GE H
EALTHCARE
PROPRIETARY TO GE
D
IRECTION
5344303-100, R
EVISION
3
VIVID P3 S
ERVICE
M
ANUAL
8-44
Section 8-4 - Mechanicals
8-4-8
Right EMI Cover (Part of EMI Cover Assy FRU P/N :5315111)
This is a description on how to remove and replace the Right EMI Cover
(P/N :5308965)
.
8-4-8-1
Tools
•
Common Phillips screwdrivers
8-4-8-2
FRU BOM.
This FRU Kit P/N:5308965 consists of following parts.
8-4-8-3
Needed Manpower
•
1 person, 15 minutes
8-4-8-4
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-4-8-5
Removal Procedure
1) Remove Right cover. Refer
for details on how to remove Right cover.
2) Use the scewdriver and loosen the eight screws (1,2,3,4,5,6,7,8)
3) Slide and remove RH-side EMI cover. Refer
for details on how to remove
Right cover.
Figure 8-37 RH-side EMI cover
PART NO
DESCRIPTION
5308965
RH EMI cover
2334842
Fastener- M3 X 6 ECO-FIX
1
2
3
4
5
6
7
8
RH-side
EMI cover
Slide to right to re-install cover
Slide to left to remove the cover