GE H
EALTHCARE
PROPRIETARY TO GE
D
IRECTION
5344303-100, R
EVISION
3
V
IVID
P3 S
ERVICE
M
ANUAL
1-16
Section 1-3 - Safety Considerations
1-3-6
Warning labels location Tables
L1 (Figure 1-1)
D1: Standard LCD monitor may rotate while
transporting. Bind the system securely to prevent
damage in transportation.
D2: To avoid injury by tipping over, DO NOT PUSH
THIS UNIT FROM THE SIDES.
D3: DO NOT place a finger, hand or any object on
the joint of the monitor or monitor arm to avoid injury
when moving the monitor and monitor arm.Placing
objects on top of the monitor may cause the monitor
to tilt with the falling objects resulting in injury to the
operator.Do not place any objects on the monitor.
L2 (Figure 1-2)
D4: This machine should be used in compliance
with law. Some jurisdictions restrict certain uses,
such as gender determination.
L3 (Figure 1-3)
D5: Identification and Rating Plate.
L4 (Figure 1-4)
D6: “Consult accompanying document” is intended
to alert the user to refer to the operator manual or
other instructions when complete information
cannot be provided on the label.
D7: Do not use the following devices near this
equipment: cellular phone, radio receiver, mobile
radio transmitter, radio controlled toy, etc. Use of
these devices near this equipment could cause this
equipment to perform outside the published
specifications. Keep power to these devices turned
off when near this equipment.
D8: Caution: Federal Law restricts this device to
sale by or on the order of a physician.
D9: The CE Mark of Conformity indicates this
equipment conforms with the Council Directive 93/
42/EEC
D10: CISPR CAUTION: The Vivid P3 conforms to
the CISPR11, Group 1, Class A of the international
standard for Electromagnetic disturbance
characteristics.
D11: WEEE Label: This symbol indicates that the
waste of electrical and electronic equipment must
not be disposed as unsorted municipal waste and
must be collected separately. Please contact an
authorized representative of the manufacturer for
information concerning the decommissioning of
your equipment.
D12: The equipment weighs approximately 80 kg
(176 lbs). To avoid possible injury and equipment
damage when transporting from one area of use to
another:
• Be sure the pathway is clear.• Limit movement to a
slow careful walk.• Use two or more persons to
move the equipment on inclines or long distance.
D14: Gost Label (Russia Only).
D15: TUV Label: TUV Listing and Certification Mark
is used to designate conformance to nationally
recognized product safety standards. The Mark
bears the name and/or logo of the testing laboratory,
product category, safety standard to which
conformity is assessed, and a control number
D16: Circuit breaker Label
D17: Line Input voltage Label.
L10 (Figure 1-5)
D18: Signal ground point label CAUTION: This is
only for “FUNCTIONAL GROUNDING”, NOT
“PROTECTIVE EARTH”.
Lamp contains Mercury Dispose According to Stete/
Local Law.