PEDIATRIC HAL®
|
User Guide
26 | Working with Pediatric HAL
Ventilation Calibration
•
The ventilation calibration wizard records the performance average of
five ventilations as the benchmark for correct ventilation.
•
Perform the actions requested by the calibration wizard following the most current
CPR guidelines. The CPR window will then evaluate the provider's performance
based on the benchmark recorded during the calibration process.
1.
Click the blue gear icon in the upper right hand
corner of the UNI software then click Setup->
Calibration -> Ventilations and click Next.
2.
After clicking Next on the Calibration
window, select to perform a Normal
Calibration if you have never calibrated
before. Then click the Next button.
3.
Click next to open the calibration
window, then click Start.
4.
Click Start and perform a correct ventilation
when prompted for a total of five (5
ventilations. A green oval will indicate that
the ventilation was successfully recorded.
5.
Follow the cues on the calibration window and
click Finish after performing all calibrations.