CRxFlex Operating Manual V1.1
25
Starting the CRxFlex
To start the CRxFlex:
Make sure:
The CRxFlex is connected to the workstation either by point-to-point network or via
the local area network.
The workstation is running the appropriate acquisition software.
1. Press the power button.
The CRxFlex starts the following boot-up sequence:
Initialization of all components.
Functional test of all components.
Check for presence of cassettes and/or image plates.
Establish connection to the workstation.
During the self-test, which takes approx. 1 minute, the CRxFlex status indicator is blinking
yellow.
Note: During the self-test, you cannot activate any function.
If the CRxFlex has completed the self-test successfully, the CRxFlex enters the operator
mode and the status indicator is continuously lighting up in green.