GE H
EALTHCARE
PROPRIETARY TO GE
D
IRECTION
5344303-100, R
EVISION
3
VIVID P3 S
ERVICE
M
ANUAL
8-46
Section 8-4 - Mechanicals
8-4-9
Left EMI Cover (Part of EMI Cover Assy FRU P/N: 5315111)
This is a description on how to remove and replace the Left EMI Cover.
(P/N : 5308964)
8-4-9-1
Tools
•
Common Phillips screwdrivers
8-4-9-2
FRU BOM.
This FRU Kit P/N:5308964 consists of following parts.
8-4-9-3
Needed Manpower
•
1 person, 15 minutes
8-4-9-4
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-4-9-5
Removal Procedure
1) Remove Left cover. Refer
for details on how to remove Left cover.
3) Eight screws to be unfastened of LH-side EMI cover of Vivid P3 system.
2) Remove LH-side EMI cover. Refer
for details on how to remove Left cover.
Figure 8-38 LH-side EMI cover
PART NO
DESCRIPTION
5308964
LH EMI cover
2334842
Fastener- M3 X 6 ECO-FIX
1
2
3
4
Slide this way to remove the cover
Slide this way to reinstall the cover
5
6
7
8