15
of the pressure gauge to be tested, "confirm" refers to the completion of a
group of test points, calculation and display of the error value, which can be
saved and printed after the completion of the test.
Click "start" to start detection. When the current value Pt in the first
column will rise to the preset value of the first set of test points 0.200mpa, it will
stop as shown in Figure 12. In the process of inflation, the status bar will
prompt "the first test". When the preset pressure value is reached, it will display
"detection completed".
Figure 12
Move the cursor to "input" in Figure 12, and click the "select" cursor to
move to the input value in the first column, as shown in Figure 13:
Figure 13
Rotate the mouse left and right to modify the value, input the displayed value of
the pressure gauge to be tested, then "click the selected" cursor to return to
"input", click "OK" to complete the first test, as shown in Figure 14: