PEDIATRIC HAL®
|
User Guide
Routine Maintenance | 73
7. TROUBLESHOOTING
7.1 CONNECTIVITY
Review the UNI user guide for the most common troubleshooting topics regarding the UNI software.
Probable Cause
Solution
Battery indicator question mark
Disconnect and reconnect battery charger and allow the simulator
to charge fully (an indicator light will turn green when finished
charging).
Virtual monitor not connecting
Check that both computers/tablets are connected to the same wifi
network, check that firewalls are turned off on both computers/
tablets, check that the IP & Port numbers match from the UNI
software on the Gaumard Monitors.
BP Cuff reading not accurate
Recalibrate through the UNI software by clicking the blue gear icon
in the upper right hand corner -> setup-> calibration.
Probable Cause
Solution
CO2 tubing is not
connected properly
Check the equipment set up and the cartridge's connection to
the simulator, and adjust Lung CO2 level in the UNI software.
7.2 CO2 NOT DETECTED