GE H
EALTHCARE
PROPRIETARY
TO
GE
D
IRECTION
5344303-100, R
EVISION
3
VIVID P3 S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8-23
8-3-2
Probe Holder (FRU P/N :5340677) / Gel Bottle Holder (FRU P/N :5340678)
This is a description on how to remove and replace the Probe holder Assembly.
8-3-2-1
Needed Manpower
•
1 person, 15 minutes
8-3-2-2
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-3-2-3
Removal procedure
1) Pull the probe holder out by slightly pushing it upwards with your hands.
Figure 8-19 Probe Holder
8-3-2-4
Mounting procedure
1.) Install the new parts in the reverse order of removal.
8-3-2-5
Functional Checkout Procedure
Functional Test
Debrief Script
Do Visual Inspection and ensure
no scratches observed and
fitment is done properly
Service Manual Direction 5344303-100, Section 8-3-2. Equipment passes all required
tests and is ready for use.
Gel Holder
Probe Holder