 ...

Страница 474: ... 2018 BASIC SERVICE MANUAL 8 126 Section 8 39 Replacement of the REAR HANDLE 5 Unscrew 2 screws and remove the REAR HANDLE Refer to the figure below 8 39 5 Installation Procedure 1 Install the new parts in the reverse order of removal Figure 8 184 Removing the REAR HANDLE ...

Страница 475: ...rations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 40 4 Removal Procedure 1 Remove 2 RUBBER CAPS on SPEAKER COVER LEFT RIGHT Refer to the figure below 2 Unscrew 2 screws then remove the SPEAKER COVER LEFT RIGHT Refer to the figure below 8 40 5 Installation Procedure 1 Install the new parts in the reverse order of removal Figure 8 185 Removing 2 RUBBER CAPS on SPEAKER COVER...

Страница 476: ...S DPS with Battery 8 41 1 Manpower One person 30 minutes 8 41 2 Tools Philips screwdriver Allen Wrench 8 41 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 2 Remove the REAR CABINET BEZEL ASSY BT16s and disconnect cables as described in 8 28 4 Removal Procedure 2 3 4 and 7 on page 8 93 3 Remove the NEST EMI BRKT BT16 as described in 8 56 4 Removal Procedure ...

Страница 477: ...er 8 Replacement Procedures 8 129 8 41 4 Removal Procedure 1 Draw the DPS out of the Nest Frame 8 41 5 Installation Procedure 1 Install the new parts in the reverse order of removal 2 Perform FRU8 28 Replacement of the DPS DPS with Battery Functional Tests Figure 8 187 DPS Figure 8 188 Drawing the DPS DPS ...

Страница 478: ...FM 8 42 1 Manpower One person 30 minutes 8 42 2 Tools Philips screwdriver Allen Wrench 8 42 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 2 Remove the REAR CABINET BEZEL ASSY BT16s and disconnect cables as described in 8 28 4 Removal Procedure 2 3 4 and 7 on page 8 93 3 Remove the NEST EMI BRKT BT16 as described in 8 56 4 Removal Procedure ...

Страница 479: ...N P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 131 8 42 4 Removal Procedure 1 Draw the DRFM out of the Nest Frame Figure 8 189 DRFM Figure 8 190 Drawing the DRFM DRFM ...

Страница 480: ...ber Reprogamming Procedure to reprogram the serial number 3 Perform 8 42 7 Installation of software patch CD NOTE Check S1 if GPU is mounted PIN number 1 2 3 of the switch should be located in left If GPU is not mounted PIN number 1 2 3 of the swtich should be located in right If UVC option is included PIN number 4 of the switch should be located in left If UVC option is not included PIN number 4 ...

Страница 481: ...D number See SYSTEM INFO page or label of DRFM board 1 Gather System Serial number and Hardware ID Press the F2 key on upper OPIO select SYSTEM SETUP select ADMINISTRATION click on SYSTEM INFO and write down the Hardware ID number NOTICE NOTICE To reprogramming the System Serial Number contact the OLC On Line Center to get the necessary Serial Key and Permanent Key calculated in OKOS Figure 8 193 ...

Страница 482: ...FM 2 Call the OLC to get the Serial Key and the Permanent Key 3 Program the System Serial number and re enter the Permanent Key To program the System SNO click the OPTIONS tab Figure 8 194 Hardware ID on label DRFM Figure 8 195 System Setup Administration Options page Hardware ID Field to enter the Serial Key Field to enter Permanent key ...

Страница 483: ...e box BT key state has been changed with YES The system reboots 4 Check System Serial number System Type Hardware ID and Software options a Press the F2 key on upper OPIO select SYSTEM SETUP select ADMINISTRATION and then select the OPTION tab System Type must be correct depending on Voluson P8 Voluson P6 Software options and serial number must be correct 5 Perform FRU8 29 Replacement of the DRFM ...

Страница 484: ...led on the system with DRFM 2 SW patch needs to be installed again for Ext 4 and previous versions NOTE Software patch does not need to be installed from software version Ext 5 or later NOTE Software patch can also be downloaded at http ge box com s yspo13ahrk3uq79gmif10api72p73f1 1 Insert Software patch CD on the system 2 Click the ASI button for patch Util System setup Administration Service tab...

Страница 485: ...EVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 137 3 Choose media to DVD CD and click Load 4 Click OK 5 Insert y or yes to install patch on command window Figure 8 197 Choosing media to DVD CD Figure 8 198 Click OK ...

Страница 486: ...TE After installing the software please keep the software patch CD with the system for later usage For S10 S6 S8 please store the CD inside the rear cover For P6 P8 please store the CD below the top cabinet cover NOTE In case of an inoperative system please replace the DRFM 2 first before installing the software patch CD After installing the software patch CD please re start the system Figure 8 19...

Страница 487: ...ower Off Shutdown the system as described in 4 3 2 on page 4 4 2 Remove the REAR CABINET BEZEL ASSY BT16 and disconnect cables as described in 8 28 4 Removal Procedure 2 3 4 and 7 on page 8 93 3 Remove the NEST EMI BRKT BT16 as described in 8 56 4 Removal Procedure 4 Remove the DRFM as described in 8 42 3 on page 8 130 8 43 4 Removal Procedure 1 Unscrew 5 screws and pull the SOM board out of the D...

Страница 488: ...rts in the reverse order of removal 2 Perform FRU8 30 Replacement of the SOM in DRFM Functional Tests CAUTION CAUTION When replacing with new SOM Push the SOM in both side of DRFM ASSEMBLY without putting weight on the board NOTICE As a new specification the system can be turned off up to 3 times due to recognizing GPU initializing Intel device and Bios In this case please turn on the system again...

Страница 489: ...INET BEZEL ASSY BT16 and disconnect cables as described in 8 28 4 Removal Procedure 2 3 4 and 7 on page 8 93 3 Remove the NEST EMI BRKT BT16 as described in 8 56 4 Removal Procedure 4 Remove the DRFM as described in 8 42 3 on page 8 130 8 44 4 Removal Procedure 1 Unscrew 4 screws and pull the HDD out of the DRFM 8 44 5 Installation Procedure 1 Install the new parts in the reverse order of removal ...

Страница 490: ... One person 30 minutes 8 45 2 Tools Philips screwdriver 8 45 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 2 Remove the REAR CABINET BEZEL ASSY BT16 and disconnect cables as described in 8 28 4 Removal Procedure 2 3 4 and 7 on page 8 93 8 45 4 Removal Procedure 1 Pull the DBM DPI out of the Nest Frame Figure 8 202 DBM DPI Figure 8 203 Drawing the DBM DPI DBM64 ...

Страница 491: ...t Procedures 8 143 2 Unscrew 6 screws 3 Remove DBM64 from the EMI SHIELD BRACKET 8 45 5 Installation Procedure 1 Install the new parts in the reverse order of removal 2 Perform FRU8 36 Replacement of the DBM64 Functional Tests Figure 8 204 Unscrewing 6 screws and removing the EMI SHIELD BRACKET 1 2 EMI SHIELD BRACKET 3 ...

Страница 492: ...er One person 30 minutes 8 46 2 Tools Philips screwdriver Allen Wrench Hirose ejector 8 46 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 2 Remove the REAR CABINET BEZEL ASSY BT16 and disconnect cables as described in 8 28 4 Removal Procedure 2 3 4 and 7 on page 8 93 3 Remove the NEST EMI BRKT BT16 as described in 8 56 4 Removal Procedure 4 Remove 3 RS Knobs ...

Страница 493: ...23243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 145 8 46 4 Removal Procedure 1 Draw the DBM T DPI out of the Nest Frame 2 Unscrew 6 screws Figure 8 205 DBM T DPI Figure 8 206 Drawing the DBM T DPI T DPI ...

Страница 494: ...46 Section 8 46 Replacement of the T DPI 3 Remove T DPI from the EMI SHIELD BRACKET 8 46 5 Installation Procedure 1 Install the new parts in the reverse order of removal 2 Perform FRU8 37 Replacement of the T DPI Functional Tests Figure 8 207 Removing the EMI SHIELD BRACKET Figure 8 208 Bottom of T DPI ...

Страница 495: ...the CSI 8 47 1 Manpower One person 30 minutes 8 47 2 Tools Philips screwdriver 8 47 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 2 Pull out the NEST as described in Section 8 28 Replacement of the NEST on page 8 93 8 47 4 Removal Procedure 1 Unscrew 4 screws to remove the LED Indicator bracket Figure 8 209 Opening the rear bracket ...

Страница 496: ...of the CSI 2 Unscrew 4 screws to remove CSI board and Separate the LED connector 8 47 5 Installation Procedure 1 Install the new parts in the reverse order of removal 2 Perform FRU8 38 Replacement of the CSI Functional Tests Figure 8 210 Seperating LED connector FAN connector HDD Status Network Status Network Speed Orange Megabit ...

Страница 497: ... 2 Tools Philips screwdriver Allen Wrench 8 48 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 2 Pull out the NEST as described in Section 8 28 Replacement of the NEST on page 8 93 8 48 4 Removal Procedure 1 Remove the CSI Refer to Section 8 47 Replacement of the CSI on page 8 147 2 Remove fan connector and unscrew 8 screws Figure 8 211 Removing fan connector and u...

Страница 498: ...Y 23 2018 BASIC SERVICE MANUAL 8 150 Section 8 48 Replacement of the CBP 3 Unscrew 6 screws 8 48 5 Installation Procedure 1 Install the new parts in the reverse order of removal 2 Perform FRU8 39 Replacement of the CBP Functional Tests Figure 8 212 Unscrewing 6 screws ...

Страница 499: ... DVD 8 49 1 Manpower One person 10 minutes 8 49 2 Tools Philips screwdriver 8 49 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 49 4 Removal Procedure 1 Remove FRONT PRINTER BEZEL ASSY BT16 Refer to the below picture 2 Unplug two cables from the DVD Figure 8 213 Removing FRONT PRINTER BEZEL ASSY BT16 Figure 8 214 Unpluging two cables ...

Страница 500: ...placement of the DVD 3 Unscrewing 2 screws to remove the DVD 4 Unscrewing 4 screws to remove the SIDE STRIPS 8 49 5 Installation Procedure 1 Install the new parts in the reverse order of removal 2 Perform FRU8 40 Replacement of the DVD Functional Tests Figure 8 215 Removing the DVD Figure 8 216 Removing the SIDE STRIPS ...

Страница 501: ...ub PWA 8 50 1 Manpower One person 30 minutes 8 50 2 Tools Philips screwdriver 8 50 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 2 Remove the OPIO covers and cables as described in 8 25 4 Removal Procedure on page 8 87 8 50 4 Removal Procedure 1 Pull 2 cables from Main Keyboard ASSY and prepare the Sub Keyboard ASSY Figure 8 217 Pulling 2 cables ...

Страница 502: ...nt of the OPIO USB Sub PWA 2 Unscrew 5 screws and disconnect 2 USB cables Then Remove 2 Brackets and OPIO USB Sub PWA 8 50 5 Installation Procedure 1 Install the new parts in the reverse order of removal 2 Perform FRU8 42 Replacement of the OPIO USB Sub PWA Functional Tests Figure 8 218 Removing 2 Brackets and OPIO USB Sub PWA ...

Страница 503: ... Flat head screw driver or Flat tool 8 51 3 Preparations 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 51 4 Key Caps Removal Procedure 1 Remove the Key should be changed with Flat tool Refer to the figure below 2 Assemble the Key Refer to the figure below 3 Perform FRU8 43 Replacement of AN KEY TOP Functional Tests Figure 8 219 Removing the Key should be changed Figure 8 220 ...

Страница 504: ...uson P8 Voluson P6 system NOTICE NOTICE READ and UNDERSTAND these instructions thoroughly before proceeding with the installation Perform each step in sequence and check it off when completed If a problem occurs after completion go back through the procedure and check for implementation errors before contacting your Ultrasound Region CAUTION CAUTION Possible Operational damage Failure to strictly ...

Страница 505: ...Assemble 2 EMI gaskets 5448934 EMI gasket side b w printer bkt 3 on printer base plate 2 Assemble 2 EMI gaskets 5447303 EMI gasket side b w printer bkt 2 on printer base plate 3 Assemble 2 EMI gaskets 5457475 EMI gasket side b w printer bkt 1 on printer base plate 4 Assemble printer base pate with B W printer using 8 screws 2306562 screw FH M3 x 6 Figure 8 221 B W Printer assembly EMI Gaskets 5457...

Страница 506: ...tional Peripherals 5 Disassemble FRONT PRINTER BEZEL ASSY BT16 6 Remove PRINTER DUMMY COVER PLATE BT16 Figure 8 222 Disassembling PRINTER DUMMY COVER PLATE BT16 7 Unscrew 2 screws to disassemble power cord fix bracket 8 Remove the power cord and REAR CABINET BEZEL ASSY BT16 Figure 8 223 Disassembling REAR CABINET BEZEL ASSY BT16 ...

Страница 507: ...3 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 159 9 Unscrew 7 screws and disassemble REAR EMI COVER 10 Unscrew 2 screws and disassemble printer EMI cover 1 and 2 Figure 8 224 Disassembling printer EMI covers ...

Страница 508: ...with sheet metal chassis Figure 8 225 Assembling printer 13 Carefully insert the printer base plate in to the chassis slot 14 Assemble the printer base plate with chassis using 2 screws 2159634 screw BH M4X10 15 Assemble FRONT CABINET BEZEL ASSY BT16 Figure 8 226 Assembling printer base plate Printer Assembly Push the printer assembly Printer base plate Assemble printer base plate with two screws ...

Страница 509: ...he nest box Figure 8 227 Inserting Printer USB and Power cable 19 Check the printer USB cable and power cable to passing through in EMI cover slot Then screw 2 screws and assemble printer EMI cover 1 and 2 20 Route the B W printer USB and power cable through the left side cable clamp Figure 8 228 Routing the printer cables Printer Power Cable Printer USB Cable Inserting Printer power cable into th...

Страница 510: ...al Peripherals 21 Screw the 7 screws and fix the REAR EMI COVER 22 Assemble the REAR CABINET BEZEL ASSY BT16 power cord and POWER CORD FIX BRACKET 23 Assemble the FRONT PRINTER BEZEL ASSY BT16 Figure 8 229 Assembling rear side parts Screw the 7 screws Assemble power cord and FRONT PRINTER BEZEL FRONT PRINTER BEZEL ASSY BT16 ASSY BT16 ...

Страница 511: ...asket side b w printer bkt 3 on printer base plate 2 Assemble 2 EMI gaskets 5448934 EMI gasket side b w printer bkt 3 on printer base plate 3 Assemble 2 EMI gaskets 5457475 EMI gasket side b w printer bkt 1 on printer base plate 4 Assemble printer base pate with B W printer using 8 screws 2306562 screw FH M3 x 6 Figure 8 230 Color Printer assembly EMI Gaskets 5448934 EMI Gaskets 5447303 EMI Gasket...

Страница 512: ... Section 8 52 Replacing optional Peripherals 5 Remove the FRONT PRINTER BEZEL ASSY BT16 Figure 8 231 FRONT PRINTER BEZEL ASSY BT16 6 Disassemble FRONT PRINTER BEZEL ASSY BT16 Figure 8 232 Removing POWER CORD FIX BRACKET and REAR CABINET BEZEL ASSY BT16 POWER CORD FIX BRACKET REAR CABINET BEZEL ASSY BT16 ...

Страница 513: ...33 Disassembling printer EMI covers 9 Disassemble FRONT CABINET BEZEL ASSY BT16 10 Remove left side printer bracket 11 Remove left side printer bracket 12 Remove right side printer bracket Figure 8 234 Removing FRONT CABINET BEZEL ASSY BT16 and printer brackets Unscrew 7 screws 2 screwsscrews and printer EMI covers 1 and 2 Rear side view after removing the EMI covers 1 and 2 FRONT CABINET BEZEL AS...

Страница 514: ...ssembly 14 Assemble the printer base plate with chassis using 2 screws 2159634 screw BH M4X10 Figure 8 235 Assembling printer base plate 15 Insert the printer USB cable and printer power cable 16 Insert the printer power cable in to the nest box Figure 8 236 Printer cables 2 screws 2159634 screw BH M4X10 Printer USB cable Printer Power cable Printer Power cable in Nest box ...

Страница 515: ...nter USB and power cable through the left side cable clamp Figure 8 237 Routing the cables 19 Screw the 7 screws and fix the REAR EMI COVER 20 Assemble the REAR CABINET BEZEL ASSY BT16 power cord and POWER CORD FIX BRACKET Figure 8 238 Assembling the rear side parts Printer USB cable in Nest box Left side cable clamp Screw the 7 screws Assemble REAR CABINET BEZEL ASSY BT16 power cord and POWER COR...

Страница 516: ...96 screw washer head M3x6 22 Assemble 4 numbers of Holding Catch female with DVD front cover using 8 screws 2306562 screw FH M3x6 Figure 8 239 Assembling DVD front cover 23 Assemble 2 numbers of holding catch male with sheet metal chassis using 4 screws 2306562 screw FH M3x6 24 Assemble ASSY BEZEL FRONT DVD PLASTIC BT16 with sheet metal chassis Figure 8 240 Assembling Holding Catch male and ASSY B...

Страница 517: ...VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 169 25 Assemble the FRONT CABINET BEZEL ASSY BT16 Figure 8 241 Assembling the FRONT CABINET BEZEL ASSY BT16 ...

Страница 518: ...placing optional Peripherals 8 52 4 HORIZONTAL TV PROBE HOLDER Installation 1 Install HORIZONTAL TV PROBE HOLDER on the left side bottom of OPIO with 3 screws Figure 8 242 Assembling HORIZONTAL TV PROBE HOLDER with 3 screws Figure 8 243 HORIZONTAL TV PROBE HOLDER is installed on Left Side only HORIZONTAL TV PROBE HOLDER ...

Страница 519: ...PROBE HOLDER Installation 1 Install VERTICAL TV PROBE HOLDER BASE on the right side bottom of OPIO with 2 screws 2 Place the TV PROBE HOLDER RUBBER on the PROBE HOLDER BASE Figure 8 244 Assembling TV PROBE HOLDER BASE with 2 screws 2159634 BH M4x10 Figure 8 245 Inserting TV PROBE HOLDER RUBBER to PROBE HOLDER BASE TV PROBE TV PROBE HOLDER BASE HOLDER RUBBER ...

Страница 520: ...ons 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 2 Pull out the NEST as described in Section 8 28 Replacement of the NEST on page 8 93 8 53 4 Removal Procedure 1 Unscrew 4 screws and remove fan connector 8 53 5 Installation Procedure 1 Install the new parts in the reverse order of removal NOTE In case of installing FAN with Cable ASSY GPN 5669586 replace screws to SCREW PH M4X...

Страница 521: ...spatch tool Use this script in the Service Comments when debriefing a Service Dispatch For a replaced FRU use the following debrief script Proprietary Service Manual Voluson P8 Voluson P6 Revision Voluson P8 Voluson P6 Section fill in appropriate section Equipment passed all required tests and is ready for use And if testing leaking current use the following text Proprietary Service Manual Voluson...

Страница 522: ...CTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 8 174 Section 8 55 Replacement of UVC Section 8 55 Replacement of UVC Replace FRU part as described in system s UVC Installation Kit Installation Instruction 5758249 in CDL ...

Страница 523: ... 56 Replacement of CWD piggy 8 56 1 Manpower One person 30 minutes 8 56 2 Tools Philips screwdriver Hirose ejector 8 56 3 Preparations Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 56 4 Removal Procedure 1 Unscrew 2 screws and remove POWER CORD FIX BRKT and power cord Figure 8 247 Power Bracket POWER CORD FIX BRACKET power cord ...

Страница 524: ...018 BASIC SERVICE MANUAL 8 176 Section 8 56 Replacement of CWD piggy 2 Remove REAR CABINET BEZEL ASSY BT16 3 Unscrew 7 screws and disassemble the REAR EMI COVER Figure 8 248 REAR CABINET BEZEL ASSY BT16 Figure 8 249 Unscrewing 7 fastners and disassembling the REAR EMI COVER ...

Страница 525: ...AY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 177 4 Unplug all cables and Unscrew 7 screws and remove the NEST EMI BRKT BT16 5 Remove 3 RS KNOBS Figure 8 250 Unpluging all cables and removing the NEST EMI BRKT BT16 Figure 8 251 Removing 3 RS Knobs ...

Страница 526: ... VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 8 178 Section 8 56 Replacement of CWD piggy 6 Pull out DBM64 T DPI Boards Figure 8 252 Pulling out DBM64 T DPI Boards ...

Страница 527: ...ent Procedures 8 179 7 Unscrew 1 screw and replace the CWD Piggy 8 56 5 Installation Procedure 1 Install the new board assy in the reverse order of removal procedure 2 Perform FRU8 48 Functional Checks after FRU replacement Debrief Guidelines Functional Tests Figure 8 253 Unscrewing 1 screw and replacing the CWD piggy ...

