CRxFlex Operating Manual V1.1
31
When the CRxFlex has finished erasing the cassette, the upper part of the status
indicator is not lit and the lower part is constantly lit in green.
4
Remove the cassette from the cassette slot.
5
To erase a second cassette, the erase mode has to be accessed again.
Cleaning the Optics Unit
Cleaning the optics unit is required if stripes parallel to the image plate movement can be
seen in the image. When you recognize this type of artifact, when using the CRxFlex, clean
the optics unit using the cleaning brush.
To clean the optics unit, proceed as follows:
1. Open the cassette unit.
2. Take out the cleaning brush.