18
CRxFlex Operating Manual V1.11
Cleaning of Image Plates
Warning: Reduced image quality possible if a wrong cleaning agent is used.
Only use the dedicated digital screen cleaner.
Follow the instructions which are part of the image plate precisely.
Each image plate is delivered with cleaning instructions.
When cleaning an image plate, follow these instructions.
Preventive Maintenance
Regular preventive maintenance needs to be done once per year for one shift operation. The
maintenance interval will also depend on the environmental condition.
Preventive maintenance has to be done by a qualified field service engineer.
Not performing the regular maintenance by appropriately certified people can have impact on
warranty commitments.
Quality Control
Apply regular quality control according local regulations.