Страница 528: ...ement of ECG Kit assy 8 57 1 Manpower One person 10 minutes 8 57 2 Tools Philips screwdriver 8 57 3 Preparations Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 57 4 Removal Procedure 1 Remove 5 probe holders 2 Unscrew 9 numbers of screws below the keyboard base Figure 8 254 Removing 5 Probe holders and unscrewing 9 screws below the keyboard base ...

Страница 529: ... Procedures 8 181 3 Open keyboard assembly 4 Remove ECG DUMMY BRKT Figure 8 255 ECG DUMMY BRKT 8 57 5 Installation Procedure 1 Assemble ECG kit on keyboard base 2 Screw ECG kit with Keyboard base using 2 screws 2159625 screw w SP PH m4x8 and fix ECG USB cable by using the hook of bracket Figure 8 256 ECG kit Installation ...

Страница 530: ...IC SERVICE MANUAL 8 182 Section 8 57 Replacement of ECG Kit assy 3 Connect the USB cable with Keyboard assembly as shown 4 Assemble Keyboard assembly Figure 8 257 USB cable 5 Screw 9 numbers of screws below the 6 Assemble 5 probe holders Figure 8 258 KEYBOARD BASE and probe holders ...

Страница 531: ...ON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures 8 183 7 Attach ECG label Figure 8 259 ECG label 8 Perform FRU8 49 Replacement of ECG Kit assy Functional Tests ECG label ...

Страница 532: ...Pack Assy 8 58 1 Manpower One person 20 minutes 8 58 2 Tools Philips screwdriver 8 58 3 Preparation Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 58 4 Removal procedure 1 Follow the removal procedure of Section 8 41 Replacement of the DPS DPS with Battery on page 8 128 2 Unscrew 14 screws and remove DPS with Battery cover Figure 8 260 Removing DPS with Battery cover ...

Страница 533: ...se order of removal 2 Perform FRU8 50 Replacement of Battery Pack Assy Functional Tests NOTE When the AC power is plugged in the system fan will be operated automatically until the battery is fully chaged Even though the AC power is plugged in with the full battery the system fan will also be operated automatically for less than one minute Figure 8 261 Disassembling Battery power and signal cables...

Страница 534: ...ndard Field Service Tool Kit 8 59 3 Preparation 1 Power Off Shutdown the system as described in 4 3 2 on page 4 4 8 59 4 Removal Procedure 1 Unscrew 2 screws to remove POWER CORD BRACKET 2 Remove REAR COVER and unscrew 7 screws to remove REAR EMI COVER Figure 8 263 Unscrewing 2 screws to remove POWER CORD BRACKET Figure 8 264 Removing REAR COVER and unscrewing 7 screws to remove REAR EMI COVER POW...

Страница 535: ... Procedures 8 187 3 Diconnect the WERELESS LAN cable from NEST BOX 4 Untie the cable clamp 5 Pull out WIRELESS LAN dongle from loop sheet Figure 8 265 Disconnecting WIRELESS LAN cable from NEST BOX Figure 8 266 Untying cable clamp to unroute WIRELESS LAN Cable Figure 8 267 Pulling out WIRELESS LAN dongle from loop sheet ...

Страница 536: ...8 59 Replacement of WIRELESS LAN 6 Unfold WIRELESS LAN and Check the dongle label before installation 7 Attach loop sheet in WIRELESS LAN dongle Refer to below figure 5499363 self adhesive hook and loop W30XL75 Figure 8 268 Checking the dongle label Figure 8 269 Attaching loop sheet WIRELESS LAN dongle AUGN EDGE ...

Страница 537: ...System setup Connectivity Device Setup WLAN Configuration The Wireless Network Configuration tool appears 2 Go to Wireless Network tab and check that available Wireless Networks connection list 3 Double click the available wireless network resource and edit the network key to connect 8 59 7 Functional Test 1 Select Utility System setup Connectivity Device Setup WLAN Configuration 2 Move to Propert...

Страница 538: ...3 1 Power On Boot Up Proprietary Service Manual Voluson P8 Voluson P6 Revision Voluson P8 Voluson P6 Section 8 2 Equipment passed all required tests and is ready for use 8 3 Software and Functional Checks after Installation Upgrade Procedure Table 8 2 System Function 4 3 2 Power Off Shutdown See Section Functional and or Leakage Current Test Debrief Script 4 3 1 Power On Boot Up Proprietary Servic...

Страница 539: ...Functional and or Leakage Current Test Debrief Script Table 8 2 System Function Proprietary Service Manual Voluson P8 Voluson P6 Revision 6 Section 8 25 Equipment passed all required tests and is ready for use See Section Functional and or Leakage Current Test Debrief Script 4 3 1 Power On Boot Up Proprietary Service Manual Voluson P8 Voluson P6 Revision 6 Section 8 25 Equipment passed all require...

Страница 540: ...ebrief Script 4 3 1 Power On Boot Up Proprietary Service Manual Voluson P8 Voluson P6 Revision 6 Section 8 33 Equipment passed all required tests and is ready for use 7 9 1 Console Test Table 8 2 System Function 4 3 2 Power Off Shutdown See Section Functional and or Leakage Current Test Debrief Script 4 3 1 Power On Boot Up Proprietary Service Manual Voluson P8 Voluson P6 Revision 6 Section 8 34 E...

Страница 541: ...l and or Leakage Current Test Debrief Script 4 3 1 Power On Boot Up Proprietary Service Manual Voluson P8 Voluson P6 Revision 6 Section 8 42 Equipment passed all required tests and is ready for use Table 8 2 System Function 4 3 2 Power Off Shutdown See Section Functional and or Leakage Current Test Debrief Script 4 3 1 Power On Boot Up Proprietary Service Manual Voluson P8 Voluson P6 Revision 6 Se...

Страница 542: ...akage Current Test Debrief Script 4 3 1 Power On Boot Up Proprietary Service Manual Voluson P8 Voluson P6 Revision 6 Section 8 46 Equipment passed all required tests and is ready for use Table 8 2 System Function 4 3 2 Power Off Shutdown See Section Functional and or Leakage Current Test Debrief Script 4 3 1 Power On Boot Up Proprietary Service Manual Voluson P8 Voluson P6 Revision 6 Section 8 47 ...

Страница 543: ...urrent Test Debrief Script 4 3 1 Power On Boot Up Proprietary Service Manual Voluson P8 Voluson P6 Revision 6 Section 8 37 Equipment passed all required tests and is ready for use 7 9 1 Console Test Table 8 2 System Function 4 3 2 Power Off Shutdown See Section Functional and or Leakage Current Test Debrief Script 4 3 1 Power On Boot Up Proprietary Service Manual Voluson P8 Voluson P6 Revision 6 S...

Страница 544: ...Console Test Table 8 2 System Function 4 3 2 Power Off Shutdown See Section Functional and or Leakage Current Test Debrief Script 10 8 5 Enclosure Leakage Current Test Proprietary Service Manual Voluson P8 Voluson P6 Revision 6 Section 8 52 Equipment passed all required tests and is ready for use 4 7 Peripheral Checks See Section Functional and or Leakage Current Test Debrief Script 4 3 1 Power On...

