background image

DU-WS-0012-01 Rev B 5/24/16 GCX Corp. Page 7 of 15 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

7.1 Adjusting Counterbalance 

The VHM-PL arm is shipped in its upmost position (top of 
height adjustment range) 

– be sure it is still in this position 

before proceeding. The Arm must be unlocked to perform 
the counterbalance adjustment. This is done by pulling the 
handle grip 

and inserting a 1/8” hex wrench into the handle 

grip holes as shown. When grip is released, the hex wrench 
will stay in position and hold the handle grip in the un-lock 
position (the arm will freely move up and down). 

Remove 

hex key after completing counterbalance adjustment. 

Installation Note: VHM-PL (locking) 

7.0 Adjusting the Arm 

Counterbalance Adjustment 

 The Counterbalance Adjustment Screw is located under the Arm in front of the rear pivot 

assembly. Grasp the mounted device or handle (making sure the hex wrench does not fall out) and move the Arm to a level 
horizontal position. Using a 5/32

” hex wrench, tighten (+) or loosen (-) the Counterbalance Adjustment Screw. 

Counterbalance is correctly adjusted when the mounted device can be moved up or down with minimal force and does not 
rise or fall after releasing the Arm.

 

Installation Note:

 Adjustments are factory pre-set. However, adjustments to counterbalance, tilt tension, swivel tension, 

and pivot tensions may all be required to achieve optimal arm performance. Detailed instructions for making adjustments 
follow in the section below. When properly adjusted the mounted device will move easily throughout the arms full range of 
motion. Refer to the Routine Maintenance Check List in Section 11 for a quick guide to these functional checks. 

 

Caution:

 The counterbalance procedure is performed with the device(s) and accessories installed on the Arm.  

When unlocking the arm to perform the counterbalance procedure, use caution as it may be necessary to 
support the weight of the arm/mounted device(s). This process may require an additional person. Use caution 
while performing this procedure. 
 

Counterbalance 

Adjustment Screw

 

Содержание VHM-P

Страница 1: ...des mounting hardware and instructions OSHPD Pre Approval WC 0002 03 05 07 09 11 and 15 Dual Cable Hooks with 3 7 6 cm Down Post for Cable Management WMM 0006 13 www gcx com GCX is an ISO 13485 regist...

Страница 2: ...range 16 40 6 cm total travel Lateral range 90 degrees Load range 15 to 30 lb 6 8 to 13 6 kg or 25 to 45 lb 11 3 to 20 4 kg Maximum weight for flat panel 30 lb 13 6 kg Available with 8 or 14 Extensio...

Страница 3: ...11 0 Routine Maintenance 14 12 0 Cleaning the Arm 14 13 0 Troubleshooting the VHM P and VHM PL 15 Installation Time Approximately 10 15 minutes The purpose of this manual is to describe general instal...

Страница 4: ...es 2 3 4 6 9 7 VHM PL locking 1b 5 8 VHM P non locking 1a 1 0 Parts Reference The following parts and hardware are included and labeled accordingly Item Description Quantity 1 VHM P a or VHM PL b Chan...

Страница 5: ...ies Arms Do not remove device when arm is in a lowered position Refer to Installation Guide for additional information Remove the Device only when the Arm is at the highest position Safety Label Locat...

Страница 6: ...result in damage to equipment or injury If you are unable to adjust the arm or require service contact GCX technical support at 800 228 2555 or 1 707 773 1100 The mounted device or arm may move sudden...

Страница 7: ...ecure position Before mounting the Arm verify that the channel has been installed and approved in accordance with the channel installation guide included with the channel 5 4 Set Screws 2 upper Set Sc...

Страница 8: ...d slide device laterally into the Mounting Plate until the Plunger snaps into the clearance hole at the front of the Mounting Adapter Nylon Locking Screws After the device is installed tighten the 4 N...

Страница 9: ...al position Using a 5 32 hex wrench tighten or loosen the Counterbalance Adjustment Screw Counterbalance is correctly adjusted when the mounted device can be moved up or down with minimal force and do...

Страница 10: ...et wrench until desired tension is achieved Total adjustment is approximately to turn Caution Do not remove the Swivel Tension Nut 7 3 Swivel Tension Adjustment Swivel Tension Nut Tilt Tension Adjustm...

Страница 11: ...arts included for cable routing Cable Sleeves 2 Length 30 5 77 5cm Cable Ties 8 Cable Cover Cable Sleeve Four 4 Cable Anchors are provided at the front and four 4 at rear of the arm to secure cables a...

Страница 12: ...being sure there is adequate length to avoid damage to a connector or unintended disconnection of a cable Insert cables into sleeve Cable Sleeve must have a 1 4 rotation overlap minimum CAUTION Cable...

Страница 13: ...d below the Arm Figure B Pivot the Cable Cover up towards the Arm Figure C Gently squeeze the front of the Cable Cover and insert tabs of the Cable Cover into the mating slots to secure the Cable Cove...

Страница 14: ...Loop 2 from the front Cable Anchor point to the rear Cable Anchor point after securing the cables to the front Cable Anchor points and already determining Service Loop 1 from device to front anchor po...

Страница 15: ...el Thread and secure the captive screw into the mounting hole under the pivot assembly with a 2 Phillips screw driver Do not overtighten the screw 9 1 Installing the upper Rear Pivot Cover Place the P...

Страница 16: ...quipment refer to the specific instructions delivered with those products 12 0 Cleaning the VHM P and VHM PL Arm 11 0 Routine Maintenance Routine Maintenance Check List Section Counterbalance With the...

Страница 17: ...m with compatible Load Rating section 3 0 Mounted device difficult to move up or down Arm not counterbalanced correctly for weight of mounted device Perform counterbalance adjustment section 7 1 Arm p...