Страница 545: ...ction Functional and or Leakage Current Test Debrief Script 4 3 1 Power On Boot Up Proprietary Service Manual Voluson P8 Voluson P6 Revision 6 Section 8 58 Equipment passed all required tests and is ready for use N A Wait for 30 minutes to charge battery N A Disconnect Circuit Breaker of AC powerPIR Table 8 2 System Function 4 3 2 Power Off Shutdown See Section Functional and or Leakage Current Te...

Страница 546: ...GE RAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 8 198 Section 8 59 Replacement of WIRELESS LAN This page was intentionally left blank ...

Страница 547: ...ailable for the Voluson P8 Voluson P6 Table 9 1 Contents in Chapter 9 Section Description Page Number 9 1 Overview 9 1 9 2 List of Abbreviations 9 2 9 3 Parts List Groups 9 3 9 4 Plastics Covers Front Sides Rear 9 4 9 5 LCD Monitor 9 6 9 6 OPIO 9 7 9 7 Nest Box Parts 9 10 9 8 Mechanical Parts 9 13 9 9 Options 9 15 9 10 Optional Peripherals and Accessories 9 16 9 11 System Manuals 9 18 9 12 Probes ...

Страница 548: ...entral Processing Unit CSI Common Status Indicator module CWDM Continuous wave doppler module DAC Digital to Analog Converter DC Direct Current DSP Digital Signal Processing DVD Digital Video Disc DVI Digital Visual Interface ECGP Electrocardiography module EUM Electronic User Manual FRU 1 Replacement part available in parts hub FRU 2 Replacement part available from the manufacturer lead time invo...

Страница 549: ...ng front back end boards 501 515 Mechanical Parts Table 9 7 on page 9 14 Mechanical Parts 601 604 Options Table 9 8 on page 9 15 Options and power cord 701 710 Optional Peripherals and Accessories Table 9 9 on page 9 17 Optional Equipment System Manuals Voluson P8 Voluson P6 Table 9 10 on page 9 18 System Manuals 801 802 803 804 806 Probes 2D curved array Transducers 2D linear array Transducers Re...

Страница 550: ...N 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 9 4 Section 9 4 Plastics Covers Front Sides Rear Section 9 4 Plastics Covers Front Sides Rear Figure 9 1 Plastic Covers 105 101 102 115 106 110 107 108 109 111 113 114 103 116 112 ...

Страница 551: ...er for DVD 1 1 N N N Y 107 FRONT PRINTER BEZEL ASSY BT16 5722371 Front Plastic cover for printer 1 1 N N N Y 108 REAR CABINET BEZEL ASSY BT16 5722373 Rear Plastic cover for Voluson P8 Voluson P6 1 1 N N N Y 109 TOP CABINET COVER BT16 5722375 Top Cover for cabinet 1 1 N N N Y 110 FRONT PEDESTAL COVER BT16 5722376 FRONT PEDESTAL COVER BT16 1 1 N N N Y 111 BOTTOM PEDESTAL COVER BT16 5722377 BOTTOM PE...

Страница 552: ...H LCDMON 5669733 17 LED Monitor 1 1 N N N Y 202 17INCH MONITOR ARM BT16 5722185 17 MONITOR ARM BT16 1 1 N N N Y 203 17INCH MONITOR ARM COVER BT16 5720705 17 MONITOR ARM COVER BT16 1 1 N N N Y 204 Tetris HDMI Cable 5717001 Tetris HDMI Cable 1 1 N N N Y 205 Tetris Monitor Power Cable 5717002 Tetris Monitor Power Cable 1 1 N N N Y 206 Tetris Monitor USB Cable 5717003 Tetris Monitor USB Cable 1 1 N N ...

Страница 553: ...SON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 9 Renewal Parts 9 7 Section 9 6 OPIO Figure 9 3 OPIO 301 302 325 308 309 303 324 323 326 327 328 329 304 307 310 311 322 ...

Страница 554: ...astic upper cover VP6_P8 BT16 1 1 N N N Y 310 OPIO Plastic Lower Covers VP6_P8 BT16 5725311 OPIO Plastic Lower Covers VP6_P8 BT16 1 1 N N N Y 311 AN Key Top with Base Assembly Swedish VP6_P8 BT16 5722469 AN Key Top with Base Assembly Swedish VP6_P8 BT16 1 1 N N N Y 312 AN Key Top with Base Assembly Spanish VP6_P8 BT16 5722470 AN Key Top with Base Assembly Spanish VP6_P8 BT16 1 1 N N N Y 313 AN Key...

Страница 555: ... 1 1 N N N Y 322 AN Key Top with Base Assembly Swiss VP6_P8 BT16 5722680 AN Key Top with Base Assembly Swiss VP6_P8 BT16 1 1 N N N Y 323 Encoder set Tetris BT16 5725314 Encoder set Tetris BT16 1 1 N N N Y 324 VOLUSON P6_P8 OPIO RUBBER SET 5726091 VOLUSON P6_P8 OPIO RUBBER SET 1 1 N N N Y 325 RUBBER CAP VP6_P8 BT16 5727408 RUBBER CAP VP6_P8 BT16 1 1 N N N Y 326 TGC Assy 5449578 TGC Assembly 1 1 Y Y...

Страница 556: ...ION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 9 10 Section 9 7 Nest Box Parts Section 9 7 Nest Box Parts Figure 9 4 Nest Box parts 401 402 403 404 405 406 407 410 408 411 414 415 409 421 416 418 413 420 417 419 422 412 423 ...

Страница 557: ... SET 1 1 Y Y Y Y 410 OPIO Cable Assy 6020400 OPIO Cable Assy 1 1 Y Y Y Y 411 System Main FAN 5669586 System Main FAN 1 1 N N N Y 412 Type 6 COM Express with 8G Memory 5721572 i3 SOM with 8G Memory 1 1 N N N Y 413 i3 Skylake with 8Gbyte RAM 7324556 i3 Skylake with 8Gbyte RAM Usable in the system which is SW Ext 7 or later version 1 1 N N N Y 414 DPS with battery Assy 6004002 2 DPS with battery Assy...

Страница 558: ...RAFT MAY 23 2018 BASIC SERVICE MANUAL 9 12 Section 9 7 Nest Box Parts 423 Manganese Dioxide Lithum Coin Battery 5173038 Manganese Dioxide Lithum Coin Battery CR2450 1 1 Y Y Y Y Table 9 6 Nest Box Item Part Name Part Number Description Qty FRU BT13 BT14 BT15 BT16 ...

Страница 559: ... VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL Chapter 9 Renewal Parts 9 13 Section 9 8 Mechanical Parts Figure 9 5 Mechanical Parts 501 502 503 504 505 506 507 508 509 510 511 512 513 514 ...

Страница 560: ... VP BT16 1 1 N N N Y 505 Cable Hook BT16 5725317 Cable Hook 1 1 N N N Y 506 Probe Holder 2 Rubber 5433118 Probe Holder 2 1 1 Y Y Y Y 507 Probe Holder 3 Rubber 5433119 Probe Holder 3 1 1 Y Y Y Y 508 Air Filter 5374942 Air Filter 1 1 Y Y Y Y 509 Power Cord Fix Bracket 110V 5497621 Power Cord Fix Bracket 110V 1 1 Y Y Y Y 510 Power Cord Fix Bracket 220V 5497620 Power Cord Fix Bracket 220V 1 1 Y Y Y Y ...

Страница 561: ... Vertical TV probe holder 1 1 Y Y Y Y 602 Horizontal TV probe holder 5408514 Horizontal TV probe holder 1 1 Y Y Y Y 603 3 button footswitch Rev 1 1 5380960 2 3 button Foot Swtich 1 1 Y Y Y Y 618 ECG Module Assy 5669519 ECG Module Assy 1 1 N Y Y Y 619 Power Cord Fix Bracket 110V 5497621 Power Cord Fix Bracket 110V 1 1 Y Y Y Y 620 Power Cord Fix Bracket 220V 5497620 Power Cord Fix Bracket 220V 1 1 Y...

Страница 562: ...018 BASIC SERVICE MANUAL 9 16 Section 9 10 Optional Peripherals and Accessories Section 9 10 Optional Peripherals and Accessories 9 10 1 Printers and options Figure 9 7 Optional Peripherals and Accessories Printers 708 707 701 703 704 705 711 709 710 713 715 714 712 702 716 ...

Страница 563: ...ital SONY UP D25MD 1 N A Y Y Y Y 705 Color Laser Printer HP M452 066E8922 For 220V 066E8923 For 110V Color Laser Printer HP M452 1 N A Y Y Y Y 706 PrinterInstallation Kit BW and Color H44892LC 5722386 BW Printer Bracket Color Printer Bracket Cable set 1 N A N N N Y 707 WLAN Stick KTZ196269 H48671DT Wireless Network Interface Netgear WLAN Adapter 1 N A Y Y Y Y 708 ECG kit assy 5669842 consists of E...

Страница 564: ...54430 143 Basic User Manual Korean 5654430 144 Basic User Manual Russian 5654430 145 Basic User Manual Polish 5654430 150 Basic User Manual Greek 5654430 151 Basic User Manual Hungarian 5654430 153 Basic User Manual Slovakian 5654430 154 Basic User Manual Czech 5654430 155 Basic User Manual Turkish 5654430 159 Basic User Manual Danish 5654430 160 Basic User Manual Norwegian 5654430 161 Basic User ...

Страница 565: ...obes Replacement Parts Sub section Description Page Number 9 12 1 2D Probes Curved Array Transducers 9 19 9 12 2 2D Probes Linear Array Transducer 9 20 9 12 3 Real Time 4D Volume Probes 9 21 Figure 9 8 2D curved array Transducers Table 9 12 2D curved array Transducers Item Part Name Part Number Description Qty CRU FRU 801 4C RS 5488477 broadband curved array transducer 2 0 5 0 MHz 128 Elements App...

Страница 566: ...2 2D Probes Linear Array Transducer Figure 9 9 2D Linear Array Transducer Table 9 13 2D linear array Transducers Item Part Name Part Number Description Qty CRU FRU 803 12L RS 5499501 Small and lightweight linear array transducer 4 0 10 0 MHz 192 Elements Applications Small parts Peripheral vascular Pediatrics Orthopedics 1 1 803 ...

Страница 567: ...4D Volume Probes Table 9 14 Real Time 4D Volume Probes Item Part Name Part Number Description Qty CRU FRU 804 RIC5 9A RS KTZ304128 Real time 4D endocavity broadband electronic curved array transducer 3 8 9 3 MHz 192 Elements field of view max 180 Applications Gynecology Obstetrics Transrectal 1 1 805 RAB2 6 RS KTZ303982 Real time 4D broadband electronic curved array transducer 2 0 5 0 MHz 128 Elem...

Страница 568: ...ed Array Transducers Figure 9 11 2D Phased Array Probe The part is only applied for VP8 Table 9 15 2D Probes Phased Array Transducers Item Part Name Part Number Description Qty CRU FRU 807 3Sc RS 47237516 H45041DL Broadband phased array transducer 1 0 4 0 MHz 64 Elements Applications Abdominal Cardiology Obstetrics Pediatrics Cephalic 1 1 1 807 ...

Страница 569: ...nformation Affected Products BT13 BT14 BT15 BT16 SPC310 Array type Part Number Part Description VP6 VP8 VP6 VP8 VP6 VP8 VP6 VP8 Curved 5488477 4C RS x x x x x x x x 5499516 E8C RS x x x x x x 5722023 IC9 RS x x Linear 5499501 12L RS x x x x x x x x Phased 47237516 3Sc RS x x x x x x x Volume KTZ304063 RIC5 9A RS x x KTZ304128 RIC5 9A RS x x KTZ303990 RIC5 9W RS x x x x x x KTZ303981 RAB2 5 RS x x ...

Страница 570: ...N P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 9 24 Section 9 13 Power Cord Section 9 13 Power Cord Figure 9 12 Power Cords 901 902 903 904 913 912 911 910 909 908 906 905 914 907 ...

Страница 571: ... Power Cord for Denmark 1 1 1 906 Power Cord for Italy 5182940 Power Cord for Italy 1 1 1 907 Power Cord for Argentina 5182942 Power Cord for Argentina 1 1 1 908 Power Cord for Israel 5182453 Power Cord for Israel 1 1 1 909 Power Cord 110V for Canada USA JPN 2388982 Power Cord 110V for Canada USA JPN 1 1 1 Power Cord for Taiwan 5792418 Power Cord for Taiwan 1 1 1 910 Power Cord 220V for EU 2327990...

Страница 572: ...le biopsy needle guide for probe RIC5 9A RS needle diameter 1 8 mm 1 N 1002 RICSingle andgle bracket H48681GF Disposable biopsy needle guide for probe RIC5 9A RS needle diameter 1 2 1 6 mm 1 N 1003 RAB Single angle bracket H48681ML Non Sterile Single Angld Bracket needle guide for porbe RAB2 6 RS needle diameter 0 6mm 2 1 mm 1 N 1004 12L RS Multiangle bracket H40432LC Non Sterile Multi Angle Brack...

Страница 573: ...8385NA Non Sterile Multi Angle Bracket needle guide starter kit for probe 4C RS needle diameter 0 6 mm 2 1 mm 1 N 1007 IC9 RS Biopsy guide disposable H48691YW Disposable Biopsy needle guide for probe IC9 RS needle diameter 1 65 mm 1 N 1008 IC9 RS Biopsy guide reuseable H48701MM Reusable Biopsy needle guide for probe IC9 RS 1 N 1009 IC9 RS Biopsy guide H48701MN Reusable Biopsy needle guide for prob...

Страница 574: ...GE RAFT VOLUSON P8 VOLUSON P6 DIRECTION 5723243 REVISION 6 DRAFT MAY 23 2018 BASIC SERVICE MANUAL 9 28 Section 9 14 Biopsy Needle Guide ...

Страница 575: ...mance Read this chapter completely and familiarize yourself with the procedures before performing a task Table 10 1 Contents in Chapter 10 Section Description Page Number 10 1 Overview 10 1 10 2 Why do Maintenance 10 2 10 3 Maintenance Task Schedule 10 2 10 5 Tools Required 10 7 10 6 System Maintenance 10 8 10 7 Using a Phantom Optional 10 14 10 8 Electrical Safety Tests 10 14 10 9 When There s To...

Страница 576: ...aining records for a quality assurance program Please contact us for coverage information and or price for service Section 10 3 Maintenance Task Schedule 10 3 1 How often should care maintenance tasks be performed The Customer Care Schedule see page 10 3 specifies how often your Voluson P8 Voluson P6 should be serviced and outlines items requiring special attention NOTE It is the customer s respon...

Страница 577: ... Unit Check Daily Control Console movement Check Translation Rotation and Height Adjustment Elevation more frequently at Mobile Units Trackball Check Check proper operation Cursor movement X Y direction If failure occurs go to trackball cleaning Trackball Cleaning Remove trackball ring open the trackball housing and take out the trackball to clean it with soft tissue and screwdriver shaft Please r...

Страница 578: ...or as required by your facilities QA program Probe Phantom optional Checks Check axial and lateral resolution see Basic User Manual Technical specifications Check Gain and TGC changes vary the focus and check reaction on screen Also after corrective maintenance or as required by your facilities QA program Functional Checks of all probes Section 10 6 2 on page 10 9 Also after corrective maintenance...

Страница 579: ...t clip 9 6757 6 pc Needle File Set 9 41594 Screwdriver Blade 1 8 in x 4 in 9 9487 Utility Knife 9 41581 Screwdriver Blade 3 16 in x 4 in 9 45341 Pliers Vice Grip 10 inch 9 39451 20 Steel Tape locking Spring load 9 3001 Xacto Pen Knife 9 GH807 Ratchet Offset Slotted 9 HT62002 Solder Aid Fork and Hook 68 412 Ratchet Offset Phillips 9 4099 Mirror Round Telescoping 9 GH130 Tapered Reamer 9 GH3001 Stee...

Страница 580: ... 1 2 inch 9 44067 Socket 1 1 16 in for 1 2 in drive 9 45092 Pliers Linesman 8 1 2 inch 9 42679 Socket 10MM Hex for 1 2 in drive 2273333 9 42882 Punch Pin 3 32 inch 9 44262 Extension 10 inch for 1 2 in drive 2273405 9 42884 Punch Pin 5 32 inch 9 4258 3 8 inch to 1 2 inch Adapter 9 42886 Punch Pin 1 4 inch 9 34374 3 8 inch Metric Socket Set 12 PT 9 42973 Cold Chisel 1 2 inch 9 44311 16mm Socket 12 p...

Страница 581: ...cables for 200 to 240 V system 3M 2204 Large adjustable wrist strap 3M 2214 Small adjustable wrist strap 3M 3051 conductive ground cord Anti Static Vacuum Cleaner 46 194427P278 46 194427P279 120V 230V Safety Analyzer Any Calibrated Electrical Safety Analyzer compliant with AAMI ESI 1993 or IEC 60601 or AS NZS 3551 QIQ Phantom Optional E8370RB RMI Grayscale Target Model 403GS B W Printer Cleaning S...

Страница 582: ... Fill in the top of the Ultrasound Inspection Certificate see page 10 32 Note all probes and system options 3 Power up Turn the system power on and verify that all fans and peripherals turn on Watch the displays during power up to verify that no warning or error messages are displayed 4 Probes Verify that the system properly recognizes all probes 5 Displays Verify proper display on the LCD monitor...

Страница 583: ... Mode Verify basic PD Mode Power Doppler Mode operation Check the basic system controls that affect this mode of operation 5 Doppler Modes Verify basic Doppler Mode operation PW Check the basic system controls that affect this mode of operation 6 3D Mode Verify basic 3D Mode operation Check the basic system controls that affect this mode of operation 7 RealTime 4D Mode optional Verify basic RealTi...

Страница 584: ...rder Verify record capabilities of the DVD Recorder Clean heads and covers if necessary 6 DICOM Verify that DICOM is functioning properly Send an image to a DICOM device 7 DVD Drive Verify that the DVD drive reads writes properly export recall image in Image Management System Archive 8 ECG Verify basic operation with customer 9 Cellular Modem Power Supply SIM Card Placing SIM Card Antenna Loop Fas...

Страница 585: ...stored away from dust and cigarette smoke Do not use alcohol or benzene to clean the drive 6 Wheels Brakes Check all wheels and casters for wear and verify operation of foot brake to stop the unit from moving and release mechanism Check all wheel locks and swivel locks for proper operation 7 Cables Connectors Check all internal cable harnesses and connectors for wear and secure connector seating P...

Страница 586: ... Thoroughly check the system probe connectors and remove dust from inside the connector sockets if necessary Visually check for bent damaged or missing pins 2 Probe Holder Clean probe holders they may need to be soaked to remove excess gel CAUTION CAUTION Failure to follow the prescribed cleaning or disinfection procedures will void the probe s warranty DO NOT soak or wipe the lens with any produc...

Страница 587: ...ocked head mechanics etc The RIC Holder especially developed for RIC Real time 4D endocavity probes guarantees that the sensitive probe head does not hit the bottom of the container during the disinfection procedure NOTE Operation instructions are supplied with each RIC Holder KTZ225469 Endocavity Soaking Cup Kit is available at GE Accessory Order Nr E8385TA To guarantee that the sensitive probe h...

Страница 588: ...ORMED AT LEAST EVERY 12 MONTHS ACCORDING TO THE REQUIREMENTS OF THE PATIENT SAFETY STANDARD IEC60601 1 standard including national deviations ONLY TRAINED PERSONS ARE ALLOWED TO PERFORM THE SAFETY INSPECTIONS MENTIONED ABOVE CAUTION CAUTION To avoid electrical shock the unit under test must not be connected to other electrical equipment Remove all interconnecting cables and wires The unit under te...

Страница 589: ...nd close Protective earth with normal polarity 0 5 mA 0 5 mA Open neutral and close Protective earth with reverse polarity 0 5 mA 0 5 mA Table 10 15 Earth Leakage Current Limits Conditions Country USA Others Close Protective earth and close neutral with normal polarity 0 3 mA 0 5 mA Close Protective earth and close neutral with reverse polarity 0 3 mA 0 5 mA Open neutral and close Protective earth...

Страница 590: ...al with reverse polarity 0 01 mA 0 01 mA Open Protective earth and close neutral with normal polarity 0 05 mA 0 05 mA Open Protective earth and close neutral with reverse polarity 0 05 mA 0 05 mA Open neutral and close Protective earth with normal polarity 0 05 mA 0 05 mA Open neutral and close Protective earth with reverse polarity 0 05 mA 0 05 mA Conditions Country USA Others ...

Страница 591: ...and close Protective earth with normal polarity 0 05 mA 0 05 mA Open neutral and close Protective earth with reverse polarity 0 05 mA 0 05 mA Conditions Country USA Others Close Protective earth and close neutral with normal polarity 5 mA 5 mA Close Protective earth and close neutral with reverse polarity 5 mA 5 mA Conditions Country USA Others Close Protective earth and close neutral with normal ...

Страница 592: ...on between Mains and Applied Limits refers to the measurement of leakage current flow which would flow from mains to patient if the patient were to come into contact with mains voltage Measurement limits per IEC 60601 1 Ed 3 Medical Electrical Equipment Safety Standards Table 3 Measurement limits per IEC 60601 1 Ed 3 Medical Electrical Equipment Safety Standards Table 4 The following tests are per...

Страница 593: ...ter and noting the combination of lights that are illuminated Any problems found should be reported to the hospital immediately and the receptacle should not be used NOTE No outlet tester can detect the condition where the Neutral grounded supply conductor and the Grounding protective earth conductor are reversed If later tests indicate high leakage currents this should be suspected as a possible ...

Страница 594: ...ug the black chassis cable into the meter s CHASSIS connector and attach the black chassis cable clamp to an exposed protectively earthed metal part of the Voluson P8 Voluson P6 unit such as Potential equilibrium connector 4 Set the meter s FUNCTION switch to the RESISTANCE position 5 Set the meter s POLARITY switch to the OFF center position 6 Measure and record the Ground wire resistance This sh...

Страница 595: ...ocedure The test verifies the isolation of the power line from the enclosure The testing meter is connected from accessible non protectively earthed parts of the case to ground Measurements should be made with the ground open and closed with power line polarity normal and reversed and with the neutral open and closed Record the highest reading of current NOTE Color of Power outlet cable i USA and ...

Страница 596: ...ty Analyzer tool compliant with AAMI ESI 1993 or IEC 60601 or AS NZS 3551 When using a test instrument its power plug may be inserted into the wall outlet and the equipment under test is plugged into the receptacle on the panel of the meter This places the meter in the grounding conductor and the current flowing from the case to ground will be indicated in any of the current ranges The maximum all...

Страница 597: ...vely earthed parts of the equipment to ground Measurements should be made with power line polarity normal and reversed and with the neutral open and closed Record the highest reading Table 10 23 Typical Data Sheet for Enclosure Leakage Current Tester Neutral Tester Polarity Switch ProtectiveEarth Switch Test 1 Wheel Test 2 LCD Optional Test 3 Optional Test 4 Enter Name of tested peripheral here Cl...

Страница 598: ...n or Green yellow ii Others Hot Brown Neutral Blue Ground Green yellow Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer tool compliant with AAMI ESI 1993 or IEC 60601 or AS NZS 3551 When using a test instrument its power plug may be inserted into the wall outlet and the equipment under test is plugged into the receptacle on the panel of the meter This p...

Страница 599: ...d to find the worst case condition 10 8 7 2 Generic Procedure Measurements should be made with the ground open and closed with power line polarity normal and reversed and with the neutral open and closed For each combination the operating controls such as the lead switch should be operated to find the worst case condition NOTE Color of Power outlet cable i USA and Canada Hot Black Tester Neutral T...

Страница 600: ...ted on each probe depending on its design Small variations in probe leakage currents are normal from probe to probe Other variations will result from differences in line voltage and test lead placement It is abnormal if no leakage current is measured If no leakage current is detected check the configuration of the test equipment Tester Neutral Tester Polarity Switch Ground Switch Measured Earth Le...

Страница 601: ...IEC 60601 1 Measure current flow from Probe to all ECG leads connected together The maximum allowable limit for earth leakage current is shown in Table 10 18 on page 10 17 Figure 10 6 Set Up for Probe Leakage Current Table 10 26 Typical Data Sheet for Type BF Patient Leakage Current Test Tester Neutral Tester Polarity Switch Ground Switch Measured Leakage Current Closed Normal Closed Closed Normal...

Страница 602: ...he system being tested must not be touched by patients users or anyone during testing CAUTION CAUTION Equipment damage possibility Never switch the Polarity and the status of Neutral when the unit is powered ON Be sure to turn the unit power OFF before switching them using the POLARITY switch and or the NEUTRAL switch Otherwise the unit may be damaged Table 10 28 Typical Data Sheet for Type BF Pat...

Страница 603: ...pen and closed For each combination ECG leads must be active simultaneously Please refer to Type CF Patient Leakage Current Test ECG Leads part for detail Table 10 30 Typical Data Sheet for Total Patient Leakage Current Type CF probes to ground Closed Reversed Closed CAUTION CAUTION Equipment damage possibility Never switch the Polarity and the status of Neutral when the unit is powered ON Be sure...

Страница 604: ...e CF probes DANGER DANGER Electric Shock Hazard Line voltage is applied to the ECG leads during this test To avoid possible electric shock hazard the system being tested must not be touched by patients users or anyone during testing CAUTION CAUTION Equipment damage possibility Never switch the Polarity and the status of Neutral when the unit is powered ON Be sure to turn the unit power OFF before ...

Страница 605: ...t lies with the probe or the System NOTE Each probe will have some amount of leakage dependent on its design Small variations in probe leakage currents are normal from probe to probe Other variations will result from differences in line voltage and test lead placement The maximum allowable leakage current for body surface contact probe differs from an inter cavity probe Be sure to enter the correc...

Страница 606: ...facture Date Probe 1 Frequency Scan Format Model Number Serial Number Probe 2 Frequency Scan Format Model Number Serial Number Probe 3 Frequency Scan Format Model Number Serial Number Probe 4 Frequency Scan Format Model Number Serial Number Probe 5 Frequency Scan Format Model Number Serial Number Probe 6 Frequency Scan Format Model Number Serial Number Probe 7 Frequency Scan Format Model Number Se...

Страница 607: ...Value Allowed Value Measured OK Comments Enter Name of tested peripheral here Closed Normal 0 3mA USA 0 5mA Others Closed Reversed 0 3mA USA 0 5mA Others Open Normal 1mA Open Reversed 1mA Tester Neutral Tester Polarity Switch Protective Earth Switch Max Value Allowed Test 1 Wheel Test 2 LCD Optional Test 3 Optional Test 4 OK Comments Enter Name of tested peripheral here Closed Normal Closed 0 1mA ...

Страница 608: ...01mA Closed Reversed Open 0 05mA Open Normal Closed 0 05mA Open Reversed Closed 0 05mA Tester Neutral Tester Polarity Switch Ground Switch Max Value Allowed Value Measured OK Comments Closed Normal Closed 0 01mA Closed Normal Open 0 05mA Closed Reversed Closed 0 01mA Closed Reversed Open 0 05mA Open Normal Closed 0 05mA Open Reversed Closed 0 05mA Tester Neutral Tester Polarity Switch Ground Switc...

Страница 609: ...ured OK Comments Closed Normal Closed 0 1mA Closed Normal Open 0 5mA Closed Reversed Closed 0 1mA Closed Reversed Open 0 5mA Open Normal Closed 0 5mA Open Reversed Closed 0 5mA Tester Neutral Tester Polarity Switch Ground Switch Max Value Allowed Value Measured OK Comments Closed Normal Closed 0 01mA Closed Normal Open 0 05mA Closed Reversed Closed 0 01mA Closed Reversed Open 0 05mA Open Normal Cl...

Страница 610: ...ith all probes as expected Accepted by ______________________________________________________________________ Tester Neutral Tester Polarity Switch Ground Switch Max Value Allowed Value Measured OK Comments Closed Normal Closed 0 05mA Closed Normal Open 0 1mA Closed Reversed Closed 0 05mA Closed Reversed Open 0 1mA Open Normal Closed 0 1mA Open Reversed Closed 0 1mA Tester Neutral Tester Polarity ...

Страница 611: ......